Linear 7.5Fr. 40cc IAB (China) 0684-00-0480-01CRecalls & reported problems

Datascope Corp. · Linear 7.5Fr. 40cc IAB (China) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Datascope Linear 7.5Fr. 40cc intra-aortic balloon catheter is used to support the heart during critical care. Most of the 32 reports describe the balloon leaking fluid or splashing, failing to inflate fully, or being hard to insert; one report ends in a death and none describe injuries. One open recall, started in June 2023, covers a dilator that can break off at the hub during insertion and leave a fragment behind.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
1
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 10607567109343), received between Aug 26, 2020 and Feb 17, 2026 — classified as 1 death, 31 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (18 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 23, 2023Class IIOpen, Classified

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

    Z-2223-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash18
  • Inflation Problem6
  • Difficult to Insert3
  • Filling Problem2
  • Break1
  • Defective Component1
  • Fluid/Blood Leak1
  • Inability to Auto-Fill1

Patient problems

  • No Clinical Signs, Symptoms or Conditions29
  • No Consequences Or Impact To Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 17, 2026Inflation Problem; Unraveled Material

    IT WAS REPORTED BY THE CUSTOMER THAT DURING THE INSERT OF INTRA-AORTIC BALLOON (IAB) CATHETER, FOUND THE IAB CAN NOT FULLY INFLATION. THERE WAS NO PATIENT HARM OR INJURY REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24367766 ↗
  • MalfunctionNov 26, 2025Leak/Splash; Material Rupture; Leak/Splash; Material Rupture

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23654082 ↗
  • MalfunctionSep 18, 2025Fluid/Blood Leak; Fluid/Blood Leak

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23082147 ↗
  • MalfunctionJan 10, 2025Inflation Problem; Inflation Problem

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21121870 ↗
  • DeathSep 27, 2024Pressure Problem; Pressure Problem

    IT WAS REPORTED THAT THE PATIENT WAS ADMITTED WITH ACUTE MYOCARDIAL INFARCTION (AMI) AND AN INTRA-AORTIC BALLOON (IAB) WAS IMMEDIATELY INSERTED. WHEN CONNECTED TO THE PUMP, AN UNKNOWN ALARM WAS GENERATED. IT WAS NOTED THAT THE BLOOD PRESSURE (BP) DISPLAYED BY THE PUMP WAS VERY DIFFERENT FROM THAT OF THE BEDSIDE MONITOR. THE PHYSICIAN IMMEDIATELY INSERTED A NEW IAB, BUT THEY THOUGHT THE BEST TIME FOR TREATMENT WAS MISSED. THE PATIENT REMAINED IN THE ICU. THE PATIENT PASSED AWAY TWO DAYS LATER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 20318534 ↗
  • MalfunctionSep 23, 2024Inflation Problem; Inflation Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20280661 ↗
  • MalfunctionSep 23, 2024Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT AFTER INSERTION, THE INTRA-AORTIC BALLOON (IAB) WOULD NOT FULLY INFLATE. THE IAB WAS REMOVED IMMEDIATELY. A NEW IAB WAS INSERTED, BUT THE SAME ISSUE OCCURRED. A THIRD IAB WAS NOT INSERTED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS REPORT IS FOR THE 2ND IAB INVOLVED IN THIS EVENT. A SEPARATE REPORT WILL BE SUBMITTED FOR THE 1ST IAB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20280600 ↗
  • MalfunctionFeb 21, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT DURING INSERTION, THE GUIDEWIRE COULD NOT BE PASSED THROUGH THE INTRA-AORTIC BALLOON (IAB). UPON INSPECTION, A DEPRESSION WAS FOUND ON THE IAB 10 CENTIMETERS FROM THE TIP. THERE WAS NO DAMAGE NOTED TO THE IAB PACKAGING. A NEW IAB WAS INSERTED SUCCESSFULLY TO CONTINUE THERAPY WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18753923 ↗
  • MalfunctionOct 16, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY 45 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, THE PUMP GENERATED A FREQUENT ALARM. BLOOD WAS SEEN IN THE TUBING. THE IAB WAS THEN REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17942335 ↗
  • MalfunctionOct 4, 2023Break; Break

    IT WAS REPORTED THAT THE CUSTOMER INSERTED THE GUIDEWIRE SMOOTHLY, HOWEVER, THE INTRA-AORTIC BALLOON (IAB) COULD NOT BE INSERTED COMPLETELY. THE IAB WAS CHECKED AND IT WAS FOUND THAT THE INNER LUMEN WAS COMPLETELY SEPARATED FROM IAB BASE. THEY DID NOT CONTINUE WITH THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17869569 ↗
  • MalfunctionNov 28, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY TWO HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS FOUND IN THE TUBING. THE IAB WAS IMMEDIATELY REMOVED AND IT WAS FOUND THAT THE INNER LUMEN WAS BROKEN. THE CUSTOMER INSERTED A NEW IAB TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15871225 ↗
  • MalfunctionSep 27, 2022Leak/Splash; Leak/Splash

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15490851 ↗
  • MalfunctionAug 30, 2022Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT AFTER INSERTING THE INTRA-AORTIC BALLOON (IAB) AND INITIATING THERAPY, THE CUSTOMER FOUND THE IAB WOULD NOT INFLATE. THEY THEN PULLED OUT THE IAB AND REPLACED IT WITH A NEW ONE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15322977 ↗
  • MalfunctionAug 24, 2022Difficult to Insert; Leak/Splash; Difficult to Insert; Leak/Splash

    IT WAS REPORTED THAT THERE WAS RESISTANCE WHILE INSERTING THE INTRA-AORTIC BALLOON (IAB) AND BLOOD WAS FOUND IN THE IAB. THEY THEN PULLED OUT THE IAB AND INSERTED A NEW ONE WITHOUT FURTHER ISSUES. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15282226 ↗
  • MalfunctionAug 16, 2022Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT WHEN THE INTRA-AORTIC BALLOON (IAB) WAS INSERTED AND THE CONSOLE WAS STARTED, THE IAB WOULD NOT INFLATE AND A CHECK IAB CATHETER ALARM WAS GENERATED. A NEW IAB WAS INSERTED TO CONTINUE THERAPY WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15234415 ↗
  • MalfunctionApr 6, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY THREE DAYS OF INTRA-AORTIC BALLOON (IAB) THERAPY THE CONSOLE GENERATED A RAPID GAS LOSS ALARM. BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14022263 ↗
  • MalfunctionApr 4, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY ONE HOUR OF INTRA-AORTIC BALLOON (IAB) THERAPY BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND REPLACED WITH A NEW ONE TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13998513 ↗
  • MalfunctionFeb 4, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER SIX DAYS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13450953 ↗
  • MalfunctionDec 31, 2021Inability to Auto-Fill; Inability to Auto-Fill

    IT WAS REPORTED THAT AFTER INSERTING AND CONNECTING THE INTRA-AORTIC BALLOON (IAB), THE CONSOLE GENERATED AN AUTO-FILL FAILURE ALARM. THE CUSTOMER CHANGED THE CATHETER EXTENDER FOR A NEW ONE AND THERAPY WAS CONTINUED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13130204 ↗
  • MalfunctionDec 14, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) HAD BEEN INSERTED SMOOTHLY. HOWEVER, AFTER APPROXIMATELY ONE DAY OF THERAPY, THE CONSOLE GENERATED A BLOOD DETECTED ALARM AND BLOOD WAS FOUND IN THE TUBING. THE IAB WAS THEN REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13005548 ↗
  • MalfunctionDec 14, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) WAS INSERTED SMOOTHLY. AFTER 43 HOURS OF THERAPY, BLOOD WAS SEEN IN THE TUBING AND THE IAB WAS REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13003000 ↗
  • MalfunctionDec 6, 2021No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT AFTER APPROXIMATELY TWO HOURS OF INTRA-AORTIC BALLOON (IAB), THE CONSOLE GENERATED AN UNABLE TO UPDATE TIMING ALARM. THE IAB WAS REMOVED. THE PATIENT WAS STABLE SO ANOTHER IAB WAS NOT INSERTED. THE PATIENT WAS TREATED WITH MEDICATION. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12934244 ↗
  • MalfunctionDec 6, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY 45 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND MEDICATION THERAPY WAS USED TO TREAT THE PATIENT. TWO DAYS LATER, A NEW IAB WAS INSERTED. AFTER APPROXIMATELY 58 HOURS OF IAB THERAPY WITH THE SECOND IAB, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND MEDICATION TREATMENT WAS CONTINUED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS REPORT IS FOR THE SECOND IAB USED. A SEPARATE REPORT WILL BE SUBMITTED FOR THE FIRST IAB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12934026 ↗
  • MalfunctionDec 6, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY 45 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND MEDICATION THERAPY WAS USED TO TREAT THE PATIENT. TWO DAYS LATER, A NEW IAB WAS INSERTED. AFTER APPROXIMATELY 58 HOURS OF IAB THERAPY WITH THE SECOND IAB, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND MEDICATION TREATMENT WAS CONTINUED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS REPORT IS FOR THE FIRST IAB USED. A SEPARATE REPORT WILL BE SUBMITTED FOR THE SECOND IAB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12934020 ↗
  • MalfunctionNov 30, 2021Leak/Splash; Leak/Splash

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12892671 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →