Sensation 7Fr. 40cc IAB with Accessories, US Only 0684-00-0470-01URecalls & reported problems

Datascope Corp. · Sensation 7Fr. 40cc IAB with Accessories, US Only · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10607567109558), received between Aug 13, 2024 and Mar 13, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 23, 2023Class IIOpen, Classified

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

    Z-2217-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Insert1
  • Fluid/Blood Leak1
  • Off-Label Use1
  • Optical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2025Optical Problem; Optical Problem

    IT WAS REPORTED DURING USE THAT INTRA-AORTIC BALLOON(IAB) FIBER OPTIC FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21591668 ↗
  • MalfunctionJan 6, 2025Fluid/Blood Leak; Off-Label Use; Fluid/Blood Leak; Off-Label Use

    IT WAS REPORTED THAT DURING THE INTRA AORTIC BALLOON (IAB) THERAPY THE CATHETER WAS PLACED THROUGH AN AXILLARY APPROACH, OFF LABLE. WHEN TRYING TO GET THE PATIENT IN A STANDING POSITION FROM SITTING THE IABP BEGAN TO ALARM "CHECK IABP CATHETER" AND BLOOD WAS NOTICED IN THE HELIUM TUBING. THE CATHETER WAS REMOVED WITH NO HARM TO THE PATIENT. .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21077544 ↗
  • MalfunctionAug 13, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE PHYSICIAN MET RESISTANCE WHILE INSERTING THE INTRA-AORTIC BALLOON (IAB) THROUGH THE SHEATH. THE IAB WAS REMOVED AND A NEW ONE WAS INSERTED WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19973401 ↗