CORAIL L971208Recalls & reported problems
Depuy (Ireland) · CORAIL HIP SYSTEM CEMENTLESS FEMORAL STEM HA COATED 12/14 AMT 125 DEGREES STANDARD COLLAR SIZE 8 · Class 2 · Implant
All models of CORAIL (121)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295455387), received between Mar 17, 2020 and Nov 4, 2025 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Loss of Osseointegration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Loss of Osseointegration2
- Device-Device Incompatibility1
- Fracture1
- Osseointegration Problem1
Patient problems
- Inadequate Osseointegration3
- Bone Fracture(s)2
- Insufficient Information2
- No Code Available2
- Pain2
- Limb Fracture1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryNov 4, 2025Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT ALL IMPLANTS APPEARED TO BE LOOSE. NO PATIENT COMPLICATIONS. AFFECTED SIDE- LEFT HIP.
Patient problems: Inadequate Osseointegration
MAUDE report 23468059 ↗MalfunctionJul 11, 2024Device-Device Incompatibility; Device-Device Incompatibility
ADDITIONAL INFORMATION RECEIVED STATES THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: Insufficient Information
MAUDE report 19722332 ↗InjuryOct 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SURGEON HAD AN INTRA OP PERIPROSTHETIC FRACTURE ON (B)(6) 2023. DURING AN ANTERIOR HIP, AFTER FINAL IMPACTION OF THE REAL IMPLANT, A LATERAL FRACTURE WAS VISIBLE ON THE INTRA OP XRAY. IT IS UNCLEAR THE PRECISE MOMENT THE FRACTURE OCCURRED. SURGEON WAS USING THE MATTA BROACH HANDLE AND A SIZE 8 CORAIL STEM. DIFFICULT EXPOSURE, PARTICULARLY HARD BONE, A TIGHT DISTAL CANAL, AN OBESE PATIENT WERE ALL FACTORS THAT LIKELY CONTRIBUTED TO THE INCIDENT, BUT ON SUNDAY SURGEON MESSAGED THAT SURGEON WAS CONCERNED THE ANGLE OF IMPACTION OF THE MATTA BROACH HANDLES MAY HAVE ALSO BEEN A CONTRIBUTING FACTOR. SURGEON ALSO REQUESTED TO KNOW IF THERE HAS BEEN ANY OTHER REPORTED INCIDENTS OF FRACTURES USING THESE BROACH HANDLES. SURGERY WAS DELAYED 15 MINUTES. TWO CABLES WERE PLACED TO ADDRESS THE FRACTURE.
Patient problems: Limb Fracture
MAUDE report 17939112 ↗InjuryNov 8, 2021Fracture; Fracture
CUSTOMER IS REPORTING A DISTAL FRACTURE OF CORAIL HIP-STEM. DOI: (B)(6) 2019, DOR: UNKNOWN, UNKNOWN SIDE.
Patient problems: Insufficient Information
MAUDE report 12771811 ↗InjuryAug 9, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO PAIN. DOI: (B)(6) 2021 - DOR: (B)(6) 2021 (LEFT HIP).
Patient problems: Pain
MAUDE report 12294885 ↗InjuryOct 13, 2020Loss of Osseointegration; Loss of Osseointegration
THE PATIENT WAS REVISED DUE TO PAIN SECONDARY TO PERIPROSTHETIC FRACTURE. IT WAS ALSO INDICATED THAT THERE WAS LOOSENING OF THE STEM AT THE BONE TO IMPLANT INTERFACE. DOI: UNKNOWN, DOR: (B)(6) 2020, AFFECTED SIDE: LEFT HIP.
Patient problems: Bone Fracture(s); Inadequate Osseointegration; No Code Available; Pain
MAUDE report 10673564 ↗InjuryMar 17, 2020Loss of Osseointegration; Loss of Osseointegration
FEMORAL COMPONENT LOOSENING THOUGHT TO HAVE CRACKED CALCAR DURING PRIMARY SURGERY. CORAIL STEM STILL HAD HA ON PROXIMAL AREA OF STEM WHEN EX PLANTED. DISTAL STEM HA APPEARED TO BE RESORTED.
Patient problems: Bone Fracture(s); Inadequate Osseointegration; No Code Available
MAUDE report 9845795 ↗
FDA clearance and classification
The FDA regulates product code KWL (prosthesis, hip, hemi-, femoral, metal) under 21 CFR 888.3360 as a Class II device. 21 CFR 888.3360 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the CORAIL L971208
Has the CORAIL L971208 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295455387), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CORAIL L971208?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Mar 17, 2020 and Nov 4, 2025. In those reports, the most common specific device problems are “Loss of Osseointegration” (2), “Device-Device Incompatibility” (1) and “Fracture” (1); the most common patient problems are “Inadequate Osseointegration” (3), “Bone Fracture(s)” (2) and “Pain” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CORAIL L971208?
None of the reports are classified as deaths, 6 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CORAIL L971208 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the CORAIL L971208, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand CORAIL, model L971208). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, hip, hemi-, femoral, metal devices
- UHR UH1-48-28Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal2 recalls3 reports
- "UNITED" USTAR II Knee System 2515-3111United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3213United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- STAGEONE 431207Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3115United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- ACTIS 101011050Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal73 reports
- UHR UH1-45-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
- UHR UH1-46-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page