Linear 7.5Fr. 25cc IAB (with APA) 0684-00-0478-02Recalls & reported problems
Datascope Corp. · Linear 7.5Fr. 25cc IAB (with APA) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10607567106533), received between Aug 21, 2024 and Aug 21, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Gas/Air Leak” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 23, 2023Class IIOpen, Classified
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Z-2223-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Gas/Air Leak1
- Inappropriate Waveform1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 21, 2024Inappropriate Waveform; Gas/Air Leak; Material Deformation; Inappropriate Waveform; Gas/Air Leak; Material Deformation
IT WAS REPORTED THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, A "GAS LOSS" ALARM HAD BEEN HAPPENING FOR OVER AN HOUR. THE CUSTOMER COULD RESUME PUMPING BUT WITHIN 1-2 MINUTES THE ALARM WOULD HAPPEN AGAIN. THE INSERTION WAS REPORTED TO BE AXILLARY, WHICH IS NOT THE METHOD DESCRIBED IN THE DEVICE INSTRUCTIONS FOR USE. THEY HAD ORDERED A CHEST FILM FOR PLACEMENT LOCATION. THERE WAS NO BLOOD OR VISIBLE KINK IN THE HELIUM TUBING. THE GETINGE REPRESENTATIVE OFFERED TROUBLESHOOTING SUGGESTIONS AS IF THE IAB WERE PLACED VIA FEMORAL ARTERY. THEY WERE UNSUCCESSFUL IN RESOLVING THE ALARM. THE IAB STATUS BAR DID NOT FILL FULLY, AND THE IAB PRESSURE WAVEFORM LOOKED ROUNDED, INDICATING A LIKELY KINK IN THE CATHETER. THE ICU NURSE PRACTITIONER WAS CONTACTED AND SPOKE WITH THE PHYSICIAN ABOUT FURTHER STEPS. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20022160 ↗
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page