ACTIS 2011-01-230Recalls & reported problems
Depuy (Ireland) · ACTIS REAMER SIZE 4/5 · Class 2
All models of ACTIS (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295495680), received between Nov 30, 2018 and Feb 20, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
1 other recall cover devices cleared under the same FDA submission (K150862); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Failure to Cut2
- Device-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information1
- Not Applicable1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE ACTIS REAMER WOULD NOT LOCK INTO THE HUDSON ADAPTER. SURGICAL DELAY WAS APPROXIMATELY FIVE MINUTES FOR OBTAINING ANOTHER TRAY. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21426032 ↗MalfunctionDec 10, 2024Break; Break
IT WAS REPORTED THAT THE REAMER HANDLE IS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20894996 ↗MalfunctionJul 19, 2023Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS WERE DULL. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17350994 ↗MalfunctionJul 14, 2023Break; Break
IT WAS REPORTED THAT THE 4/5 ACTIS DISTAL REAMER BROKE WHILE REAMING. IT BROKE CLEANLY WITH NO DEBRIS OR PIECES BROKEN OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17327341 ↗MalfunctionJul 22, 2021Break; Break
IT WAS REPORTED AS "WHILE REAMING THE FEMORAL CANAL, THE ACTIS 4/5 REAMER SHAFT SHEARED AND BROKE OFF, LEAVING THE REAMER TIP BEHIND. REAMER TIP WAS RETRIEVED AND REMOVED FROM FEMUR." SURGICAL DELAY OF 15 MINUTES.
Patient problems: Insufficient Information
MAUDE report 12211712 ↗MalfunctionFeb 5, 2021Break; Failure to Cut; Break; Failure to Cut
SURGEON NOTICED THE PROBLEM ON ACTIS REAMER STABILIT AND SAW THE RISK OF THE BROKEN REAMER DURING SURGERY. THICKNESS OF THE REAMER BODY SEAMS ARE TOO WEAK EVEN THOUGH NO EVENT HAPPENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11289490 ↗MalfunctionNov 30, 2018Break; Break
Recalls under the same FDA submission
These recalls involve devices cleared under K150862. One submission often covers many models, so check the product and lot details in each record.
- Z-0650-2017 · Oct 25, 2016 · TerminatedACTIS Flex Reamers SZ 0/1; 2/3; 4/5; 6/7; 8/9; 10/11; 12, Product is sold non-sterile, these instruments are packed non-sterile, within a rigid tube with foam protectors, with labe
Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices
- Novation 130-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls156 reports
- Novation 130-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls79 reports
- CORAIL 3L92511Depuy (Ireland) · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate158 reports
- Acumatch A Series 138-36-29Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate3 recalls1 report
- Novation 136-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls43 reports
- Novation 130-32-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls33 reports
- Novation 132-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls26 reports
- Novation 136-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.