Skip to content

Linear 7.5Fr. 25cc IAB with Accessories, US Only 0684-00-0478-01URecalls & reported problems

Datascope Corp. · Linear 7.5Fr. 25cc IAB with Accessories, US Only · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10607567109480), received between Sep 3, 2021 and Sep 17, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 23, 2023Class IIOpen, Classified

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

    Z-2223-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2024Leak/Splash; Leak/Splash

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20237697 ↗
  • MalfunctionSep 6, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY ONE MONTH OF INTRA-AORTIC BALLOON (IAB) THERAPY, THE NURSE NOTICED FROTHY BLOOD IN THE GAS CHAMBER OF THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP). THE INSERTION WAS REPORTED TO BE AXILLARY, WHICH IS NOT THE METHOD DESCRIBED IN THE DEVICE INSTRUCTIONS FOR USE. THERE WERE NO ALARMS NOTED ON THE IABP, BUT IT WAS IMMEDIATELY PLACED IN STAND-BY. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO REPORTED INJURY OR ADVERSE EVENT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15364215 ↗
  • MalfunctionOct 18, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER 9 DAYS OF INTRA-AORTIC BALLOON (IAB) THERAPY, A LEAK OCCURRED AND BLOOD WAS SEEN IN THE TUBING. THE INSERTION WAS REPORTED TO BE AXILLARY, WHICH IS NOT THE METHOD DESCRIBED IN THE DEVICE INSTRUCTIONS FOR USE. THE IAB WAS REMOVED AND NOT REPLACED BECAUSE THE PATIENT WAS STABLE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12647477 ↗
  • MalfunctionSep 3, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, A LEAK OCCURRED AND BLOOD WAS SEEN IN THE TUBING. THE INSERTION WAS REPORTED TO BE AXILLARY, WHICH IS NOT THE METHOD DESCRIBED IN THE DEVICE INSTRUCTIONS FOR USE. THE ALARM HISTORY OF THE CONSOLE SHOWED SEVERAL CATHETER ALARMS THAT ARE RELATED TO A CATHETER THAT HAD A MINOR LEAK. THE ALARMS INCLUDED AUTO-FILL FAILURE, IAB DISCONNECTED, AND IAB CATHETER RESTRICTION. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12424760 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page