Linear 7.5Fr. 34cc IAB (China) 0684-00-0479-01CRecalls & reported problems

Datascope Corp. · Linear 7.5Fr. 34cc IAB (China) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10607567109329), received between Feb 12, 2021 and Mar 25, 2025 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash7
  • Difficult to Insert5
  • Difficult to Remove3
  • Inability to Auto-Fill1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2025Difficult to Insert; Difficult to Insert

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21685680 ↗
  • MalfunctionNov 14, 2023Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) COULD NOT BE INSERTED THROUGH THE SHEATH. A NEW IAB WAS INSERTED SUCCESSFULLY TO CONTINUE WITH THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18130830 ↗
  • MalfunctionNov 2, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER ABOUT TWO HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS FOUND IN THE TUBING. THE CATHETER WAS REMOVED AND NOT REPLACED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18057447 ↗
  • MalfunctionApr 10, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT DURING INSERTION OF THE INTRA-AORTIC BALLOON (IAB), BLOOD WAS FOUND IN THE TUBING. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16707969 ↗
  • MalfunctionJan 11, 2023Inability to Auto-Fill; Inability to Auto-Fill

    IT WAS REPORTED THAT AFTER INSERTING THE INTRA-AORTIC BALLOON (IAB) AND INITIATING THERAPY, THE CONSOLE GENERATED AN AUTOFILL FAILURE ALARM. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16141587 ↗
  • MalfunctionAug 31, 2022Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE CUSTOMER COULD NOT REMOVE THE INTRA-AORTIC BALLOON (IAB) FROM THE T-HANDLE. A NEW IAB WAS INSERTED SUCCESSFULLY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15331346 ↗
  • MalfunctionAug 16, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY 11 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY THE CONSOLE GENERATED A LEAK IN IAB CIRCUIT ALARM. BLOOD WAS THEN FOUND IN THE TUBING. A NEW IAB WAS INSERTED TO CONTINUE THERAPY WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15234809 ↗
  • MalfunctionApr 13, 2022Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT PRIOR TO INSERTING THE INTRA-AORTIC BALLOON (IAB) THE CUSTOMER NOTICED THAT THE TOP OF THE GUIDEWIRE WAS DAMAGED AND IT COULD NOT BE INSERTED. A NEW IAB WAS INSERTED INSTEAD. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14106886 ↗
  • MalfunctionApr 6, 2022Leak/Splash; Difficult to Remove; Leak/Splash; Difficult to Remove

    IT WAS REPORTED THAT AFTER APPROXIMATELY TWO DAYS OF INTRA-AORTIC BALLOON (IAB) THERAPY THE CONSOLE GENERATED A CHECK IAB CATHETER ALARM. THE CUSTOMER CHECKED THE IAB BUT FOUND NO ISSUE. TWO HOURS LATER, THE CONSOLE GENERATED THE SAME ALARM AND BLOOD WAS SEEN IN THE TUBING. THE IAB HAD TO BE REMOVED BY SURGERY IN THE INTENSIVE CARE UNIT (ICU). THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14022386 ↗
  • MalfunctionFeb 4, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER 2 DAYS OF INTRA-AORTIC BALLOON(IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND LUMEN WAS FOUND TO BE BROKEN. THERE WAS NO REPORTED INJURY OR ADVERSE EVENT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13457469 ↗
  • MalfunctionJan 7, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER ABOUT 8 HOURS OF INTRA-AORTIC BALLOON(IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE BALLOON WAS REMOVED AND REPLACED WITH A NEW ONE. THERE WAS NO REPORTED INJURY OR ADVERSE EVENT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13195149 ↗
  • MalfunctionDec 6, 2021Difficult to Remove; Difficult to Remove

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12942409 ↗
  • MalfunctionNov 22, 2021Difficult to Insert; Difficult to Insert

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12852361 ↗
  • MalfunctionNov 22, 2021Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE GUIDEWIRE WAS INSERTED SMOOTHLY, BUT THE INTRA-AORTIC BALLOON (IAB) COULD NOT BE INSERTED. A NEW IAB WAS INSERTED, HOWEVER, THE SAME ISSUE OCCURRED. A THIRD IAB WAS INSERTED SUCCESSFULLY TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS REPORT IS FOR THE SECOND IAB USED IN THIS EVENT. A SEPARATE REPORT WILL BE SUBMITTED FOR THE FIRST IAB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12852356 ↗
  • MalfunctionNov 1, 2021Use of Device Problem; Use of Device Problem

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12728061 ↗
  • MalfunctionFeb 12, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY 17 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS FOUND IN THE TUBING. THE IAB WAS REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11320801 ↗

Other System, Balloon, Intra-Aortic And Control devices