DELTA XTEND 1307-64-142Recalls & reported problems

Depuy (Ireland) · DELTA XTEND Lateralized Glenosphere +4mm Standard 042mm · Class 2 · Implant

All models of DELTA XTEND (194)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295502463), received between Mar 15, 2021 and Jul 10, 2026 — classified as 9 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert9
  • Adverse Event Without Identified Device or Use Problem7
  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Unspecified Infection4
  • Joint Laxity3
  • Insufficient Information2
  • Pain1
  • Unspecified Musculoskeletal problem1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED FOR LOOSENING AND INFECTION. IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO HUMERAL STEM LOOSENING AT BONE TO CEMENT INTERFACE. AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Unspecified Infection

    MAUDE report 25778428 ↗
  • InjuryJun 8, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT HAS HAD MULTIPLE FALLS AND THE GLENOSPHERE DISASSOCIATED FROM THE BASE PLATE. POLY WAS DEFORMED FROM RUBBING AGAINST BASEPLATE. DOCTOR DECIDED THAT A CTA HEAD WOULD BE THE BEST CHOICE FOR THIS PATIENT. TOOK OUT POLY, GLENOSPHERE, BASEPLATE, AND SCREWS AND REPLACED WITH THE CTA HEAD. IT WAS STATED THAT THE LOOSENING OF SHOULDER GLENOSPHERE WAS UNKNOWN. DOI: UNKNOWN DOR: (B)(6), 2026 AFFECTED SIDE: UNKNOWN.

    Patient problems: Unspecified Musculoskeletal problem

    MAUDE report 25433246 ↗
  • MalfunctionMar 25, 2026Difficult to Insert

    IT WAS REPORTED THAT INTRAOPERATIVELY DURING SHOULDER SURGERY, IT WAS NOT POSSIBLE TO FIX BOTH GLENOSHERES IN THE METAGLENE. ANOTHER GLENOSPHERE WAS USED AS A THIRD ATTEMPT AND THE SURGERY WAS COMPLETED. THERE WAS LESS THAN 10 MINUTES OF SURGICAL DELAY. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24691223 ↗
  • MalfunctionMar 25, 2026Difficult to Insert

    IT WAS REPORTED THAT INTRAOPERATIVELY DURING SHOULDER SURGERY, IT WAS NOT POSSIBLE TO FIX BOTH GLENOSHERES IN THE METAGLENE. ANOTHER GLENOSPHERE WAS USED AS A THIRD ATTEMPT AND THE SURGERY WAS COMPLETED. THERE WAS LESS THAN 10 MINUTES OF SURGICAL DELAY. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24691165 ↗
  • InjuryDec 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INSTABILITY. DOI: (B)(6) 2022, DOR: (B)(6) 2025, AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Joint Laxity

    MAUDE report 23790845 ↗
  • InjuryNov 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT THE REVISION OF REVERSE LEFT SHOULDER DUE TO INFECTION. THIS REPORT IS RELATED COMPLAINT (B)(4) WHICH CAPTURES ADDITIONAL IMPACTED PRODUCTS.

    Patient problems: Unspecified Infection

    MAUDE report 23569949 ↗
  • InjuryNov 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS COMPLAINING OF HUMERAL PAIN. STEM WAS FOUND TO BE IN GOOD STABLE CONDITION AS WELL AS METAGLENE. SURGEON REPLACED SCREWS, GLENOSPHERE AND PLOY TO BIGGER SIZES TO GIVE MORE STABILITY. DOI: (B)(6) 2024. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT SHOULDER.

    Patient problems: Joint Laxity; Pain

    MAUDE report 23488232 ↗
  • InjuryAug 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTS WERE REMOVED DUE TO INSTABILITY. THE PATIENT HAD REPORTED NO INSTABILITY DURING THE FIRST FEW YEARS AND WAS SATISFIED WITH THE RESULTS UNTIL EXPERIENCING ISSUES OVER THE PAST 6 MONTHS. TO ACHIEVE APPROPRIATE STABILITY, THE SURGEON UPSIZED THE GLENOSPHERE AND HUMERAL CUP. THERE WERE NO PATIENT CONSEQUENCES AND NO SURGICAL DELAY OCCURRED. DOI: ON (B)(6) 2022, DOR: ON (B)(6) 2025, AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Joint Laxity

    MAUDE report 22751472 ↗
  • MalfunctionJul 19, 2024Difficult to Insert; Difficult to Insert

    INTRAOPERATIVE IT WAS NOT POSSIBLE TO ENGAGE THE GLENOSPHERE WITH THE METAGLENE. ANOTHER IMPLANT WAS USED. THIS WAS POSSIBLE WITHOUT ISSUES. A SURGICAL DELAY WAS NOT REPORTED, NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19781798 ↗
  • MalfunctionFeb 15, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT IT IS DIFFICULT TO ENGAGE THE SCREWS OF THE TWO GLENOSPHERES WITH THE THE METAGLENE. SURGICAL DELAY ABOUT TEN MINUTES, NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18710639 ↗
  • MalfunctionFeb 12, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT INTERIOR SCREW OF GLENOSPHERE WOULD NOT ENGAGE INTO COLLET. WOULD THREAD AND ENGAGE BUT WOULD RELEASE ONCE MORE TORQUE WAS APPLIED. IT WAS UNKNOWN, IF THERE WAS A SURGICAL DELAY. DOI- UNKNOWN, DOR- ON (B)(6) 2024, AFFECTED SIDE- UNKNOWN SHOULDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18685479 ↗
  • MalfunctionNov 29, 2023Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT DURING IMPLANTATION OF GLENOSPHERE INTO METAGLENE SURGEON FELT TIGHTENING OF THE SCREW INTO THE COLLET THEN THE FEELING IT HAD RELEASED. REPEATED THE PROCESS AGAIN WITH SAME RESULT. ULTIMATELY FELT SCREW TIGHTEN INTO COLLET AND MOVED ON WITH PROCEDURE. NO TIME DELAY. AFFECTED SIDE: LEFT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18229588 ↗
  • InjuryJun 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OF A DELTA XTEND DUE TO INFECTION. THE PATIENT HAD A PREVIOUS REVISION SURGERY A FEW YEARS AGO. UNABLE TO CONFIRM EXACT DATE. REASON FOR THIS WAS DUE TO GLENOID NOTCHING AND PAIN. AT THIS REVISION THE 155 EPI WAS REMOVED, GLENOID WAS REMOVED. UNSURE IF THE METAGLENE AND SCREWS WERE TOO. THE STEM WAS RETAINED, AND IT WAS CONVERTED TO A 145 EPI WITH LATERALISED GLENOSPHERE. PATIENT NOTES WERE UNAVAILABLE TO CONFIRM EXACT DATES AND LOT NUMBERS. I'M UNABLE TO CONFIRM WHAT PC THIS WAS LOGGED AGAINST TOO. TODAY WE REMOVED ALL IMPLANTS AS PER IMPACTED PRODUCTS. 4X SCREWS WERE ALSO REMOVED, BUT I'M UNABLE TO CONFIRM THE REF AND LOT NUMBERS FOR THESE. A SPACER WAS IMPLANTED UNTIL THE INFECTION CLEARS. DOI: UNKNOWN, DOR; (B)(6) 2023, AFFECTED SIDE: RIGHT.

    Patient problems: Unspecified Infection

    MAUDE report 17059069 ↗
  • MalfunctionJan 6, 2023Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT UPON INSERTION OF THE GLENOSPHERE INTO THE METAGLENE, THE SCREW WOULD NOT PROPERLY TIGHTEN INTO THE COLLET. SCREW WOULD TIGHTEN AND THEN SLIP AND BECOME LOOSE AFTER SUFFICIENT FORCE TO TIGHTEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16112309 ↗
  • InjuryApr 18, 2022Difficult to Insert; Difficult to Insert

    THE POST ON GLENOSPHERE WASN'T SEATING IN METAGLENE. DOE: (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 14138825 ↗
  • InjuryNov 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A. REVISION FOR INFECTION - WAS THERE ANY SUGGESTION BY ANYONE INDICATING THAT THERE WAS OR MAY BE DEFICIENCY WITH THE PRODUCT(S)? IF YES, PLEASE PROVIDE DETAILS. THE SURGEON DID NOT NOTE ANY DEFICIENCY WITH THE IMPLANTS CAUSING THE INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 12843880 ↗
  • MalfunctionMar 15, 2021Difficult to Insert; Difficult to Insert

    INTRAOPERATIVELY DURING PRIMARY SHOULDER-TEP RIGHT SIDE IT WAS NOT POSSIBLE TO CONNECT THE GLENOSPHERE TO THE METAGLENE. ANOTHER GLENOSPHERE WAS AVAILABLE AND PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT HARM, NO ADVERSE CONSEQUENCES. SURGICAL DELAY < 10 MIN.

    Patient problems: Insufficient Information

    MAUDE report 11486115 ↗

FDA clearance and classification

Cleared through 510(k) K192855 on Feb 24, 2020 (decision: Substantially Equivalent). Submitted by Depuy(Ireland) as “DELTA XTEND(TM) Reverse Shoulder System”. The FDA review took 143 days. FDA record ↗

Cleared through 510(k) K183077 on May 17, 2019 (decision: Substantially Equivalent). Submitted by Depuy(Ireland) as “Delta Xtend Reverse Shoulder System”. The FDA review took 193 days. FDA record ↗

The FDA regulates product code PHX (shoulder prosthesis, reverse configuration) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Intended to be used to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.

Common questions about the DELTA XTEND 1307-64-142

Has the DELTA XTEND 1307-64-142 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295502463), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the DELTA XTEND 1307-64-142?

The FDA's MAUDE database lists 17 adverse event reports for this device, received between Mar 15, 2021 and Jul 10, 2026. In those reports, the most common specific device problems are “Difficult to Insert” (9) and “Device Dislodged or Dislocated” (1); the most common patient problems are “Unspecified Infection” (4), “Joint Laxity” (3) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the DELTA XTEND 1307-64-142?

None of the reports are classified as deaths, 9 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the DELTA XTEND 1307-64-142 FDA approved?

It was cleared through 510(k) K183077 on May 17, 2019; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the DELTA XTEND 1307-64-142 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the DELTA XTEND 1307-64-142, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand DELTA XTEND, model 1307-64-142). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page