Skip to content

INHANCE 6200-42-145Recalls & reported problems

Depuy (Ireland) · INHANCE SHOULDER SYSTEM HUMERAL HEAD TRIAL diameter 42mm X 14.5mm · Class 1

All models of INHANCE (202)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295539179), received between Feb 16, 2022 and Jan 20, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Crack1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 20, 2026Break

    IT WAS REPORTED THAT THE LOCKING TABS WERE BROKEN FROM SCRUB TURNING ADAPTOR WRONG WAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24116631 ↗
  • MalfunctionDec 19, 2025Break; Break

    IT WAS REPORTED THAT THE HUMERAL HEAD TRIAL DEVICE WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23850065 ↗
  • MalfunctionDec 2, 2025Crack; Crack

    IT WAS REPORTED THAT THE TAB CRACKED/BROKEN OFF HUMERAL HEAD TRIAL FOR INHANCE. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23693244 ↗
  • MalfunctionMay 27, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTATION BROKE FROM THE SET. IT WAS NOTICED BEFORE THE CASE. A NEW SET WAS BROUGHT TO NOT DELAY THE CASE. THERE WAS NO DELAY. NO BROKEN PIECES LEFT ON FIELD OR PATIENT. NO SURGEON COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22088757 ↗
  • MalfunctionFeb 19, 2024Break; Break

    IT WAS REPORTED THAT THE TAB BROKE. SURGERY WAS NOT DELAYED DUE TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18730776 ↗
  • MalfunctionJan 11, 2023Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT SEVERAL TRIAL BROKEN INSTRUMENTS. WEAR AND TEAR OF PRODUCT. TRIAL GLENOID CHIPPED WHILE IMPACTING. TRIAL HEADS SNAPPED WHILE ENGAGING/DISENGAGING OFFSET TAPER ADAPTER. HEAD IMPACTOR BROKE WHILE IMPLANTING FINAL IMPLANTS. THERE WAS 10 MINUTES SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 16137021 ↗
  • MalfunctionFeb 16, 2022Break; Break

    IT WAS REPORTED THAT ON INSPECTION 3 BROKEN TABS ON THE HEAD TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13543133 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page