INHANCE 6500-40-104Recalls & reported problems
Depuy (Ireland) · INHANCE SHOULDER SYSTEM REVERSE LINER TRIAL RETENTIVE Ø40 +4mm · Class 1
All models of INHANCE (202)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295547419), received between Sep 5, 2025 and Sep 25, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility3
- Break2
- Material Deformation1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Device-Device Incompatibility; Scratched Material
It was reported that during a Shoulder Repair trial reduction of the shoulder, the surgeon informed that the poly trial was disassociating from the shell trial. The representative noticed the poly was pretty scratched up and the tabs to click into the shell trial were damaged. All parts were retrieved. There were no delays or adverse events that took place.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50897306 ↗MalfunctionFeb 12, 2026Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT DURING SURGERY WHILST TRIALING, OUR TRIAL POLYS WERE NOT ATTACHING WELL TO THE SHELL TRIAL. TECH POINTED OUT THE LOCKING TAB MECHANISMS WERE BENT AND NOT FULLY ENGAGING. AFTER INSPECTION, THE TRIALS WERE FULLY INTACT AND NOTHING WAS BROKEN, JUST BENT, THEREFORE NOT LEAVING ANYTHING INTO BODY. NO ADVERSE AFFECTS TOWARDS THE PATIENT AND NO TIME WAS LOST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24321769 ↗MalfunctionSep 11, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT THE TABS ON TRIAL LINERS WERE BROKEN AND THEY WOULD NOT STAY LOCKED IN TO SHELL TRIAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23026524 ↗MalfunctionSep 5, 2025Break; Break
IT WAS REPORTED THAT THE TRIAL LINER SHELL TAB WAS BROKEN AND NOTICED AFTER TRIALING BETWEEN MULTIPLE RETENTIVE AND REGULAR LINERS. AS WELL AS THE 30 STAR DRIVE WENT THRU SETS AND BOTH WERE STRIPPED AFTER PREVIOUS USAGE OF IMPLANTATION OF GLENOSHOHERES. IT CAUSED ISSUES WHEN REMOVING THE GLENOSPHERE FOR THE REVISION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22976282 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page