Skip to content

CS300 Intra-Aortic Balloon Pump, Chinese, 220V 0998-00-3023-45Recalls & reported problems

Datascope Corp. · CS300 Intra-Aortic Balloon Pump, Chinese, 220V · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Datascope CS300 intra-aortic balloon pump has 233 adverse event reports, including 2 injuries and 1 death. The reports mostly describe the pump failing to fill its balloon automatically, air or gas leaks, and error or power messages on startup. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
235
Deaths reported
1
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 235 adverse event reports that list this exact device (DI 10607567108575), received between Oct 12, 2020 and Sep 11, 2026 — classified as 1 death, 2 injuries, 232 malfunctions.

The most frequently reported specific device problem is “Inability to Auto-Fill” (55 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 8 reportsDec 2023: 6 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 4 reportsMay 2024: 10 reportsJun 2024: 6 reportsJul 2024: 8 reportsAug 2024: 6 reportsSep 2024: 7 reportsOct 2024: 12 reportsNov 2024: 16 reportsDec 2024: 1 reportsJan 2025: 4 reportsFeb 2025: 2 reportsMar 2025: 15 reportsApr 2025: 0 reportsMay 2025: 5 reportsJun 2025: 2 reportsJul 2025: 6 reportsAug 2025: 4 reportsSep 2025: 5 reportsOct 2025: 3 reportsNov 2025: 11 reportsDec 2025: 6 reportsJan 2026: 6 reportsFeb 2026: 10 reportsMar 2026: 4 reportsApr 2026: 5 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inability to Auto-Fill55
  • Gas/Air Leak33
  • Appropriate Device Problem Term/Code Not Available32
  • Leak/Splash24
  • Inflation Problem13
  • Failure to Power Up11
  • Battery Problem10
  • Electrical /Electronic Property Problem9

Patient problems

  • No Clinical Signs, Symptoms or Conditions223
  • Insufficient Information9
  • No Patient Involvement4
  • No Consequences Or Impact To Patient3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Cardiac Arrest1
  • Cardiogenic Shock1
  • Hemodynamic instability1

Recent adverse event reports

The 25 most recent of 235 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2026Gas/Air Leak

    It was reported by the customer that prior to use, the CS300 Intra-Aortic Balloon Pump (IABP) experienced a Leak in IAB Circuit Alarm. There was no patient involvement. No harm is reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE reported that the safety disk was faulty. After replacement, it passed pre-use inspection. All machine functions and factory-set safety performance metrics were verified, and the unit was approved for clinical use following price negotiation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50727840 ↗
  • MalfunctionSep 2, 2026Inability to Auto-Fill

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50566025 ↗
  • MalfunctionAug 11, 2026Inability to Auto-Fill

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50208131 ↗
  • MalfunctionAug 6, 2026Inability to Auto-Fill; Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50156581 ↗
  • MalfunctionMay 11, 2026Inability to Auto-Fill; Battery Problem

    IT WAS REPORTED BY THE CUSTOMER THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) DISPLAYED A BATTERY NEEDS MAINTENANCE MESSAGE AFTER POWER-ON. THE CUSTOMER ALSO REPORTED AN AUTOMATIC INFLATION FAILED MESSAGE. NO PATIENT INVOLVEMENT OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25145322 ↗
  • MalfunctionApr 29, 2026Gas/Air Leak

    IT WAS REPORTED BY THE CUSTOMER THAT DURING USE, THE CS300 INTRA AORTIC BALLOON PUMP (IABP) HAD AIR LEAK IN THE IAB LOOP. THERE WAS NO PATIENT HARM AND INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25025615 ↗
  • MalfunctionApr 21, 2026Unexpected Shutdown

    IT WAS REPORTED BY THE CUSTOMER THAT DURING ROUTINE INSPECTION THE CS300, CHINESE, NON-UTS INTRA-AORTIC BALLOON PUMP (IABP)DEVICE DISPLAYED A SYSTEM FAILURE ERROR MESSAGE AFTER POWERING ON, ACCOMPANIED BY A BUZZER ALARM. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24939351 ↗
  • MalfunctionApr 20, 2026Inability to Auto-Fill

    IT WAS REPORTED THAT DURING USE THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD AN AUTOMATIC INFLATION MALFUNCTION. PATIENT INVOLVEMENT WAS REPORTED, HOWEVER NO PATIENT HARM OR INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24922939 ↗
  • MalfunctionApr 10, 2026Leak/Splash; Battery Problem

    IT WAS REPORTED BY THE CUSTOMER THAT DURING USE, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD A CIRCUIT LEAK ALARM. THE BALLOON WAS CHECKED AND FOUND TO HAVE NO ABNORMALITIES. PATIENT INVOLVEMENT WAS REPORTED, HOWEVER NO PATIENT HARM OR INJURY OCCURRED, AS THE PATIENT WAS TRANSFERRED TO ANOTHER DEVICE TO CONTINUE THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24838976 ↗
  • MalfunctionApr 7, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE CUSTOMER THAT THE CS300 INTRA-AORTIC BALLOON PUMP(IABP) DEVICE OCCASIONALLY DISPLAYS A MESSAGE MAINTENANCE REQUIRED, CODE 7, WHILE OTHER FUNCTIONS OF THE DEVICE ARE NORMAL. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24796612 ↗
  • MalfunctionMar 30, 2026Restricted Flow rate; Appropriate Device Problem Term/Code Not Available

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24722638 ↗
  • MalfunctionMar 30, 2026Inability to Auto-Fill; Gas/Air Leak

    IT WAS REPORTED THAT DURING USE, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) EXPERIENCED AN ALARM LOOP LEAK INDICATING AUTOMATIC INFLATION FAILURE. . THERE WAS PATIENT INVOLVEMENT BUT NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24711652 ↗
  • MalfunctionMar 6, 2026Gas/Air Leak

    IT WAS REPORTED BY CUSTOMER THAT DURING ROUTINE MAINTENANCE, THE CS300 INTRA AORTIC BALLOON PUMP HAD A LEAK IN IAB CIRCUIT ALARM DUE TO A LEAKING SAFETY DISC. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24528553 ↗
  • MalfunctionMar 2, 2026Gas/Air Leak

    IT WAS REPORTED BY THE CUSTOMER THAT DURING MAINTENANCE AN ERROR WAS FOUND AFTER POWERING ON THE CS300 INTRA-AORTIC BALLOON PUMP(IABP) MACHINE: LOOP LEAK. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24479929 ↗
  • MalfunctionFeb 18, 2026Inability to Auto-Fill

    IT WAS REPORTED BY THE CUSTOMER THAT DURING PRE-USE SELF TEST THE CS300 INTRA-AORTIC BALLOON PUMP(IABP) EQUIPMENT ALARMED AND FAILED TO AUTOMATICALLY INFLATE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24376086 ↗
  • MalfunctionFeb 18, 2026Inability to Auto-Fill

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24375520 ↗
  • MalfunctionFeb 17, 2026Gas/Air Leak

    IT WAS REPORTED BY THE CUSTOMER THAT DURING A ROUTINE INSPECTION, AN AIR LEAK WAS FOUND IN THE ALARM LOOP OF THE CS300 INTRA-AORTIC BALLOON PUMP(IABP) DEVICE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24369256 ↗
  • MalfunctionFeb 16, 2026Decrease in Pressure

    IT WAS REPORTED BY CUSTOMER THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAD ALARMED DUE TO LOW NEGATIVE PRESSURE DURING USE. THERE WAS NO PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24342048 ↗
  • MalfunctionFeb 12, 2026Moisture or Humidity Problem; Noise, Audible

    IT WAS REPORTED DURING PRE-USE CHECK INSPECTION THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD A CONDENSER MALFUNCTION. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320261 ↗
  • MalfunctionFeb 11, 2026Inability to Auto-Fill; Gas/Air Leak

    IT WAS REPORTED THAT DURING ROUTINE MAINTENANCE THE CS300 IABP HAD AN AUTOMATIC INFLATION MALFUNCTION AND THE DRIVE VALVE ASSEMBLY TEST FAILED. NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24311781 ↗
  • MalfunctionFeb 5, 2026Battery Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24265556 ↗
  • MalfunctionFeb 3, 2026Inability to Auto-Fill; Gas/Air Leak

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24244420 ↗
  • MalfunctionFeb 3, 2026Inability to Auto-Fill

    IT WAS REPORTED BY THE CUSTOMER THAT THE DEVICE CS300 INTRA-AORTIC BALLOON PUMP (IABP) DISPLAYED AN ERROR MESSAGE THAT AUTOMATIC INFLATION FAILED DURING TESTING, RENDERING THE DEVICE UNUSABLE. THERE WERE NO PATIENTS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24243856 ↗
  • MalfunctionFeb 3, 2026Leak/Splash; Pressure Problem

    IT WAS REPORTED BY THE CUSTOMER THAT DURING ROUTINE INSPECTION THE CS300 INTRA-AORTIC BALLOON PUMP(IABP) SYSTEM'S SELF-TEST REVEALED AN EXCESSIVELY LOW POSITIVE MOTOR PRESSURE, CAUSING THE VALVE ASSEMBLY LEAKAGE TEST TO FAIL. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24243592 ↗
  • MalfunctionJan 30, 2026Inability to Auto-Fill

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24219465 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page