Skip to content

CS300 Intra-Aortic Balloon Pump, Spanish, 220V 0998-00-3023-58Recalls & reported problems

Datascope Corp. · CS300 Intra-Aortic Balloon Pump, Spanish, 220V · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Datascope CS300 intra-aortic balloon pump is a bedside machine that inflates a balloon in the aorta to help the heart pump blood during cardiac care. The 64 reports mostly describe equipment problems found during servicing — screens or control buttons not working, batteries needing replacement, gas or air leaks, and units shutting off on their own — with no patient involved and no deaths or injuries reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
66
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 66 adverse event reports that list this exact device (DI 10607567108339), received between Jan 21, 2021 and Sep 28, 2026 — classified as 66 malfunctions.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available17
  • Display or Visual Feedback Problem7
  • Gas/Air Leak6
  • Battery Problem5
  • Break5
  • Inability to Auto-Fill5
  • Failure to Power Up3
  • Unexpected Shutdown3

Patient problems

  • No Clinical Signs, Symptoms or Conditions58
  • Insufficient Information20
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 66 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Failure to Read Input Signal

    It was reported by the customer that during a routine check the CS300 Intra-Aortic Balloon Pump (IABP) malfunctioned. Optical display not working has been reported. The customer stated the optical sensor was not functional. No patient was involved. No harm was reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE tested the fiber optic function and it worked as expected, the issue with the pump could not be reproduced. The values are correct and pass the test. Equipment is suitable for clinical use. No parts to be replaced or returned. The unit passed all functional and safety tests to manufacturer specifications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50912777 ↗
  • MalfunctionSep 24, 2026Battery Problem

    It was reported by the customer that during routine check, the CS300 Intra-Aortic Balloon Pump (IABP) the batteries need to be replaced because they do not last the minimum required time. There was no patient involvement and no harm was reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE reported that batteries replaced and proper operation verified for clinical use. Product passed all functional and safety tests per manufacturer specifications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50877703 ↗
  • MalfunctionJan 5, 2026Display or Visual Feedback Problem

    IT WAS REPORTED BY THE CUSTOMER THAT DURING PM THE CS300 INTRA-AORTIC BALLOON PUMP(IABP) SCREEN NOT WORKING. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23954944 ↗
  • MalfunctionDec 11, 2025No Display/Image; No Display/Image

    IT WAS REPORTED BY THE CUSTOMER THAT CS300 INTRA-AORTIC BALLOON PUMP, SPANISH, 220V INTRA-AORTIC BALLOON PUMP (IABP) HAD A SCREEN FAILURE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23780598 ↗
  • MalfunctionDec 11, 2025No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23778108 ↗
  • MalfunctionJul 31, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    IT WAS REPORTED BY THE FIELD SERVICE ENGINEER (FSE) THAT CS300 INTRA-AORTIC BALLOON PUMP (IABP) BUTTONS DIDN'T WORK. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22673961 ↗
  • MalfunctionJul 31, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT BATTRIES OF CS300 INTRA AORTIC BALLOON PUMP (IABP) NEED TO CHANGE. THERE WAS NO HARM OR INJURY REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22672991 ↗
  • MalfunctionJul 21, 2025Display or Visual Feedback Problem; Image Display Error/Artifact; Display or Visual Feedback Problem; Image Display Error/Artifact

    IT WAS REPORTED BY CUSTOMER THAT THE DISPLAY OF CS300 INTRA AORTIC BALLOON PUMP ( IABP) WAS SHOWING HORIZONTAL LINES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22553174 ↗
  • MalfunctionJun 4, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) SHUTS OFF BY ITSELF.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22140872 ↗
  • MalfunctionApr 4, 2025Electrical /Electronic Property Problem; Gas/Air Leak; Appropriate Device Problem Term/Code Not Available; Electrical /Electronic Property Problem; Gas/Air Leak; Appropriate Device Problem Term/Code Not Available

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21771499 ↗
  • MalfunctionMar 24, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD GAS LOSS ALARM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21675550 ↗
  • MalfunctionMar 17, 2025Display or Visual Feedback Problem; Appropriate Device Problem Term/Code Not Available; Display or Visual Feedback Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE CUSTOMER THAT DURING USE ON THE PATIENT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD ALARM ON THE DISPLAY. THERE WAS NO HARM OR INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21616670 ↗
  • MalfunctionMar 12, 2025Gas Output Problem; Gas Output Problem

    IT WAS REPORTED BY CUSTOMER DURING USE THAT CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) COUNTERPULSATION BALLOON CONSOLE GIVES A LOW VACUUM ERROR. NO PATIENT HARM OR INJURY REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21581610 ↗
  • MalfunctionJan 21, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT, DURING PREVENTIVE MAINTENANCE THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD KEYPAD USER INTERFACE MALFUNCTION. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21195915 ↗
  • MalfunctionJan 10, 2025Break; Break

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21122592 ↗
  • MalfunctionJan 7, 2025Pressure Problem; Pressure Problem

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 21090196 ↗
  • MalfunctionNov 19, 2024Battery Problem; Battery Problem

    IT WAS REPORTED DURING A ROUTINE CHECK ,THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) BATTERIES ARE FAILING .THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20727235 ↗
  • MalfunctionNov 1, 2024Break; Appropriate Device Problem Term/Code Not Available; Break; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) WHEEL BROKE, THE RUBBER FAILED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20588048 ↗
  • MalfunctionOct 22, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT BEFORE USE, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAS HELIUM LEAK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20505822 ↗
  • MalfunctionSep 6, 2024Battery Problem; Battery Problem

    IT WAS REPORTED DURING THE STARTUP ROUTINE, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) BATTERY NEEDED TO BE CHANGED. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20162514 ↗
  • MalfunctionSep 4, 2024Decrease in Suction; Decrease in Suction

    IT WAS REPORTED THAT BEFORE THERAPY DURING THE START-UP ROUTINE, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAS LOW VACUUM ALARM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20137675 ↗
  • MalfunctionAug 7, 2024Break; Break

    IT WAS REPORTED THAT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) ENTRANCE FOR CONNECTING THE COUNTERPULSATION BALLOON CATHETER IS BROKEN.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19928384 ↗
  • MalfunctionJul 5, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING A ROUTINE CHECK, PRIOR TO USE, THE SCREEN ON THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT IS BROKEN. THERE WAS NO PATIENT INVOLVEMENT .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19679768 ↗
  • MalfunctionMay 14, 2024False Alarm; False Alarm

    IT WAS REPORTED THAT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT GIVES FALSE BALLOON CHECK ALARM TO CHECK THE BALL.

    Patient problems: Insufficient Information

    MAUDE report 19310696 ↗
  • MalfunctionApr 22, 2024Break; Break

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19152280 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →

Sources and updates