Linear 7.5Fr. 34cc IAB (with APA) 0684-00-0479-02Recalls & reported problems
Datascope Corp. · Linear 7.5Fr. 34cc IAB (with APA) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10607567106540), received between Nov 9, 2021 and Jul 3, 2025 — classified as 1 death, 8 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Insert3
- Leak/Splash2
- Appropriate Device Problem Term/Code Not Available1
- Burst Container or Vessel1
- Inflation Problem1
- No Apparent Adverse Event1
- Optical Problem1
- Restricted Flow rate1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
DeathJul 3, 2025Burst Container or Vessel; Leak/Splash; Optical Problem; Burst Container or Vessel; Leak/Splash; Optical Problem
IT WAS REPORTED FIRST THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, A FAILURE OCCURRED IN THE FIBER OPTIC SENSOR MODULE, RESULTING IN THE BALLOON RUPTURING INSIDE THE PATIENT. ALTHOUGH NO COMPLICATIONS DIRECTLY RELATED TO THE BALLOON RUPTURE WERE REPORTED, THE PATIENT SUBSEQUENTLY DIED.
Patient problems: Insufficient Information
MAUDE report 22396736 ↗MalfunctionDec 27, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING IAB THERAPY, THERE WAS A RESTRICTION IN THE CATHETER. THERE WAS NO HARM OR INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21025314 ↗MalfunctionNov 21, 2024Restricted Flow rate; Restricted Flow rate
IT WAS REPORTED THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, WHEN PERFORMING THE TEST TO START THE PASSAGE OF THE IAB, OCCLUSION OF THE SAME WAS IDENTIFIED. IT WAS NECESSARY TO OPEN A NEW KIT. THERE WAS NO REPORTED INJURY OR ADVERSE EVENT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20753982 ↗MalfunctionMar 14, 2023Inflation Problem; Inflation Problem
MalfunctionFeb 14, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT A PATIENT WITH DECOMPENSATED HEART FAILURE AND A HISTORY OF HEART TRANSPLANTATION IN 2019 WAS IN NEED OF URGENT INTRA-AORTIC BALLOON (IAB) THERAPY. THE IAB WAS INSERTED WITHOUT IMPROVEMENT IN BLOOD PRESSURE. THE CONSOLE GENERATED A GAS LOSS IN IAB CIRCUIT ALARM AND BLOOD WAS SEEN ENTERING THE IAB CATHETER. THE IAB WAS REMOVED AND REPLACED WITH A NEW ONE TO CONTINUE THERAPY WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16371284 ↗MalfunctionOct 4, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING INSERTION OF THE INTRA-AORTIC BALLOON (IAB) THE SHEATH COULD NOT BE INSERTED THROUGH THE GUIDEWIRE BECAUSE THE DILATOR IN THE SHEATH WAS BLOCKED. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15538089 ↗MalfunctionSep 16, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) COULD NOT BE INSERTED THROUGH THE SHEATH. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15440892 ↗MalfunctionDec 16, 2021Difficult to Insert; Unraveled Material; Difficult to Insert; Unraveled Material
IT WAS REPORTED THAT THE CUSTOMER COULD NOT INSERT THE INTRA-AORTIC BALLOON (IAB) DUE TO THE BALLOON MEMBRANE BEING UNFURLED. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13019879 ↗MalfunctionNov 9, 2021No Apparent Adverse Event; No Apparent Adverse Event
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page