Skip to content

MEGA 7.5Fr. 40cc IAB with Accessories, US Only 0684-00-0295-01URecalls & reported problems

Datascope Corp. · MEGA 7.5Fr. 40cc IAB with Accessories, US Only · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10607567109596), received between Jul 29, 2026 and Aug 11, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove1
  • Fluid/Blood Leak1
  • Material Rupture1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50208066 ↗
  • InjuryJul 29, 2026Fluid/Blood Leak; Difficult to Remove; Material Rupture

    On (B)(6) at 08:05, while assisting the patient to the toilet, the bedside Registered Nurse (RN) observed blood in the intra-aortic balloon pump (IABP) helium line, indicating a suspected MEGA 40cc Intra-Aortic Balloon (IAB) rupture. The Respiratory Therapist (RT) responded promptly to assess the situation. In accordance with standard practice, the IABP was immediately turned off, and the helium line was clamped to prevent further blood entry into the balloon. A critical care physician attempted to remove the IAB catheter; however, significant resistance was encountered during extraction. During the removal attempt, the patient developed stroke-like symptoms, including weakness and headache. Cardiothoracic Surgery was consulted and successfully removed the catheter without further complication.Following catheter removal, the patient's neurological symptoms resolved completely. Computed tomography (CT) scans of the head and chest were performed to evaluate for potential complications. Both studies were unremarkable, with no evidence of acute intracranial pathology, vascular injury, or other procedure-related complications. The patient remained hemodynamically stable, and a replaceme

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50065656 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page