ACTIS 3010-01-050Recalls & reported problems
Depuy (Ireland) · ACTIS HIP SYSTEM BROACH SIZE 5 · Class 2
All models of ACTIS (97)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10603295548645), received between Jan 18, 2023 and Sep 22, 2026 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility9
- Break8
- Material Deformation1
- Material Integrity Problem1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- Insufficient Information2
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Break
It was reported that the size 5 actis broach snapped.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853258 ↗MalfunctionNov 20, 2025Device-Device Incompatibility; Material Integrity Problem; Device-Device Incompatibility; Material Integrity Problem
IT WAS REPORTED THAT THE 5 ACTIS BROACH WAS GETTING STUCK TO KINCISE ATTACHMENT AND NECK TRIAL WOULDN¿T COME OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23605985 ↗MalfunctionOct 10, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT THERE ARE SEVERE STRESS RINGS ON ACTIS BROACHES WHERE THEY ATTACH TO BROACH HANDLE. SURGEON FREQUENTLY BREAKS THEM AND WHEN THEY BREAK IT IS EXTREMELY HARD TO REMOVE FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23272708 ↗MalfunctionNov 7, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT THE PROXIMAL PORTION OF THE SIZE 5 ACTIS BROACH WAS SLIGHTLY BENT, NOT ALLOWING THE NECK TRIALS TO SIT PROPERLY ON THE TRIAL BROACH.
Patient problems: Insufficient Information
MAUDE report 20630371 ↗MalfunctionJul 9, 2024Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE CALCAR PLANAR WAS NOT GOING DOWN FLUSH ON THE ACTIS 5 FEMORAL BROACH. IT WAS BELIEVED TO BE DUE TO WEAR AROUND THE NECK/BROACH ATTACHMENT. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19706616 ↗MalfunctionMar 6, 2024Break; Break
ADDITIONAL INFORMATION RECEIVED: THERE WAS NO SPECIFIC ALLEGATIONS OF DEFICIENCY AGAINST THE FOUR OTHER SIZE 5 BROACHES THAT WERE RETURNED, NOT THAT WE KNOW OF.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18844488 ↗MalfunctionSep 1, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT SURGEON WAS REAMING A CALCAR REAMER OVER THE BROACH. NOTHING UNUSUAL HAPPENED DURING THE REAMING , BUT WHEN THE SURGEON FINISHED REAMING, THE CALCAR REAMER WOULD NOT COME OFF THE BROACH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17669811 ↗MalfunctionAug 2, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT THE EVENT DID NOT HAPPEN IN SURGERY. THE BROACHES ARE ¿STICKING¿ IN THE HANDLES AND A FEW SIZE 5 HAVE ACTUALLY BROKEN IN SPD WHEN TRYING TO SEPARATE FROM THE HANDLES. NO OTHER INFORMATION IS AVAILABLE. ADDITIONAL REPORTS ARE CAPTURED UNDER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17440058 ↗MalfunctionAug 1, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT THE EVENT DID NOT HAPPEN IN SURGERY. THE BROACHES ARE ¿STICKING¿ IN THE HANDLES AND A FEW SIZE 5 HAVE ACTUALLY BROKEN IN SPD WHEN TRYING TO SEPARATE FROM THE HANDLES. NO OTHER INFORMATION IS AVAILABLE. ADDITIONAL REPORTS ARE CAPTURED UNDER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17434976 ↗MalfunctionJun 12, 2023Break; Break
IT WAS REPORTED THAT THE BROACH BROKE AFTER THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17108762 ↗MalfunctionJun 12, 2023Break; Break
IT WAS REPORTED THAT THE BROACH BROKE AFTER THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17108518 ↗MalfunctionJun 12, 2023Break; Break
IT WAS REPORTED THAT THE BROACH BROKE AFTER THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17108373 ↗MalfunctionMar 16, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT 5 ACTIS BROACH WAS CATCHING ON THE BROACH HANDLE. IT DIDN¿T SMOOTHLY GO ON AND OFF THE HANDLE. THE CONNECTION WAS INSPECTED AND CAN¿T IDENTIFY ANYTHING SPECIFICALLY WRONG WITH IT. SO, IT WAS PULLED OUT FROM TRAY. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16555678 ↗MalfunctionJan 18, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT BOTH THE BROACHES WOULD NOT FIT ONTO KINCISE ADAPTORS. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16196905 ↗
Other Instruments designed for press-fit osteochondral implants devices
- ACTIS 3010-01-060Depuy (Ireland) · Instruments designed for press-fit osteochondral implants14 reports
- ACTIS 3010-01-070Depuy (Ireland) · Instruments designed for press-fit osteochondral implants12 reports
- ACTIS 3010-01-040Depuy (Ireland) · Instruments designed for press-fit osteochondral implants12 reports
- ACTIS 3010-01-080Depuy (Ireland) · Instruments designed for press-fit osteochondral implants7 reports
- ACTIS 3010-01-010Depuy (Ireland) · Instruments designed for press-fit osteochondral implants5 reports
- ACTIS 3010-01-020Depuy (Ireland) · Instruments designed for press-fit osteochondral implants4 reports
- ACTIS 3010-01-030Depuy (Ireland) · Instruments designed for press-fit osteochondral implants4 reports
- ACTIS 3010-01-090Depuy (Ireland) · Instruments designed for press-fit osteochondral implants4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page