ACTIS 2011-01-260Recalls & reported problems
Depuy (Ireland) · ACTIS REAMER SIZE 10/11 · Class 2
All models of ACTIS (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295495710), received between Feb 5, 2021 and Jul 19, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K150862); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Failure to Cut2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 19, 2023Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS WERE DULL. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17351028 ↗MalfunctionFeb 5, 2021Break; Failure to Cut; Break; Failure to Cut
SURGEON NOTICED THE PROBLEM ON ACTIS REAMER STABILIT AND SAW THE RISK OF THE BROKEN REAMER DURING SURGERY. THICKNESS OF THE REAMER BODY SEAMS ARE TOO WEAK EVEN THOUGH NO EVENT HAPPENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11289540 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K150862. One submission often covers many models, so check the product and lot details in each record.
- Z-0650-2017 · Oct 25, 2016 · TerminatedACTIS Flex Reamers SZ 0/1; 2/3; 4/5; 6/7; 8/9; 10/11; 12, Product is sold non-sterile, these instruments are packed non-sterile, within a rigid tube with foam protectors, with labe
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page