Linear 7.5Fr. 25cc IAB 0684-00-0478-01Recalls & reported problems
Datascope Corp. · Linear 7.5Fr. 25cc IAB · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10607567106526), received between Oct 23, 2020 and Jan 24, 2024 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 23, 2023Class IIOpen, Classified
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Z-2223-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Contamination with Chemical or Other Material3
- Leak/Splash2
- Defective Component1
- Difficult to Insert1
- No Apparent Adverse Event1
- Off-Label Use1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient2
- No Patient Involvement2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 24, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AN INTRA-AORTIC BALLOON (IAB) LEAK OCCURRED DURING THERAPY. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THE IAB HAD BEEN INSERTED ON (B)(6) 2023. THE INSERTION WAS REPORTED TO BE AXILLARY, WHICH IS NOT A METHOD DESCRIBED IN THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18576041 ↗MalfunctionFeb 14, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
MalfunctionDec 19, 2022No Apparent Adverse Event; No Apparent Adverse Event
MalfunctionOct 19, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT 11 DEVICES HAD A VAPOR-LIKE SUBSTANCE INSIDE THE INTRA-AORTIC BALLOON (IAB) INSERTION KIT. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE EVENT REPORTED. THIS REPORT IS FOR #4 OF THE 11 DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15634915 ↗MalfunctionOct 3, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, THE PATIENT WAS GETTING READY FOR AN X-RAY AT THE BED SIDE. ALL OF A SUDDEN, THE GAS EXTENSION OF THE IAB BECAME FILLED WITH BLOOD. THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) GENERATED A BLOOD DETECTED ALARM AND THE IAB WAS REMOVED. THE PATIENT WAS WAITING FOR SURGERY THE NEXT DAY, BUT WAS IN STABLE CONDITION. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15530399 ↗MalfunctionAug 6, 2021Off-Label Use; Defective Component; Off-Label Use; Defective Component
IT WAS REPORTED THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, THE CUSTOMER WAS CHANGING THE FLUSH LINE TUBING PER THEIR PROTOCOL WHEN THE STOPCOCK BROKE. THIS RESULTED IN THE LINE CLOTTING OFF AS THE PROBLEM WAS NOT REMEDIED QUICKLY. THE INSERTION WAS REPORTED TO BE AXILLARY, WHICH IS NOT THE METHOD DESCRIBED IN THE DEVICE INSTRUCTIONS FOR USE. THE CUSTOMER WAS MADE AWARE OF THE OFF-LABEL USE. THE CUSTOMER WAS THEN NOTIFIED OF THE NEED FOR THE ARTERIAL PRESSURE WAVEFORM ON THE CONSOLE AND THE MANY OPTIONS OF HOW TO OBTAIN IT. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12291818 ↗MalfunctionOct 23, 2020Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING INSERTION OF AN INTRA-AORTIC BALLOON (IAB), THE GUIDE WIRE WAS OBSTRUCTED. THE CUSTOMER ALSO NOTED FOREIGN PARTICLES ON THE DEVICE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 10727694 ↗MalfunctionOct 23, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT DURING INSERTION OF AN INTRA-AORTIC BALLOON (IAB), THE GUIDE WIRE WAS OBSTRUCTED. THE CUSTOMER ALSO NOTED FOREIGN PARTICLES ON THE DEVICE. THERE WAS NO PATIENT HARM, OR ADVERSE EVENT REPORTED.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 10727691 ↗
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page