ATTUNE 2546-00-421Recalls & reported problems
Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION SLEEVE COUNTER TORQUE WRENCH · Class 1
All models of ATTUNE (1475)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295440345), received between Nov 2, 2018 and Jul 3, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility4
- Material Deformation1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE WRENCH IS STRIPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22393372 ↗MalfunctionAug 8, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT WHILE TORQUING/TIGHTENING THE STEM TO THE SLEEVE DURING ASSEMBLY, THE SLEEVE WRENCH BECAME STUCK TO THE SLEEVE. THE SURGEON WAS ABLE TO REMOVE IT WITHOUT FURTHER ISSUE BUT THE WRENCH SHOULD NOT BECOME BOUND UP ON THE IMPLANT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19939317 ↗MalfunctionAug 5, 2024Device-Device Incompatibility; Device-Device Incompatibility
ADDITIONAL INFORMATION WAS RECEIVED: A. WAS THERE ANY ISSUE WITH THE SLEEVE? NO B. WAS THERE ANY PATIENT HARM/CONSEQUENCE RELATED TO THE EVENT? NO C. WAS THERE ANY SURGICAL DELAY DUE TO REPORTED EVENT? - A SMALL DELAY WHILE DR FOSTER WORKED THROUGH THE EVENT DESCRIBED ABOVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19906123 ↗MalfunctionJun 12, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE SLEEVE WAS SLIPPING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19521850 ↗MalfunctionMay 16, 2022Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE WRENCH IN THE ATTUNE REVISION FEMORAL BROACHING TRAY IS STARTING TO STRIP. THERE WAS NO ADDITIONAL TIME ADDED TO CASE. PATIENT WASN'T INJURED DURING THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14407723 ↗MalfunctionNov 2, 2018Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING ASSEMBLY OF THE FEMORAL ATTUNE SLEEVE STEM CONSTRUCT, THE SLEEVE WRENCH WOULD NOT STRIP/SLIP AROUND AS THE SURGEON TORQUED IT WITH THE TORQUE WRENCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8032530 ↗
Other WRENCH devices
- MONARCH 277040510Medos International Sàrl · WRENCH199 reports
- PINNACLE 2217-50-008Depuy (Ireland) · WRENCH108 reports
- SIGMA 96-1673Depuy (Ireland) · WRENCH102 reports
- P.F.C. 86-5189Depuy (Ireland) · WRENCH92 reports
- Veptr 03.641.004Synthes GmbH · WRENCH86 reports
- RECLAIM 2975-00-300Depuy (Ireland) · WRENCH51 reports
- PINNACLE 2217-50-006Depuy (Ireland) · WRENCH34 reports
- NA 388.261Synthes GmbH · WRENCH32 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page