Sensation 7Fr. 40cc IAB (China) 0684-00-0470-01CRecalls & reported problems
Datascope Corp. · Sensation 7Fr. 40cc IAB (China) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10607567109374), received between Jan 14, 2021 and Jul 7, 2026 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 23, 2023Class IIOpen, Classified
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Z-2217-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Sensing Problem2
- Optical Problem2
- Calibration Problem1
- Difficult to Remove1
- Inappropriate Waveform1
- Leak/Splash1
- Material Deformation1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information2
- Foreign Body In Patient1
- Ischemia1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJul 7, 2026Difficult to Remove; Material Rupture
ON (B)(6) 2026, THE PATIENT SUFFERED AN ACUTE MYOCARDIAL INFARCTION AND WAS IN CRITICAL CONDITION. AS A LIFE-SAVING MEASURE, AN INTRA-AORTIC BALLOON (IAB) WAS INSERTED. THE PROCEDURE WAS TECHNICALLY CHALLENGING BECAUSE ULTRASONOGRAPHY REVEALED SEVERE CALCIFICATION OF THE LOWER-EXTREMITY ARTERIES. UNDER ULTRASOUND GUIDANCE, THE IAB WAS EVENTUALLY INSERTED WITH CONSIDERABLE DIFFICULTY, RESULTING IN STABILIZATION OF THE PATIENT'S HEMODYNAMIC STATUS. THE PATIENT SUBSEQUENTLY UNDERWENT SUCCESSFUL CORONARY ARTERY BYPASS GRAFTING (CABG). ON (B)(6) 2026, DURING ATTEMPTED REMOVAL OF THE IAB, THE BALLOON BECAME ENTRAPPED WITHIN ATHEROSCLEROTIC PLAQUE AND COULD NOT BE WITHDRAWN THROUGH THE VASCULAR ACCESS ROUTE. VASCULAR SURGEONS AND GENERAL SURGEONS WERE CONSULTED, AND SURGICAL EXPLORATION OF THE ABDOMEN AND GROIN WAS PERFORMED WITH ARTERIOTOMY TO FACILITATE REMOVAL OF THE BALLOON. INTRAOPERATIVELY, THE ABDOMINAL AORTA AND COMMON ILIAC ARTERIES WERE FOUND TO BE SEVERELY CALCIFIED, WITH MARKED STENOSIS AND NEAR-COMPLETE OCCLUSION IN SOME SEGMENTS. DUE TO THE EXTENSIVE VASCULAR CALCIFICATION AND UNFAVORABLE ANATOMY, COMPLETE RETRIEVAL OF THE DEVICE WAS NOT POSSIBLE. CONSEQUENTLY, A SMALL STERI
Patient problems: Foreign Body In Patient; Insufficient Information; Ischemia
MAUDE report 25729827 ↗MalfunctionFeb 24, 2026Device Sensing Problem
IT WAS REPORTED THAT, WHILE USING SENSATION 7FR. 40CC INTRA-AORTIC BALLOON PUMP (IABP), THE PATIENT'S ELECTROCARDIOGRAM (ECG) WAS PACING ABNORMALITIES AND UNABLE TO PERFORM PACING; SIMULTANEOUSLY, DUE TO THE FIBER OPTIC FAILURE, THE IABP COULD NOT PROVIDE NORMAL COUNTERPULSATION. NO HARM OR INJURY TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24430439 ↗MalfunctionMar 27, 2025Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT INTRA AORTIC BALLOON (IAB) FIBER OPTIC WAS NOT WORKING EVEN AFTER CHECKED ON TWO MACHINES.
Patient problems: Insufficient Information
MAUDE report 21706631 ↗MalfunctionApr 10, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AFTER INSERTING THE INTRA-AORTIC BALLOON (IAB), BLOOD WAS SEEN IN THE TUBING. A NEW IAB WAS THEN INSERTED TO CONTINUE THERAPY, BUT THE SAME ISSUE OCCURRED AND BLOOD WAS SEEN IN THE TUBING AGAIN. THE PATIENT WAS THEN TREATED USING MEDICATION INSTEAD OF IAB THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS RECORD IS FOR THE 2ND IAB USED. A SEPARATE REPORT HAS BEEN SUBMITTED FOR THE 1ST IAB UNDER MFG REPORT NUMBER 2248146-2023-00245.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16707757 ↗MalfunctionFeb 20, 2023Inappropriate Waveform; Optical Problem; Pressure Problem; Inappropriate Waveform; Optical Problem; Pressure Problem
MalfunctionJan 10, 2023Calibration Problem; Calibration Problem
MalfunctionDec 20, 2022Material Deformation; Material Deformation
MalfunctionJan 14, 2021Optical Problem; Optical Problem
IT WAS REPORTED THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, THE OPTICAL FIBER PRESSURE SIGNAL IS SOMETIMES NOT AVAILABLE. THE IAB WAS REPLACED WITH A NEW ONE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11169713 ↗
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page