Linear 7.5Fr. 34cc IAB (China) 0684-00-0479-01CRecalls & reported problems
Datascope Corp. · Linear 7.5Fr. 34cc IAB (China) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10607567109329), received between Feb 12, 2021 and Mar 25, 2025 — classified as 16 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash7
- Difficult to Insert5
- Difficult to Remove3
- Inability to Auto-Fill1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2025Difficult to Insert; Difficult to Insert
MalfunctionNov 14, 2023Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) COULD NOT BE INSERTED THROUGH THE SHEATH. A NEW IAB WAS INSERTED SUCCESSFULLY TO CONTINUE WITH THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18130830 ↗MalfunctionNov 2, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER ABOUT TWO HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS FOUND IN THE TUBING. THE CATHETER WAS REMOVED AND NOT REPLACED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18057447 ↗MalfunctionApr 10, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING INSERTION OF THE INTRA-AORTIC BALLOON (IAB), BLOOD WAS FOUND IN THE TUBING. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16707969 ↗MalfunctionJan 11, 2023Inability to Auto-Fill; Inability to Auto-Fill
IT WAS REPORTED THAT AFTER INSERTING THE INTRA-AORTIC BALLOON (IAB) AND INITIATING THERAPY, THE CONSOLE GENERATED AN AUTOFILL FAILURE ALARM. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16141587 ↗MalfunctionAug 31, 2022Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT THE CUSTOMER COULD NOT REMOVE THE INTRA-AORTIC BALLOON (IAB) FROM THE T-HANDLE. A NEW IAB WAS INSERTED SUCCESSFULLY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15331346 ↗MalfunctionAug 16, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER APPROXIMATELY 11 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY THE CONSOLE GENERATED A LEAK IN IAB CIRCUIT ALARM. BLOOD WAS THEN FOUND IN THE TUBING. A NEW IAB WAS INSERTED TO CONTINUE THERAPY WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15234809 ↗MalfunctionApr 13, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT PRIOR TO INSERTING THE INTRA-AORTIC BALLOON (IAB) THE CUSTOMER NOTICED THAT THE TOP OF THE GUIDEWIRE WAS DAMAGED AND IT COULD NOT BE INSERTED. A NEW IAB WAS INSERTED INSTEAD. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14106886 ↗MalfunctionApr 6, 2022Leak/Splash; Difficult to Remove; Leak/Splash; Difficult to Remove
IT WAS REPORTED THAT AFTER APPROXIMATELY TWO DAYS OF INTRA-AORTIC BALLOON (IAB) THERAPY THE CONSOLE GENERATED A CHECK IAB CATHETER ALARM. THE CUSTOMER CHECKED THE IAB BUT FOUND NO ISSUE. TWO HOURS LATER, THE CONSOLE GENERATED THE SAME ALARM AND BLOOD WAS SEEN IN THE TUBING. THE IAB HAD TO BE REMOVED BY SURGERY IN THE INTENSIVE CARE UNIT (ICU). THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14022386 ↗MalfunctionFeb 4, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER 2 DAYS OF INTRA-AORTIC BALLOON(IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND LUMEN WAS FOUND TO BE BROKEN. THERE WAS NO REPORTED INJURY OR ADVERSE EVENT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13457469 ↗MalfunctionJan 7, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER ABOUT 8 HOURS OF INTRA-AORTIC BALLOON(IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE BALLOON WAS REMOVED AND REPLACED WITH A NEW ONE. THERE WAS NO REPORTED INJURY OR ADVERSE EVENT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13195149 ↗MalfunctionDec 6, 2021Difficult to Remove; Difficult to Remove
MalfunctionNov 22, 2021Difficult to Insert; Difficult to Insert
MalfunctionNov 22, 2021Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE GUIDEWIRE WAS INSERTED SMOOTHLY, BUT THE INTRA-AORTIC BALLOON (IAB) COULD NOT BE INSERTED. A NEW IAB WAS INSERTED, HOWEVER, THE SAME ISSUE OCCURRED. A THIRD IAB WAS INSERTED SUCCESSFULLY TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS REPORT IS FOR THE SECOND IAB USED IN THIS EVENT. A SEPARATE REPORT WILL BE SUBMITTED FOR THE FIRST IAB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12852356 ↗MalfunctionNov 1, 2021Use of Device Problem; Use of Device Problem
MalfunctionFeb 12, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER APPROXIMATELY 17 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS FOUND IN THE TUBING. THE IAB WAS REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11320801 ↗
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page