Skip to content

MEGA 8Fr. 50cc IAB (INDIA)(with APA) 0684-00-0498-08Recalls & reported problems

Datascope Corp. · MEGA 8Fr. 50cc IAB (INDIA)(with APA) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 23, 2023Class IIOpen, Classified

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

    Z-2220-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page