Skip to content

INHANCE 6507-00-101Recalls & reported problems

Depuy (Ireland) · INHANCE SHOULDER SYSTEM KICKSTAND GUIDE PILOT ROD · Class 1

All models of INHANCE (202)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295549666), received between Nov 9, 2023 and Apr 16, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility4
  • Material Deformation2
  • Crack1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT PILOT ROD GOT STUCK IN SMALL KICKSTAND GUIDE, COULD NOT BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24895628 ↗
  • MalfunctionApr 8, 2026Crack; Device-Device Incompatibility

    IT WAS REPORTED THAT THE INHANCE KICKSTAND LARGE GUIDE AND PILOT ROD STUCK TOGETHER, AND THE LARGE GUIDE HAS A CRACK NEAR WHERE THE PILOT ROD SLIDES INTO. THIS DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24820737 ↗
  • MalfunctionJan 24, 2026Material Deformation

    IT WAS REPORTED THAT DURING A SHOULDER REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE VERSION ROD SNAPPED WHILE TRYING TO CHANGE VERSION, AND THE PILOT ROD WAS CROSS THREADED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24173627 ↗
  • MalfunctionJan 5, 2026Material Deformation

    IT WAS REPORTED THAT DRILL BIT WAS DAMAGED AND FOUND WHILE CROSS THREADING. THIS EVENT HAPPENED IN SURGERY. NO FURTHER INFORMATION IS KNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23960152 ↗
  • MalfunctionFeb 20, 2024Device-Device Incompatibility; Device-Device Incompatibility

    INHANCE KICKSTAND THREADED SIZER WILL NOT THREAD FULLY. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18741394 ↗
  • MalfunctionNov 9, 2023Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT AFTER SURGERY, SALES REP TRIED TO TAKE PILOT ROD OUT OF THE LG KICKSTAND GUIDE HOWEVER THE TWO PIECES WOULD NOT BUDGE AND SEEMED TO BE STRIPPED/STUCK TOGETHER. THIS HAPPENED POST OP SO THERE WAS NO NEGATIVE EFFECT TO THE PATIENT, NOR WAS SURGERY TIME EXTENDED DUE TO THIS. UNFORTUNATELY BECAUSE THE ITEMS DO NOT COME APART, SALES REP WAS UNABLE TO OBTAIN THE LOT NUMBER FROM THE PILOT ROD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18105942 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page