Skip to content

ATTUNE 1518-20-906Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM TIBIAL INSERT FIXED BEARING MEDIAL STABILIZED LEFT AOX 9, 6mm · Class 2 · Implant

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295541974), received between Jul 31, 2025 and Sep 24, 2026 — classified as 5 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5

Patient problems

  • Joint Laxity2
  • Fall1
  • Pain1
  • Unspecified Infection1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient was described as globally lax and the physician believed the patient may have stretched out. The poly was upsized from 6 mm to 10 mm, and the patient required device revision or replacement. DOI-Unknown.DOR (b)(6) 2026.Affected side-Left.

    Patient problems: Joint Laxity

    MAUDE report 50886306 ↗
  • InjuryJun 8, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WAS DOING FINE UNTIL HE RUPTURED HIS PATELLA TENDON. SURGEON REPLACED INSERT AND REPAIRED HIS PATELLA TENDON. DOI: (B)(6) 2024, DOR: (B)(6) 2026, AFFECTED SIDE- LEFT KNEE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 25426011 ↗
  • InjuryFeb 17, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT FELL. DOI: UNKNOWN. DOR: (B)(6) 2026. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Fall

    MAUDE report 24367451 ↗
  • InjuryDec 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT INFECTED TKA. LINER EXCHANGE WITH I&D. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 23876499 ↗
  • InjuryJul 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO PAIN AND INSTABILITY AFTER PRIMARY TK. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Joint Laxity; Pain

    MAUDE report 22675229 ↗

Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page