DELTA XTEND 2317-70-024Recalls & reported problems

Depuy (Ireland) · DELTA XTEND Central Screw Metaglene augmented metaglene reamer handle · Class 1

All models of DELTA XTEND (194)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295532460), received between Aug 19, 2022 and Dec 9, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility5
  • Mechanical Jam1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 9, 2024Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT TWO INSTRUMENTS IN THIS CASE MALFUNCTIONED: THE FIRST AUGMENT REAMER HANDLE HAD A BUTTON NEAR THE BACK END WHICH IS SPRING LOADED TO CAPTURE THE HUDSON ADAPTOR CONNECTION. THIS BUTTON WAS STICKING AND REQUIRED THE TECH TO TAP IT ON THE SIDE REPEATEDLY TO SPRING THE BOTTOM UP. THIS SHOULD NOT BE NECESSARY. THE SECOND CENTRAL SCREW, METAGLENE AUGMENTED METAGLENE REAMER HEAD BROKE WHILE REAMING THE GLENOID. THIS LEFT FRAGMENTS IN THE PATIENT WHICH REQUIRED THE SURGEON TO REMOVE.

    Patient problems: Insufficient Information

    MAUDE report 20885764 ↗
  • MalfunctionMar 27, 2024Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement

    UPON RUNNING THE CSM REAMER, IT CAUSED THE DRILL TO SHAKE NOTICEABLY, AND REMAINED DOING SO AFTER RE-ASSEMBLY A NUMBER OF TIMES. THERE WAS A 5 MINUTE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18987355 ↗
  • MalfunctionJun 9, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE REAMER LOCKED UP AND WOULDN'T SPIN. IT ALSO GOT STUCK ON THE HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17098714 ↗
  • MalfunctionFeb 21, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT WHEN PUTTING TOGETHER THE FULL WEDGE AUGMENTED REAMER, THE CONSTRUCT WOULD NOT PROPERLY ENGAGE. THERE WAS A 15-20 MINUTES OF SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16409367 ↗
  • MalfunctionAug 30, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE CSM AUGMENT REAMER HAND PIECE TOLERANCE IS DEFECTIVE AND WILL NOT ACCEPT THE REAMER SHAFT BECAUSE IT DOESN¿T FIT. THIS INSTRUMENT HAS NEVER BEEN USED. THE REAMER SHAFT TOLERANCES WERE DOUBLE CHECKED WITH A DIFFERENT HANDLE AND THEY WERE CORRECT. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15324422 ↗
  • MalfunctionAug 19, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE AUGMENT HANDLE REAMER WILL NOT TAKE AUGMENT REAMER HOUSING. AUGMENT REAMER HOUSING WILL SEAT INTO A DIFFERENT AUGMENT HANDLE REAMER OF THE SAME PART AND LOT NUMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15257989 ↗

Other Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment devices

All Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment devices →