Sensation 7Fr. 40cc IAB (China) 0684-00-0470-01CRecalls & reported problems

Datascope Corp. · Sensation 7Fr. 40cc IAB (China) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10607567109374), received between Jan 14, 2021 and Jul 7, 2026 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 23, 2023Class IIOpen, Classified

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

    Z-2217-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Sensing Problem2
  • Optical Problem2
  • Calibration Problem1
  • Difficult to Remove1
  • Inappropriate Waveform1
  • Leak/Splash1
  • Material Deformation1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information2
  • Foreign Body In Patient1
  • Ischemia1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 7, 2026Difficult to Remove; Material Rupture

    ON (B)(6) 2026, THE PATIENT SUFFERED AN ACUTE MYOCARDIAL INFARCTION AND WAS IN CRITICAL CONDITION. AS A LIFE-SAVING MEASURE, AN INTRA-AORTIC BALLOON (IAB) WAS INSERTED. THE PROCEDURE WAS TECHNICALLY CHALLENGING BECAUSE ULTRASONOGRAPHY REVEALED SEVERE CALCIFICATION OF THE LOWER-EXTREMITY ARTERIES. UNDER ULTRASOUND GUIDANCE, THE IAB WAS EVENTUALLY INSERTED WITH CONSIDERABLE DIFFICULTY, RESULTING IN STABILIZATION OF THE PATIENT'S HEMODYNAMIC STATUS. THE PATIENT SUBSEQUENTLY UNDERWENT SUCCESSFUL CORONARY ARTERY BYPASS GRAFTING (CABG). ON (B)(6) 2026, DURING ATTEMPTED REMOVAL OF THE IAB, THE BALLOON BECAME ENTRAPPED WITHIN ATHEROSCLEROTIC PLAQUE AND COULD NOT BE WITHDRAWN THROUGH THE VASCULAR ACCESS ROUTE. VASCULAR SURGEONS AND GENERAL SURGEONS WERE CONSULTED, AND SURGICAL EXPLORATION OF THE ABDOMEN AND GROIN WAS PERFORMED WITH ARTERIOTOMY TO FACILITATE REMOVAL OF THE BALLOON. INTRAOPERATIVELY, THE ABDOMINAL AORTA AND COMMON ILIAC ARTERIES WERE FOUND TO BE SEVERELY CALCIFIED, WITH MARKED STENOSIS AND NEAR-COMPLETE OCCLUSION IN SOME SEGMENTS. DUE TO THE EXTENSIVE VASCULAR CALCIFICATION AND UNFAVORABLE ANATOMY, COMPLETE RETRIEVAL OF THE DEVICE WAS NOT POSSIBLE. CONSEQUENTLY, A SMALL STERI

    Patient problems: Foreign Body In Patient; Insufficient Information; Ischemia

    MAUDE report 25729827 ↗
  • MalfunctionFeb 24, 2026Device Sensing Problem

    IT WAS REPORTED THAT, WHILE USING SENSATION 7FR. 40CC INTRA-AORTIC BALLOON PUMP (IABP), THE PATIENT'S ELECTROCARDIOGRAM (ECG) WAS PACING ABNORMALITIES AND UNABLE TO PERFORM PACING; SIMULTANEOUSLY, DUE TO THE FIBER OPTIC FAILURE, THE IABP COULD NOT PROVIDE NORMAL COUNTERPULSATION. NO HARM OR INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24430439 ↗
  • MalfunctionMar 27, 2025Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT INTRA AORTIC BALLOON (IAB) FIBER OPTIC WAS NOT WORKING EVEN AFTER CHECKED ON TWO MACHINES.

    Patient problems: Insufficient Information

    MAUDE report 21706631 ↗
  • MalfunctionApr 10, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER INSERTING THE INTRA-AORTIC BALLOON (IAB), BLOOD WAS SEEN IN THE TUBING. A NEW IAB WAS THEN INSERTED TO CONTINUE THERAPY, BUT THE SAME ISSUE OCCURRED AND BLOOD WAS SEEN IN THE TUBING AGAIN. THE PATIENT WAS THEN TREATED USING MEDICATION INSTEAD OF IAB THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED. THIS RECORD IS FOR THE 2ND IAB USED. A SEPARATE REPORT HAS BEEN SUBMITTED FOR THE 1ST IAB UNDER MFG REPORT NUMBER 2248146-2023-00245.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16707757 ↗
  • MalfunctionFeb 20, 2023Inappropriate Waveform; Optical Problem; Pressure Problem; Inappropriate Waveform; Optical Problem; Pressure Problem

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16404531 ↗
  • MalfunctionJan 10, 2023Calibration Problem; Calibration Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16133902 ↗
  • MalfunctionDec 20, 2022Material Deformation; Material Deformation

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16017978 ↗
  • MalfunctionJan 14, 2021Optical Problem; Optical Problem

    IT WAS REPORTED THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, THE OPTICAL FIBER PRESSURE SIGNAL IS SOMETIMES NOT AVAILABLE. THE IAB WAS REPLACED WITH A NEW ONE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11169713 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →