Sensation 7Fr. 40cc IAB (INDIA) 0684-00-0470-07Recalls & reported problems
Datascope Corp. · Sensation 7Fr. 40cc IAB (INDIA) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10607567109749), received between Oct 24, 2022 and Apr 24, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 23, 2023Class IIOpen, Classified
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Z-2217-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Inflation Problem2
- Leak/Splash2
- Device Sensing Problem1
- Difficult to Insert1
- Inability to Auto-Fill1
- Material Deformation1
- Pressure Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 24, 2026Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT WAS INITIATED ON INTRA AORTIC BALLOON PUMP (IABP) SUPPORT WITH 40 CC SENSATION BALLOON CATHETER. AFTER INSERTION AND CONNECTION TO THE IABP CONSOLE, NO ARTERIAL PRESSURE SIGNAL WAS DETECTED FROM IAB. A PERSISTENT IAB OPTICAL SENSOR FAILURE ALARM INDICATED MALFUNCTION OF THE FIBER OPTIC PRESSURE SENSING SYSTEM. IABP SUPPORT WAS CONTINUED IN ECG TRIGGER MODE USING THE SAME BALLOON CATHETER TO MAINTAIN SYNCHRONIZATION WITH THE CARDIAC CYCLE. ARTERIAL BLOOD PRESSURE WAS MONITORED VIA AN ALTERNATE INVASIVE ARTERIAL LINE CONNECTED TO THE BEDSIDE MONITOR. NO HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24983119 ↗MalfunctionJul 24, 2023Inflation Problem; Leak/Splash; Pressure Problem; Inflation Problem; Leak/Splash; Pressure Problem
MalfunctionJul 12, 2023Inability to Auto-Fill; Inflation Problem; Material Deformation; Inability to Auto-Fill; Inflation Problem; Material Deformation
IT WAS REPORTED THAT AFTER INSERTION, THE CONSOLE GENERATED AN AUTOFILL FAILURE ALARM AND THE INTRA-AORTIC BALLOON (IAB) WAS NOT ABLE TO INFLATE. THE CUSTOMER FOUND A KINK IN MIDDLE OF THE IAB. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17309521 ↗MalfunctionNov 22, 2022Difficult to Insert; Difficult to Insert
MalfunctionOct 24, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE CUSTOMER ATTEMPTED A SHEATH-LESS INSERTION OF THE INTRA-AORTIC BALLOON (IAB) AND FOUND BLOOD INSIDE THE CENTRAL LUMEN. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15663562 ↗
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page