The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10607567108292), received between Mar 17, 2021 and Nov 19, 2024 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Body Fluid” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Body Fluid3
  • Unexpected Shutdown2
  • Device Alarm System1
  • Display Difficult to Read1
  • Display or Visual Feedback Problem1
  • Failure to Run on Battery1
  • Inadequate User Interface1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Insufficient Information6

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2024Insufficient Device Problem Information; Noise, Audible; Insufficient Device Problem Information; Noise, Audible

    IT WAS REPORTED THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT SCREEN CRACKLING DURING TURNING ON. NO PATIENT DAMAGE WAS REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20724489 ↗
  • MalfunctionMay 8, 2024Device Contamination with Body Fluid; Device Contamination with Body Fluid

    N/A

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19271020 ↗
  • MalfunctionApr 3, 2023Failure to Run on Battery; Failure to Run on Battery

    IT WAS REPORTED THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAS A BATTERY MALFUNCTION THE UNIT SWITCHES OFF WHEN UNPLUGGED FROM THE MAINS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16667625 ↗
  • MalfunctionMar 28, 2023Inadequate User Interface; Inadequate User Interface

    N/A.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16626623 ↗
  • MalfunctionDec 28, 2022Device Contamination with Body Fluid; Device Contamination with Body Fluid

    IT WAS REPORTED THAT A BALLOON BROKE DURING USE ON A PATIENT CAUSING BLOOD TO GO INSIDE THE CS300 INTRA-AORTIC BALLOON PUMP (IABP).

    Patient problems: Insufficient Information

    MAUDE report 16062274 ↗
  • MalfunctionApr 29, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT DURING A REPAIR EVALUATION FOR A PRIOR REPAIR CALL BY A GETINGE FIELD SERVICE ENGINEER (FSE), THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD A LEAK THAT WAS NOTICED ON THE DRIVE MANIFOLD. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14254358 ↗
  • MalfunctionFeb 15, 2022Device Contamination with Body Fluid; Device Contamination with Body Fluid

    IT WAS REPORTED THAT DURING USE ON A PATIENT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD BLOOD BACKED ALL THE WAY THROUGH THE TUBING AND POTENTIALLY INTO THE MACHINE. THERE WAS NO PATIENT HARM AND NO ADVERSE EVENT WAS REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13536784 ↗
  • MalfunctionFeb 8, 2022Device Alarm System; Device Alarm System

    IT WAS REPORTED THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) GAVE AN ALARM AND A MESSAGE. IT IS UNKNOWN THE CIRCUMSTANCES UNDER WHICH THE EVENT OCCURRED. IT IS ALSO UNKNOWN IF THERE WAS A PATIENT INVOLVEMENT. HOWEVER, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13481363 ↗
  • MalfunctionJul 15, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12172294 ↗
  • MalfunctionJun 30, 2021Unexpected Shutdown; Unexpected Shutdown

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12092776 ↗
  • MalfunctionJun 21, 2021Unexpected Shutdown; Unexpected Shutdown

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12032271 ↗
  • MalfunctionApr 26, 2021Display Difficult to Read; Display Difficult to Read

    IT WAS REPORTED THAT DURING A ROUTINE CHECK, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD BLURRY LINES ON THE DISPLAY SCREEN. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11726737 ↗
  • MalfunctionMar 17, 2021Power Problem; Power Problem

    IT WAS REPORTED BY THE CUSTOMER'S BIOMED THAT DURING USE ON A PATIENT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) WAS ON BATTERIES AND WHEN SWITCHED OFF, THERE WAS NO AC POWER. SUBSEQUENTLY, THE USER SWITCHED OFF AND ON THE AC POWER AND AFTER THAT THE MACHINE SWITCHED ON DIRECTLY. THERE WAS NO HARM OR INJURY TO PATIENT AND NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11506907 ↗