Skip to content

CS100 Intra-Aortic Balloon Pump, Refurbished, ENGLISH,NON-UTS,INTL 0998-UC-3013-55Recalls & reported problems

Datascope Corp. · CS100 Intra-Aortic Balloon Pump, Refurbished, ENGLISH,NON-UTS,INTL · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Datascope CS100 intra-aortic balloon pump is a refurbished heart support machine that inflates a balloon in the aorta. Of its 128 reports, 126 are malfunctions, 1 is an injury and 1 describes a death; some describe the pump failing to power up, failing to auto-fill, or showing erratic displays. Reports fell to 13 in the last 12 months from 31 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
129
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 129 adverse event reports that list this exact device (DI 10607567107875), received between Oct 15, 2020 and Aug 31, 2026 — classified as 1 death, 1 injury, 127 malfunctions.

The most frequently reported specific device problem is “Inability to Auto-Fill” (16 reports).

8 other recalls cover devices cleared under the same FDA submission (K031636); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 4 reportsMar 2024: 4 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 8 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 4 reportsOct 2024: 2 reportsNov 2024: 5 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 3 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 6 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 4 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 4 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inability to Auto-Fill16
  • Appropriate Device Problem Term/Code Not Available14
  • Failure to Power Up13
  • Electrical /Electronic Property Problem10
  • Display or Visual Feedback Problem9
  • Battery Problem8
  • Erratic or Intermittent Display8
  • Break5

Patient problems

  • No Clinical Signs, Symptoms or Conditions124
  • Insufficient Information8
  • No Patient Involvement3
  • No Known Impact Or Consequence To Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Headache1
  • Nausea1
  • Pain1

Recent adverse event reports

The 25 most recent of 129 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2026Infusion or Flow Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50515668 ↗
  • MalfunctionJul 27, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50017358 ↗
  • MalfunctionMar 25, 2026Battery Problem

    IT WAS REPORTED THAT CS100 INTRA-AORTIC BALLOON PUMP (IABP)¿S DAMAGED BATTERIES WERE REPLACED DURING A TECHNICAL INSPECTION. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24685673 ↗
  • MalfunctionMar 12, 2026Failure to Power Up

    IT WAS REPORTED DURING A ROUTINE CHECK, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) WAS NOT SWITCHING ON MAINS. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24579545 ↗
  • MalfunctionMar 6, 2026Appropriate Device Problem Term/Code Not Available; Unexpected Shutdown

    IT WAS REPORTED BY CUSTOMER DURING ROUTINE CHECK THAT CS100 INTRA AORTIC BALLOON PUMP (IABP) HAD SYSTEM FAILURE ALARM AND PRESSURE PORT WAS NOT WORKING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24531770 ↗
  • MalfunctionMar 2, 2026Electrical /Electronic Property Problem; Calibration Problem

    IT WAS REPORTED BY CUSTOMER DURING ROUTINE CHECK THAT THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD MAINTENANCE REQUIRED CODE #5. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24479037 ↗
  • MalfunctionFeb 26, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING ROUTINE CHECK THE AUGMENTATION OF CS100 INTRA-AORTIC BALLOON PUMP(IABP) MALFUNCTIONED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24453110 ↗
  • MalfunctionFeb 24, 2026Calibration Problem; Appropriate Device Problem Term/Code Not Available

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24433592 ↗
  • MalfunctionFeb 20, 2026Inability to Auto-Fill

    IT WAS REPORTED THAT DURING A ROUTINE CHECK THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD AN AUTO FAILURE ALARM. NO PATIENT INVOLVEMENT OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24398829 ↗
  • MalfunctionDec 29, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT BEFORE USE BY CUSTOMER, CS100 INTRA-AORTIC BALLOON PUMP (IABP) NOT SWITCHING ON. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23905573 ↗
  • MalfunctionDec 14, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THAT BEFORE USE, CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD ELECTRICAL TEST FAILURE #39. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23799488 ↗
  • MalfunctionDec 11, 2025Decrease in Pressure; Suction Problem; Output below Specifications; Decrease in Pressure; Suction Problem; Output below Specifications

    IT WAS REPORTED BY THE CUSTOMER THAT THE CS100 INTRA-AORTIC BALLOON PUMP(IABP) WHEN TURNING ON THE DEVICE, IT FIRST STARTED REPORTING LOW VACUUM AND THEN ELECTRICAL ERROR CODE #50. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23777297 ↗
  • MalfunctionDec 4, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23717692 ↗
  • MalfunctionNov 21, 2025Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT DURING ROUTINE CHECK, CS100 INTRA-AORTIC BALLOON PUMP (IABP) NOT SWITCHING ON MAINS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23614687 ↗
  • MalfunctionOct 16, 2025Incorrect, Inadequate or Imprecise Result or Readings; Appropriate Device Problem Term/Code Not Available; Incorrect, Inadequate or Imprecise Result or Readings; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING ROUTINE CHECK, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) FAILED SYSTEM TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310945 ↗
  • MalfunctionSep 16, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT DURING ROUTINE CHECK, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD A CS100 MAINTENANCE REQUIRED CODE#5. THERE WAS NO PATIENT INVOLVED AND NO HARM OR INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23060767 ↗
  • MalfunctionSep 3, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22951247 ↗
  • MalfunctionSep 2, 2025Inability to Auto-Fill; Inability to Auto-Fill

    IT WAS REPORTED THAT THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD AUTO-FILL FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22932212 ↗
  • MalfunctionAug 26, 2025Inability to Auto-Fill; Gas/Air Leak; Inability to Auto-Fill; Gas/Air Leak

    IT WAS REPORTED THAT DURING PLANNED INSPECTION BY FSE, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD ODOMETER READING 4645 H PUMP INOPERATIVE. IT DID NOT REACH THE SET VACUUM VALUES. THERE WERE FEW TECHNICAL ERRORS GENERATED. ERROR #57, #62. THE PUMP FAILS THE FOLLOWING TESTS: AUTOFILL TEST, K6, K6A, K7, K8 LEAK TEST AND PNEUMATIC PERFORMANCE TEST. THERE WAS NO PATIENT INVOLVED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22899071 ↗
  • MalfunctionAug 14, 2025Failure to Read Input Signal; Failure to Read Input Signal

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22801889 ↗
  • MalfunctionAug 12, 2025Failure to Read Input Signal; No Pressure; Failure to Read Input Signal; No Pressure

    IT WAS REPORTED THAT CS100 INTRA AORTIC BALLOON PUMP ( IABP) SHOWED PRESSURE AND ECG PROBLEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22781721 ↗
  • MalfunctionAug 8, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22746798 ↗
  • MalfunctionAug 1, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22685395 ↗
  • MalfunctionAug 1, 2025Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available; Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY CUSTOMER, DURING ROUTINE CHECK CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD ELECTRICAL TEST FAILS CODE #50. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22684412 ↗
  • MalfunctionJul 29, 2025Gas/Air Leak; Gas/Air Leak

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22643013 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K031636. One submission often covers many models, so check the product and lot details in each record.

  • Z-1486-2026 · Jan 23, 2026 · Open, Classified
    CS100 IABP. Software Version CS100 IABP Q.01.
  • Z-3085-2017 · Jul 17, 2017 · Open, Classified
    CS 100i Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by a
  • Z-3086-2017 · Jul 17, 2017 · Open, Classified
    CS 100 Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by au
  • Z-2736-2017 · Jun 16, 2017 · Terminated
    CS 100i Intra-Aortic Balloon Pump
  • Z-2737-2017 · Jun 16, 2017 · Terminated
    CS 100 Intra-Aortic Balloon Pump
  • Z-1523-2014 · Mar 21, 2014 · Terminated
    Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate in
  • Z-2414-2011 · Mar 16, 2011 · Terminated
    Datascope/ Maquet IABPs. System 98/98XT Intra-Aortic Balloon Pump; CS100/CS100i Intra-Aortic Balloon Pump; CS300 Intra-Aortic Balloon Pump; Manufactured by Datascope Corp., 130
  • Z-1811-2010 · May 10, 2010 · Terminated
    CS100 Intra-Aortic Balloon Pump.

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →