CS100 Intra-Aortic Balloon Pump, Refurbished, CHINESE,NON-UTS,INTL 0998-UC-3013-45Recalls & reported problems
Datascope Corp. · CS100 Intra-Aortic Balloon Pump, Refurbished, CHINESE,NON-UTS,INTL · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10607567111957), received between Sep 28, 2020 and Jul 17, 2023 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (3 reports).
8 other recalls cover devices cleared under the same FDA submission (K031636); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available3
- Gas/Air Leak3
- Pressure Problem2
- Unexpected Shutdown2
- Device Emits Odor1
- Display or Visual Feedback Problem1
- Electrical /Electronic Property Problem1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Patient Involvement6
- Nausea1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM), THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) UNIT DISPLAYS FAULT CODE #53. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17333921 ↗MalfunctionDec 20, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD A LOOP AIR LEAKAGE FAULT. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16014087 ↗InjuryAug 12, 2021Device Emits Odor; Appropriate Device Problem Term/Code Not Available; Device Emits Odor; Appropriate Device Problem Term/Code Not Available
MalfunctionAug 10, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem
MalfunctionAug 3, 2021Leak/Splash; Gas/Air Leak; Leak/Splash; Gas/Air Leak
IT WAS REPORTED THAT DURING USE ON A PATIENT, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) ALARMED "LEAK IN IAB CIRCUIT". NO PATIENT HARM, SERIOUS INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12267412 ↗MalfunctionMay 11, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
MalfunctionJan 22, 2021Gas/Air Leak; Gas/Air Leak
N/A.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11215976 ↗MalfunctionDec 23, 2020Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED BY THE DISTRIBUTOR THAT PRIOR TO USE, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) UNEXPECTEDLY SHUTDOWN AFTER THE UNIT WAS TURNED ON FOR LESS THAN ONE MINUTE. IT WAS ALSO REPORTED THAT THE CUSTOMER CHECKED TO ENSURE THAT THE POWER SUPPLY WAS CONNECTED AND ATTEMPTED TO RESTART THE UNIT. HOWEVER, THE REPORTED ISSUE PERSISTED AND THE CUSTOMER SWITCHED TO ANOTHER IABP UNIT. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 11067125 ↗MalfunctionDec 11, 2020Pressure Problem; Pressure Problem
IT WAS REPORTED THAT DURING A ROUTINE INSPECTION AND TESTING OF THE CS100 INTRA-AORTIC BALLOON PUMP (IABP), THE UNIT ALARMED AN ISSUE OF "NEGATIVE PRESSURE TO LOW." THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10993753 ↗MalfunctionNov 18, 2020Pressure Problem; Pressure Problem
IT WAS REPORTED THAT THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD A NEGATIVE PRESSURE TOO LOW MESSAGE. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10863339 ↗MalfunctionNov 5, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE DAY AFTER THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) WAS INSTALLED AN ERROR CODE # 50 FOR THE ELECTRICAL SAFETY TEST FAILURE OCCURRED. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10792651 ↗MalfunctionOct 28, 2020Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING A ROUTINE INSPECTION, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD AN AUTOMATIC INFLATION FAILURE. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10749892 ↗MalfunctionSep 28, 2020Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT BEFORE USE THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) AUTOMATICALLY SHUT DOWN AFTER THE INTERVENTIONAL CATHETERIZATION ROOM NURSE TURNED IT ON AND COULD NOT BE USED. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10592484 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K031636. One submission often covers many models, so check the product and lot details in each record.
- Z-1486-2026 · Jan 23, 2026 · Open, ClassifiedCS100 IABP. Software Version CS100 IABP Q.01.
- Z-3086-2017 · Jul 17, 2017 · Open, ClassifiedCS 100 Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by au
- Z-3085-2017 · Jul 17, 2017 · Open, ClassifiedCS 100i Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by a
- Z-2736-2017 · Jun 16, 2017 · TerminatedCS 100i Intra-Aortic Balloon Pump
- Z-2737-2017 · Jun 16, 2017 · TerminatedCS 100 Intra-Aortic Balloon Pump
- Z-1523-2014 · Mar 21, 2014 · TerminatedDatascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate in
- Z-2414-2011 · Mar 16, 2011 · TerminatedDatascope/ Maquet IABPs. System 98/98XT Intra-Aortic Balloon Pump; CS100/CS100i Intra-Aortic Balloon Pump; CS300 Intra-Aortic Balloon Pump; Manufactured by Datascope Corp., 130
- Z-1811-2010 · May 10, 2010 · TerminatedCS100 Intra-Aortic Balloon Pump.
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page