Skip to content

ACTIS 2011-01-230Recalls & reported problems

Depuy (Ireland) · ACTIS REAMER SIZE 4/5 · Class 2

All models of ACTIS (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295495680), received between Nov 30, 2018 and Feb 20, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

1 other recall cover devices cleared under the same FDA submission (K150862); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Failure to Cut2
  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1
  • Not Applicable1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS REAMER WOULD NOT LOCK INTO THE HUDSON ADAPTER. SURGICAL DELAY WAS APPROXIMATELY FIVE MINUTES FOR OBTAINING ANOTHER TRAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21426032 ↗
  • MalfunctionDec 10, 2024Break; Break

    IT WAS REPORTED THAT THE REAMER HANDLE IS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20894996 ↗
  • MalfunctionJul 19, 2023Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS WERE DULL. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17350994 ↗
  • MalfunctionJul 14, 2023Break; Break

    IT WAS REPORTED THAT THE 4/5 ACTIS DISTAL REAMER BROKE WHILE REAMING. IT BROKE CLEANLY WITH NO DEBRIS OR PIECES BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17327341 ↗
  • MalfunctionJul 22, 2021Break; Break

    IT WAS REPORTED AS "WHILE REAMING THE FEMORAL CANAL, THE ACTIS 4/5 REAMER SHAFT SHEARED AND BROKE OFF, LEAVING THE REAMER TIP BEHIND. REAMER TIP WAS RETRIEVED AND REMOVED FROM FEMUR." SURGICAL DELAY OF 15 MINUTES.

    Patient problems: Insufficient Information

    MAUDE report 12211712 ↗
  • MalfunctionFeb 5, 2021Break; Failure to Cut; Break; Failure to Cut

    SURGEON NOTICED THE PROBLEM ON ACTIS REAMER STABILIT AND SAW THE RISK OF THE BROKEN REAMER DURING SURGERY. THICKNESS OF THE REAMER BODY SEAMS ARE TOO WEAK EVEN THOUGH NO EVENT HAPPENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11289490 ↗
  • MalfunctionNov 30, 2018Break; Break

    ADAPTER BROKE.

    Patient problems: Not Applicable

    MAUDE report 8121389 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150862. One submission often covers many models, so check the product and lot details in each record.

  • Z-0650-2017 · Oct 25, 2016 · Terminated
    ACTIS Flex Reamers SZ 0/1; 2/3; 4/5; 6/7; 8/9; 10/11; 12, Product is sold non-sterile, these instruments are packed non-sterile, within a rigid tube with foam protectors, with labe

Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices

All Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page