Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10607567108711), received between Feb 25, 2021 and Aug 26, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Battery Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Battery Problem2
- Unexpected Shutdown2
- Display or Visual Feedback Problem1
- Failure to Power Up1
- Failure to Sense1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information1
- No Patient Involvement1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 26, 2026Failure to Power Up
It was reported by the customer that during routine check, the CS300 Intra-Aortic Balloon Pump (IABP) is not turning on. No patient involved. No harm was reported. FSE Replaced power supply which resolved issue. FSE completed repair successfully. Product passed all functional and safety tests per manufacturer specifications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50446874 ↗MalfunctionAug 8, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING A ROUTINE CHECK, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) EXHIBITED SCREEN FLICKERING, WITH FLASHING LINES VISIBLE ON THE DISPLAY. NO PATIENT WAS INVOLVED OR HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22747257 ↗MalfunctionFeb 5, 2025Unexpected Shutdown; Unexpected Shutdown
MalfunctionJan 8, 2025Battery Problem; Battery Problem
IT WAS REPORTED THAT CS300 INTRA-AORTIC BALLOON PUMP (IABP) HAS BATTERY CHARGING ISSUES. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21100277 ↗MalfunctionAug 22, 2022Battery Problem; Battery Problem
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM) THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT FAILED BATTERY RUN TIME. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15269150 ↗MalfunctionSep 9, 2021Failure to Sense; Failure to Sense
IT WAS REPORTED THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) COULD NOT OBTAIN THE ECG TRACING. IT IS UNKNOWN THE CIRCUMSTANCES UNDER WHICH THE EVENT OCCURRED. IT IS ALSO UNKNOWN IF THERE WAS A PATIENT INVOLVEMENT. HOWEVER, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12449426 ↗MalfunctionMar 30, 2021Mechanical Problem; Mechanical Problem
MalfunctionFeb 25, 2021Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT DURING A ROUTINE CHECK, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) POWERED DOWN ON ITS OWN, WHILE PLUGGED IN. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 11381186 ↗
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.