Linear 7.5Fr. 40cc IAB (China) 0684-00-0480-01CRecalls & reported problems
Datascope Corp. · Linear 7.5Fr. 40cc IAB (China) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Datascope Linear 7.5Fr. 40cc intra-aortic balloon catheter is used to support the heart during critical care. Most of the 32 reports describe the balloon leaking fluid or splashing, failing to inflate fully, or being hard to insert; one report ends in a death and none describe injuries. One open recall, started in June 2023, covers a dilator that can break off at the hub during insertion and leave a fragment behind.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 10607567109343), received between Aug 26, 2020 and Feb 17, 2026 — classified as 1 death, 31 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (18 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 23, 2023Class IIOpen, Classified
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Z-2223-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash18
- Inflation Problem6
- Difficult to Insert3
- Filling Problem2
- Break1
- Defective Component1
- Fluid/Blood Leak1
- Inability to Auto-Fill1
Patient problems
- No Clinical Signs, Symptoms or Conditions29
- No Consequences Or Impact To Patient2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 17, 2026Inflation Problem; Unraveled Material
IT WAS REPORTED BY THE CUSTOMER THAT DURING THE INSERT OF INTRA-AORTIC BALLOON (IAB) CATHETER, FOUND THE IAB CAN NOT FULLY INFLATION. THERE WAS NO PATIENT HARM OR INJURY REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24367766 ↗MalfunctionNov 26, 2025Leak/Splash; Material Rupture; Leak/Splash; Material Rupture
MalfunctionSep 18, 2025Fluid/Blood Leak; Fluid/Blood Leak
MalfunctionJan 10, 2025Inflation Problem; Inflation Problem
DeathSep 27, 2024Pressure Problem; Pressure Problem
IT WAS REPORTED THAT THE PATIENT WAS ADMITTED WITH ACUTE MYOCARDIAL INFARCTION (AMI) AND AN INTRA-AORTIC BALLOON (IAB) WAS IMMEDIATELY INSERTED. WHEN CONNECTED TO THE PUMP, AN UNKNOWN ALARM WAS GENERATED. IT WAS NOTED THAT THE BLOOD PRESSURE (BP) DISPLAYED BY THE PUMP WAS VERY DIFFERENT FROM THAT OF THE BEDSIDE MONITOR. THE PHYSICIAN IMMEDIATELY INSERTED A NEW IAB, BUT THEY THOUGHT THE BEST TIME FOR TREATMENT WAS MISSED. THE PATIENT REMAINED IN THE ICU. THE PATIENT PASSED AWAY TWO DAYS LATER.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20318534 ↗MalfunctionSep 23, 2024Inflation Problem; Inflation Problem
MalfunctionSep 23, 2024Inflation Problem; Inflation Problem
IT WAS REPORTED THAT AFTER INSERTION, THE INTRA-AORTIC BALLOON (IAB) WOULD NOT FULLY INFLATE. THE IAB WAS REMOVED IMMEDIATELY. A NEW IAB WAS INSERTED, BUT THE SAME ISSUE OCCURRED. A THIRD IAB WAS NOT INSERTED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS REPORT IS FOR THE 2ND IAB INVOLVED IN THIS EVENT. A SEPARATE REPORT WILL BE SUBMITTED FOR THE 1ST IAB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20280600 ↗MalfunctionFeb 21, 2024Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING INSERTION, THE GUIDEWIRE COULD NOT BE PASSED THROUGH THE INTRA-AORTIC BALLOON (IAB). UPON INSPECTION, A DEPRESSION WAS FOUND ON THE IAB 10 CENTIMETERS FROM THE TIP. THERE WAS NO DAMAGE NOTED TO THE IAB PACKAGING. A NEW IAB WAS INSERTED SUCCESSFULLY TO CONTINUE THERAPY WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18753923 ↗MalfunctionOct 16, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER APPROXIMATELY 45 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, THE PUMP GENERATED A FREQUENT ALARM. BLOOD WAS SEEN IN THE TUBING. THE IAB WAS THEN REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17942335 ↗MalfunctionOct 4, 2023Break; Break
IT WAS REPORTED THAT THE CUSTOMER INSERTED THE GUIDEWIRE SMOOTHLY, HOWEVER, THE INTRA-AORTIC BALLOON (IAB) COULD NOT BE INSERTED COMPLETELY. THE IAB WAS CHECKED AND IT WAS FOUND THAT THE INNER LUMEN WAS COMPLETELY SEPARATED FROM IAB BASE. THEY DID NOT CONTINUE WITH THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17869569 ↗MalfunctionNov 28, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER APPROXIMATELY TWO HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS FOUND IN THE TUBING. THE IAB WAS IMMEDIATELY REMOVED AND IT WAS FOUND THAT THE INNER LUMEN WAS BROKEN. THE CUSTOMER INSERTED A NEW IAB TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15871225 ↗MalfunctionSep 27, 2022Leak/Splash; Leak/Splash
MalfunctionAug 30, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT AFTER INSERTING THE INTRA-AORTIC BALLOON (IAB) AND INITIATING THERAPY, THE CUSTOMER FOUND THE IAB WOULD NOT INFLATE. THEY THEN PULLED OUT THE IAB AND REPLACED IT WITH A NEW ONE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15322977 ↗MalfunctionAug 24, 2022Difficult to Insert; Leak/Splash; Difficult to Insert; Leak/Splash
IT WAS REPORTED THAT THERE WAS RESISTANCE WHILE INSERTING THE INTRA-AORTIC BALLOON (IAB) AND BLOOD WAS FOUND IN THE IAB. THEY THEN PULLED OUT THE IAB AND INSERTED A NEW ONE WITHOUT FURTHER ISSUES. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15282226 ↗MalfunctionAug 16, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT WHEN THE INTRA-AORTIC BALLOON (IAB) WAS INSERTED AND THE CONSOLE WAS STARTED, THE IAB WOULD NOT INFLATE AND A CHECK IAB CATHETER ALARM WAS GENERATED. A NEW IAB WAS INSERTED TO CONTINUE THERAPY WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15234415 ↗MalfunctionApr 6, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER APPROXIMATELY THREE DAYS OF INTRA-AORTIC BALLOON (IAB) THERAPY THE CONSOLE GENERATED A RAPID GAS LOSS ALARM. BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14022263 ↗MalfunctionApr 4, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER APPROXIMATELY ONE HOUR OF INTRA-AORTIC BALLOON (IAB) THERAPY BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND REPLACED WITH A NEW ONE TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13998513 ↗MalfunctionFeb 4, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER SIX DAYS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13450953 ↗MalfunctionDec 31, 2021Inability to Auto-Fill; Inability to Auto-Fill
IT WAS REPORTED THAT AFTER INSERTING AND CONNECTING THE INTRA-AORTIC BALLOON (IAB), THE CONSOLE GENERATED AN AUTO-FILL FAILURE ALARM. THE CUSTOMER CHANGED THE CATHETER EXTENDER FOR A NEW ONE AND THERAPY WAS CONTINUED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13130204 ↗MalfunctionDec 14, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) HAD BEEN INSERTED SMOOTHLY. HOWEVER, AFTER APPROXIMATELY ONE DAY OF THERAPY, THE CONSOLE GENERATED A BLOOD DETECTED ALARM AND BLOOD WAS FOUND IN THE TUBING. THE IAB WAS THEN REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13005548 ↗MalfunctionDec 14, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) WAS INSERTED SMOOTHLY. AFTER 43 HOURS OF THERAPY, BLOOD WAS SEEN IN THE TUBING AND THE IAB WAS REMOVED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13003000 ↗MalfunctionDec 6, 2021No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT AFTER APPROXIMATELY TWO HOURS OF INTRA-AORTIC BALLOON (IAB), THE CONSOLE GENERATED AN UNABLE TO UPDATE TIMING ALARM. THE IAB WAS REMOVED. THE PATIENT WAS STABLE SO ANOTHER IAB WAS NOT INSERTED. THE PATIENT WAS TREATED WITH MEDICATION. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12934244 ↗MalfunctionDec 6, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER APPROXIMATELY 45 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND MEDICATION THERAPY WAS USED TO TREAT THE PATIENT. TWO DAYS LATER, A NEW IAB WAS INSERTED. AFTER APPROXIMATELY 58 HOURS OF IAB THERAPY WITH THE SECOND IAB, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND MEDICATION TREATMENT WAS CONTINUED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS REPORT IS FOR THE SECOND IAB USED. A SEPARATE REPORT WILL BE SUBMITTED FOR THE FIRST IAB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12934026 ↗MalfunctionDec 6, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER APPROXIMATELY 45 HOURS OF INTRA-AORTIC BALLOON (IAB) THERAPY, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND MEDICATION THERAPY WAS USED TO TREAT THE PATIENT. TWO DAYS LATER, A NEW IAB WAS INSERTED. AFTER APPROXIMATELY 58 HOURS OF IAB THERAPY WITH THE SECOND IAB, BLOOD WAS SEEN IN THE TUBING. THE IAB WAS REMOVED AND MEDICATION TREATMENT WAS CONTINUED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS REPORT IS FOR THE FIRST IAB USED. A SEPARATE REPORT WILL BE SUBMITTED FOR THE SECOND IAB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12934020 ↗MalfunctionNov 30, 2021Leak/Splash; Leak/Splash
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page