This Datascope CS300 intra-aortic balloon pump is a machine that inflates and deflates a balloon in the aorta to help the heart pump during surgery or critical care. The 45 reports mostly describe display and screen problems, the monitor not working or showing only lines, and low vacuum or error codes found during testing or maintenance. Two deaths and no injuries are reported. There are no recalls naming this device, and reports have dropped from 10 in the earlier12 months to 3 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
45
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 10607567108278), received between Sep 30, 2020 and Mar 30, 2026 — classified as 2 deaths, 43 malfunctions.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available7
  • Display or Visual Feedback Problem6
  • Failure to Sense4
  • Electrical /Electronic Property Problem3
  • Erratic or Intermittent Display3
  • No Display/Image3
  • Battery Problem2
  • Decrease in Pressure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions39
  • Insufficient Information8
  • No Known Impact Or Consequence To Patient3
  • Aneurysm2
  • No Patient Involvement2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Decrease in Pressure

    IT WAS REPORTED THAT DURING MAINTENANCE TESTING CS300 INTRA-AORTIC BALLOON PUMP (IABP), IT WAS FOUND THAT THE VACUUM WAS BELOW MINIMUM VALUES. AFTER SEVERAL TESTS, IT WAS CONCLUDED THAT THE PROBLEM WAS WITH THE COMPRESSOR ITSELF. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24710847 ↗
  • MalfunctionFeb 25, 2026Electrical /Electronic Property Problem; Noise, Audible

    IT WAS REPORTED THAT BEFORE USE, THE CS300 INTRA AORTIC BALLOON PUMP (IABP) DISPLAYED AN ELECTRICAL TEST FAILURE CODE # 50 (MOTOR SPEED OUT OF SPECIFICATION). NO PATIENT WAS INVOLVED, AND NO HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24444409 ↗
  • MalfunctionDec 8, 2025No Display/Image; Display or Visual Feedback Problem; No Display/Image; Display or Visual Feedback Problem

    IT WAS REPORTED BY THE CUSTOMER THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) MONITOR WAS NOT WORKING AND ONLY LINES WERE VISIBLE ON THE SCREEN. THE ISSUE OCCURRED DURING INITIAL TEST BEFORE USE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23736604 ↗
  • MalfunctionSep 2, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22931769 ↗
  • MalfunctionJul 3, 2025Electrical /Electronic Property Problem; Decrease in Pressure; Infusion or Flow Problem; Electrical /Electronic Property Problem; Decrease in Pressure; Infusion or Flow Problem

    IT WAS REPORTED BY CUSTOMER THAT DURING USE THE CS300 INTRA AORTIC BALLOON PUMP HAD DISPLAYED LOW VACUUM ALARM FOLLOWED BY ELECTRICAL CODE ERROR 50. THERE WAS NO INJURY OR THREAT TO THE PATIENT OR OTHER PERSON. USE ANOTHER DEVICE CS100 PUMP TO CONTINUE THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22393942 ↗
  • MalfunctionJun 20, 2025Fluid/Blood Leak; Device Contamination with Body Fluid; Gas/Air Leak; Fluid/Blood Leak; Device Contamination with Body Fluid; Gas/Air Leak

    IT WAS REPORTED BY GETINGE FIELD SAFETY ENGINEER DURING MAINTENANCE THE CS300 INTRA AORTIC BALLOON PUMP (IABP) CUSTOMER DISCOVERED THAT THERE WAS BLOOD IN THE SYSTEM AND LEAK IN THE IABP CATHETER. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE EVENT REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22295045 ↗
  • MalfunctionJun 11, 2025Inability to Auto-Fill; Inability to Auto-Fill

    IT WAS REPORTED THAT CS300 INTRA-AORTIC BALLOON PUMP (IABP) FAILED DURING AUTOFILL AND EXPERIENCED AN AUTOFILL FAILURE ALARM DURING USE ON PATIENT. THERE WAS NO INJURY OR HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22182280 ↗
  • MalfunctionApr 22, 2025Incorrect, Inadequate or Imprecise Result or Readings; Appropriate Device Problem Term/Code Not Available; Unexpected Shutdown; Incorrect, Inadequate or Imprecise Result or Readings; Appropriate Device Problem Term/Code Not Available; Unexpected Shutdown

    IT WAS REPORTED BEFORE USE THAT A CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD UNSPECIFIED MALFUNCTION. NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21882490 ↗
  • MalfunctionMar 17, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT DURING USE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD AUTOLOAD SYSTEM FAILURE (FAULT 57) . THERE WAS NO PATIENT HARM OR INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21617630 ↗
  • MalfunctionMar 17, 2025Optical Problem; Optical Problem

    IT WAS REPORTED BY THE CUSTOMER THAT DURING DAILY INSPECTION, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD AN OPTICAL SENSOR FAILURE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21616090 ↗
  • MalfunctionFeb 24, 2025Power Problem; Power Problem

    IT WAS REPORTED THAT DURING USE ON PATIENT, CS300 INTRA-AORTIC BALLOON PUMP (IABP) DISPLAYED THE MESSAGE "MAINTENANCE REQUIRED CODE #7". THERE WAS NO INJURY OR HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21444269 ↗
  • MalfunctionFeb 3, 2025Calibration Problem; Calibration Problem

    N/A

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21292379 ↗
  • MalfunctionDec 2, 2024Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20824285 ↗
  • MalfunctionOct 22, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT WHEN TURNING ON THE EQUIPMENT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAS ELECTRICAL FAULT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20505911 ↗
  • MalfunctionSep 13, 2024Decrease in Suction; Decrease in Suction

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20212920 ↗
  • MalfunctionJul 31, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT DURING DAILY USE, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) DISPLAY IS DEFECTIVE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19878182 ↗
  • MalfunctionMay 20, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    N/A

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19355112 ↗
  • MalfunctionMay 20, 2024Failure to Sense; Failure to Sense

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19351496 ↗
  • MalfunctionApr 30, 2024Device Alarm System; Appropriate Device Problem Term/Code Not Available; Device Alarm System; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING ROUTINE CHECK, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) ALARM WENT OFF. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19216431 ↗
  • MalfunctionApr 29, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM) , THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) FIBER OPTIC COVER WAS MISSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19209942 ↗
  • MalfunctionFeb 19, 2024Break; Break

    IT WAS REPORTED THAT WHILE MOVING THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) DEVICE FROM CARDIOLOGY STORAGE UNIT TO THE OPERATING ROOM, THE HOLDING ARM BROKE OFF AND NEEDS TO BE REPLACED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18730630 ↗
  • MalfunctionNov 17, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT DURING A ROUTINE CHECL AFTER SWITCHING ON THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) FLAT SCREEN MONITOR CONSTANTLY FLICKERS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18163076 ↗
  • MalfunctionAug 15, 2023Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT'S DISPLAY WAS DAMAGED. THE DISPLAY FLICKERED AS SOON AS IT WAS SWITCHED ON. A DIFFERENT DEVICE/THERAPY WAS THEN USED FOR THE TREATMENT. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17547916 ↗
  • MalfunctionMay 4, 2023Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM) THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) THE DEVICE FAILED THE BATTERY TEST, THE BATTERIES DID NOT REACH THE REQUIRED MINIMUM RUNTIME AND THE DEVICE SWITCHED OFF AFTER AROUND 98 MINUTES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16868754 ↗
  • MalfunctionMar 24, 2023Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM) , THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD A BROKEN LCD SCREEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16610979 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →