Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295532460), received between Aug 19, 2022 and Dec 9, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility5
- Mechanical Jam1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 9, 2024Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT TWO INSTRUMENTS IN THIS CASE MALFUNCTIONED: THE FIRST AUGMENT REAMER HANDLE HAD A BUTTON NEAR THE BACK END WHICH IS SPRING LOADED TO CAPTURE THE HUDSON ADAPTOR CONNECTION. THIS BUTTON WAS STICKING AND REQUIRED THE TECH TO TAP IT ON THE SIDE REPEATEDLY TO SPRING THE BOTTOM UP. THIS SHOULD NOT BE NECESSARY. THE SECOND CENTRAL SCREW, METAGLENE AUGMENTED METAGLENE REAMER HEAD BROKE WHILE REAMING THE GLENOID. THIS LEFT FRAGMENTS IN THE PATIENT WHICH REQUIRED THE SURGEON TO REMOVE.
Patient problems: Insufficient Information
MAUDE report 20885764 ↗MalfunctionMar 27, 2024Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement
UPON RUNNING THE CSM REAMER, IT CAUSED THE DRILL TO SHAKE NOTICEABLY, AND REMAINED DOING SO AFTER RE-ASSEMBLY A NUMBER OF TIMES. THERE WAS A 5 MINUTE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18987355 ↗MalfunctionJun 9, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE REAMER LOCKED UP AND WOULDN'T SPIN. IT ALSO GOT STUCK ON THE HANDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17098714 ↗MalfunctionFeb 21, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT WHEN PUTTING TOGETHER THE FULL WEDGE AUGMENTED REAMER, THE CONSTRUCT WOULD NOT PROPERLY ENGAGE. THERE WAS A 15-20 MINUTES OF SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16409367 ↗MalfunctionAug 30, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE CSM AUGMENT REAMER HAND PIECE TOLERANCE IS DEFECTIVE AND WILL NOT ACCEPT THE REAMER SHAFT BECAUSE IT DOESN¿T FIT. THIS INSTRUMENT HAS NEVER BEEN USED. THE REAMER SHAFT TOLERANCES WERE DOUBLE CHECKED WITH A DIFFERENT HANDLE AND THEY WERE CORRECT. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15324422 ↗MalfunctionAug 19, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE AUGMENT HANDLE REAMER WILL NOT TAKE AUGMENT REAMER HOUSING. AUGMENT REAMER HOUSING WILL SEAT INTO A DIFFERENT AUGMENT HANDLE REAMER OF THE SAME PART AND LOT NUMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15257989 ↗
Other Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment devices
- INHANCE 6505-10-103Depuy (Ireland) · Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment209 reports
- INHANCE 6205-10-102Depuy (Ireland) · Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment33 reports
- GLOBAL CAP 2230-81-110Depuy Orthopaedics, Inc. · Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment5 reports
- INHANCE 6205-60-020Depuy (Ireland) · Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment4 reports
- DELTA XTEND 2317-70-424Depuy (Ireland) · Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment3 reports
- Stainless Steel Guide Wires 01-9473-3115Orthopediatrics Corp. · Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment2 reports
- DELTA XTEND 2317-70-324Depuy (Ireland) · Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment2 reports
- INHANCE 6105-30-050Depuy (Ireland) · Instrument, surgical, orthopedic, dc-powered motor and accessory/attachment2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗