Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10607567106540), received between Nov 9, 2021 and Jul 3, 2025 — classified as 1 death, 8 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert3
  • Leak/Splash2
  • Appropriate Device Problem Term/Code Not Available1
  • Burst Container or Vessel1
  • Inflation Problem1
  • No Apparent Adverse Event1
  • Optical Problem1
  • Restricted Flow rate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Insufficient Information1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • DeathJul 3, 2025Burst Container or Vessel; Leak/Splash; Optical Problem; Burst Container or Vessel; Leak/Splash; Optical Problem

    IT WAS REPORTED FIRST THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, A FAILURE OCCURRED IN THE FIBER OPTIC SENSOR MODULE, RESULTING IN THE BALLOON RUPTURING INSIDE THE PATIENT. ALTHOUGH NO COMPLICATIONS DIRECTLY RELATED TO THE BALLOON RUPTURE WERE REPORTED, THE PATIENT SUBSEQUENTLY DIED.

    Patient problems: Insufficient Information

    MAUDE report 22396736 ↗
  • MalfunctionDec 27, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING IAB THERAPY, THERE WAS A RESTRICTION IN THE CATHETER. THERE WAS NO HARM OR INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21025314 ↗
  • MalfunctionNov 21, 2024Restricted Flow rate; Restricted Flow rate

    IT WAS REPORTED THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, WHEN PERFORMING THE TEST TO START THE PASSAGE OF THE IAB, OCCLUSION OF THE SAME WAS IDENTIFIED. IT WAS NECESSARY TO OPEN A NEW KIT. THERE WAS NO REPORTED INJURY OR ADVERSE EVENT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20753982 ↗
  • MalfunctionMar 14, 2023Inflation Problem; Inflation Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16537772 ↗
  • MalfunctionFeb 14, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT A PATIENT WITH DECOMPENSATED HEART FAILURE AND A HISTORY OF HEART TRANSPLANTATION IN 2019 WAS IN NEED OF URGENT INTRA-AORTIC BALLOON (IAB) THERAPY. THE IAB WAS INSERTED WITHOUT IMPROVEMENT IN BLOOD PRESSURE. THE CONSOLE GENERATED A GAS LOSS IN IAB CIRCUIT ALARM AND BLOOD WAS SEEN ENTERING THE IAB CATHETER. THE IAB WAS REMOVED AND REPLACED WITH A NEW ONE TO CONTINUE THERAPY WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16371284 ↗
  • MalfunctionOct 4, 2022Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT DURING INSERTION OF THE INTRA-AORTIC BALLOON (IAB) THE SHEATH COULD NOT BE INSERTED THROUGH THE GUIDEWIRE BECAUSE THE DILATOR IN THE SHEATH WAS BLOCKED. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538089 ↗
  • MalfunctionSep 16, 2022Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) COULD NOT BE INSERTED THROUGH THE SHEATH. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15440892 ↗
  • MalfunctionDec 16, 2021Difficult to Insert; Unraveled Material; Difficult to Insert; Unraveled Material

    IT WAS REPORTED THAT THE CUSTOMER COULD NOT INSERT THE INTRA-AORTIC BALLOON (IAB) DUE TO THE BALLOON MEMBRANE BEING UNFURLED. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13019879 ↗
  • MalfunctionNov 9, 2021No Apparent Adverse Event; No Apparent Adverse Event

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12781665 ↗

Other System, Balloon, Intra-Aortic And Control devices