Skip to content

MEGA 7.5 Fr. 40cc IAB W/O Statlock - INDIA 0684-00-0295-07Recalls & reported problems

Datascope Corp. · MEGA 7.5 Fr. 40cc IAB W/O Statlock - INDIA · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10607567111339), received between Jun 7, 2021 and Apr 17, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 23, 2023Class IIOpen, Classified

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

    Z-2219-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inflation Problem2
  • Component Missing1
  • Leak/Splash1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2025Material Rupture; Material Rupture

    N/A

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21854969 ↗
  • MalfunctionNov 21, 2023Inflation Problem; Inflation Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18187398 ↗
  • MalfunctionJun 1, 2023Component Missing; Component Missing

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17038109 ↗
  • MalfunctionAug 8, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AFTER APPROXIMATELY 30 MINUTES OF INTRA-AORTIC BALLOON (IAB) THERAPY, THE CONSOLE GENERATED A LEAK IN IAB CIRCUIT ALARM AND BLOOD WAS SEEN IN THE HELIUM TUBING. AFTER REMOVAL, IT WAS FOUND THAT THE INNER LUMEN WAS BROKEN. A NEW IAB WAS INSERTED TO CONTINUE THERAPY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15187314 ↗
  • MalfunctionJun 7, 2021Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT AFTER THE INSERTION OF THE INTRA-AORTIC BALLOON (IAB), THE IAB WAS NOT AUGMENTING AND COULD NOT INFLATE. THE CONSOLE GENERATED CHECK IAB CATHETER ALARM. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11949442 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →