The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10607567107509), received between Jan 15, 2021 and Mar 31, 2026 — classified as 16 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (3 reports).
8 other recalls cover devices cleared under the same FDA submission (K031636); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image3
- Break2
- Display or Visual Feedback Problem2
- Erratic or Intermittent Display2
- Appropriate Device Problem Term/Code Not Available1
- Battery Problem1
- Connection Problem1
- Device Displays Incorrect Message1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- Insufficient Information5
- No Patient Involvement1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 31, 2026Battery Problem; Gas/Air Leak
IT WAS REPORTED DURING ROUTINE CHECK BY CUSTOMER THE CS100 HAD MESSAGE ABOUT A SECURITY DISK LEAK AND BATTERY MAINTENANCE TO BE PERFORMED. NO PATIENT WAS INVOLVED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24739860 ↗MalfunctionFeb 14, 2026Failure to Read Input Signal
MalfunctionFeb 10, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED BY THE CUSTOMER THAT CS100 INTRA AORTIC BALLOON PUMP (IABP) HAD A POWER CABLE CUT. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21341683 ↗MalfunctionNov 15, 2024Incomplete or Missing Packaging; Appropriate Device Problem Term/Code Not Available; Incomplete or Missing Packaging; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) LOANER ECG CABLE WAS MISSING.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20694069 ↗MalfunctionAug 7, 2024Break; Break
IT WAS REPORTED THAT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD A DAMAGED ECG CABLE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19928382 ↗MalfunctionJul 5, 2024Erratic or Intermittent Display; Connection Problem; Erratic or Intermittent Display; Connection Problem
IT WAS REPORTED AFTER USE THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) SCREEN TURNED WHITE FOLLOWING A SHOCK. THE COMPLAINANT FOLDED THE SCREEN DOWN ONTO THE CABLES, THE SHOCK OCCURRED AT THE THE END OF THE PROCEDURE WHEN IT WAS TIME TO PUT AWAY THE EQUIPMENT. NO PATIENT INVOLVEMENT OR NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19680373 ↗MalfunctionDec 22, 2023Protective Measures Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE BY GETINGE FIELD SERVICE ENGINEER, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT'S BLOOD DETECTION CELL WAS DEFECTIVE AND REQUIRED REPLACEMENT. THERE WAS NO PATIENT INVOLVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18384556 ↗MalfunctionDec 19, 2023Failure to Power Up; Failure to Power Up
MalfunctionOct 30, 2023Erratic or Intermittent Display; Display or Visual Feedback Problem; Erratic or Intermittent Display; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING START UP, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD A SCREEN PROBLEM. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18034432 ↗MalfunctionSep 28, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD SCREEN PROBLEM.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17835427 ↗MalfunctionJan 26, 2023Break; Break
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE FIELD SERVICE ENGINEER (FSE), THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD A BROKEN IV POLE. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16249004 ↗MalfunctionAug 12, 2022No Display/Image; No Display/Image
IT WAS REPORTED THAT, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) UNITS SCREEN IS NOT WORKING PROPERLY. IT IS UNKNOWN UNDER WHICH CIRCUMSTANCES THE EVENT OCCURRED OR IF THERE WAS ANY PATIENT INVOLVEMENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 15218397 ↗MalfunctionAug 4, 2022No Display/Image; No Display/Image
IT WAS REPORTED THAT DURING USE ON A PATIENT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD A DISPLAY FAILURE. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15164511 ↗MalfunctionOct 21, 2021Device Displays Incorrect Message; Device Displays Incorrect Message
MalfunctionMar 30, 2021Display Difficult to Read; Display Difficult to Read
MalfunctionJan 15, 2021No Display/Image
IT WAS REPORTED THAT DURING A SELF-TEST, THE DISPLAY OF THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) WAS UNREADABLE. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 11178624 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K031636. One submission often covers many models, so check the product and lot details in each record.
- Z-1486-2026 · Jan 23, 2026 · Open, ClassifiedCS100 IABP. Software Version CS100 IABP Q.01.
- Z-3085-2017 · Jul 17, 2017 · Open, ClassifiedCS 100i Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by a
- Z-3086-2017 · Jul 17, 2017 · Open, ClassifiedCS 100 Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by au
- Z-2737-2017 · Jun 16, 2017 · TerminatedCS 100 Intra-Aortic Balloon Pump
- Z-2736-2017 · Jun 16, 2017 · TerminatedCS 100i Intra-Aortic Balloon Pump
- Z-1523-2014 · Mar 21, 2014 · TerminatedDatascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate in
- Z-2414-2011 · Mar 16, 2011 · TerminatedDatascope/ Maquet IABPs. System 98/98XT Intra-Aortic Balloon Pump; CS100/CS100i Intra-Aortic Balloon Pump; CS300 Intra-Aortic Balloon Pump; Manufactured by Datascope Corp., 130
- Z-1811-2010 · May 10, 2010 · TerminatedCS100 Intra-Aortic Balloon Pump.
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.