Skip to content

CS300 Intra-Aortic Balloon Pump, English, 220V 0998-00-3023-55Recalls & reported problems

Datascope Corp. · CS300 Intra-Aortic Balloon Pump, English, 220V · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Datascope CS300 intra-aortic balloon pump is the subject of 1,040 adverse event reports, which mention 2 deaths and 1 injury. Many reports describe the pump not powering up, battery problems, failed auto-fill and incorrect messages. Reports fell from 213 in the previous 12 months to 86 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,062
Deaths reported
2
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,062 adverse event reports that list this exact device (DI 10607567112541), received between Aug 26, 2020 and Sep 29, 2026 — classified as 2 deaths, 1 injury, 1,059 malfunctions.

The most frequently reported specific device problem is “Inability to Auto-Fill” (117 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 13 reportsDec 2023: 28 reportsJan 2024: 6 reportsFeb 2024: 20 reportsMar 2024: 11 reportsApr 2024: 17 reportsMay 2024: 22 reportsJun 2024: 19 reportsJul 2024: 21 reportsAug 2024: 30 reportsSep 2024: 24 reportsOct 2024: 32 reportsNov 2024: 21 reportsDec 2024: 13 reportsJan 2025: 33 reportsFeb 2025: 11 reportsMar 2025: 31 reportsApr 2025: 16 reportsMay 2025: 17 reportsJun 2025: 9 reportsJul 2025: 21 reportsAug 2025: 12 reportsSep 2025: 16 reportsOct 2025: 13 reportsNov 2025: 17 reportsDec 2025: 18 reportsJan 2026: 14 reportsFeb 2026: 7 reportsMar 2026: 16 reportsApr 2026: 8 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 6 reportsSep 2026: 19 reportsOct 2026: 0 reports20242025202633
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available127
  • Inability to Auto-Fill117
  • Battery Problem109
  • Failure to Power Up83
  • Device Displays Incorrect Message58
  • Electrical /Electronic Property Problem57
  • Incorrect, Inadequate or Imprecise Result or Readings56
  • Failure to Sense48

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,022
  • Insufficient Information131
  • No Patient Involvement15
  • No Known Impact Or Consequence To Patient7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
  • No Consequences Or Impact To Patient5
  • Low Blood Pressure/ Hypotension1

Recent adverse event reports

The 25 most recent of 1,062 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Noise, Audible

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928275 ↗
  • MalfunctionSep 28, 2026Power Problem

    It was reported by the customer that during routine check, the CS300 Intra-Aortic Balloon Pump (IABP) was not turning on. There was no patient involvement, no harm reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE checked the unit and found that the Power Supply was defective and needed to be replaced. After the Power Supply was replaced, the machine operated correctly. Product passed all functional and safety tests per manufacturer specifications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50912984 ↗
  • MalfunctionSep 28, 2026Calibration Problem; Unexpected Shutdown

    It was reported by the customer that during a routine check, the CS300 Intra-Aortic Balloon Pump (IABP) is displaying an error message "system failure". There was no patient involvement. No harm is reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE confirmed that they found the machine showing the "System failure" error. The FSE replaced the Main Board and then the device was showing Error Code 5 (Maintenance Required # 5 - Helium Pressure Transducer Out of Calibration). The required Helium Tank calibration was performed. Product passed all functional and safety tests per manufacturer specifications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50912905 ↗
  • MalfunctionSep 24, 2026Device Alarm System; Battery Problem; Power Problem

    It was reported by the customer that during routine check, the CS300 Intra-Aortic Balloon Pump (IABP) had no battery back up. There was no patient involvement, no harm reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE reported that unit drained batteries immediately, after replacing batteries additional low battery alarm was present. Power supply Board was found to be suspected and it needs to be replaced for further diagnostics.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50878046 ↗
  • MalfunctionSep 24, 2026Battery Problem; Power Problem

    It was reported by FSE that during Preventive Maintenance the CS300 Intra-Aortic Balloon Pump (IABP) malfunctioned. No patient involved. No harm reported. The FSE reported that machine is not working on battery. Battery was replaced. During evaluation it was also found out that power supply is defective and needs to be replaced.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50877585 ↗
  • MalfunctionSep 23, 2026Battery Problem

    It was reported by the customer that during a routine check, the CS300 Intra-Aortic Balloon Pump (IABP) displayed a ¿Low Battery¿ message and had insufficient battery backup. There was no patient involvement and no harm reported. A Getinge Field Service Engineer (FSE) evaluated the unit. During the evaluation, the device was tested after disconnecting it from AC power and was found to switch off within approximately one minute. The FSE determined that the battery was defective and was not providing adequate backup power. The battery requires replacement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867290 ↗
  • MalfunctionSep 23, 2026Inability to Auto-Fill

    It was reported by the FSE that during PM, the CS300 Intra-Aortic Balloon Pump (IABP) had Autofill Failure. There was no patient involved, no harm reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE confirmed the Autofill Failure in device logs. The device is currently pending a repair.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50866532 ↗
  • MalfunctionSep 21, 2026Electrical /Electronic Property Problem

    It was reported that during routine check by customer the CS300 Intra-Aortic Balloon Pump (IABP) malfunctioned. Electrical test fail code 50 has been reported by distributor's Field Service Engineer (FSE). No harm reported. No patient involvement reported. The Getinge FSE was dispatched to evaluate the unit. It was reported that motor control PCB was found to be defective and it was replaced. Electrical Test confirmed in logs attached. Safe for clinical use and can be released to the users. Product passed all functional and safety tests per manufacturer specifications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50835646 ↗
  • MalfunctionSep 18, 2026No Display/Image; Charging Problem; Power Problem

    It was reported by FSE that during machine repair, the CS300 Intra-Aortic Balloon Pump (IABP) power supply is faulty. The battery that was previously replaced for a customer was found to be faulty - battery is not charging. The device screen remained white all the time and could not display anything. There was no patient involvement. No harm was reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE reported that power supply was replaced. After that screen remain white and is not display anything. The machine is not functioning normally.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50814116 ↗
  • MalfunctionSep 17, 2026Gas/Air Leak

    It was reported by the customer that during a routine check the CS300 Intra-Aortic Balloon Pump (IABP) had a helium leak. No patient was involved. No harm was reported. Pending acceptance of repair from the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50800702 ↗
  • MalfunctionSep 17, 2026No Display/Image; Electrical /Electronic Property Problem

    It was reported by the customer that during routine check the CS300 Intra-Aortic Balloon Pump (IABP) failed electrical test #50 and display was not visible. No patient involved and no harm is reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE reported that they checked the machine and found the issue with Inverter PCB and Motor control PCB. The FSE replaced both the parts. Product passed all functional and safety tests per manufacturer specifications. Handed over the machine in good working condition.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50800143 ↗
  • MalfunctionSep 17, 2026Electrical /Electronic Property Problem; Failure to Read Input Signal

    It was reported that during a routine check the CS300 Intra-Aortic Balloon Pump (IABP) electrical test fails 39 and electrical test fails 65 occurred. No patient involved. No harm was reported. The issue discovered by the customer. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. The FSE confirm that the eletrical faile codes occurred. Board is suspected to be defective. After the replacement of main board, errors 39 and 65 no longer appeared. After further testing a machine exhibited A.P. Optical Sensing Module Failure. Fiber Optic Sensor Module (FOS) is suspected to be defective and needs to be replaced. After replacement of FOS module machine is working properly and was handed to user.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50800045 ↗
  • MalfunctionSep 17, 2026Alarm Not Visible

    It was reported by the Field Service Engineer that, the CS300 Intra-Aortic Balloon Pump (IABP) is showing battery backup low. There was no information about patient involvement. No harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50798463 ↗
  • MalfunctionSep 16, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50784934 ↗
  • MalfunctionSep 16, 2026Alarm Not Visible

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50783740 ↗
  • MalfunctionSep 2, 2026Inability to Auto-Fill; Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50567425 ↗
  • MalfunctionSep 2, 2026Device Contamination with Body Fluid

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50559082 ↗
  • MalfunctionSep 2, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50558915 ↗
  • MalfunctionSep 1, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50541178 ↗
  • MalfunctionAug 31, 2026Failure to Power Up

    It was reported by customer that during a routine check the CS300 Intra-Aortic Balloon Pump (IABP) malfunctioned. Device not turning on has been reported. There was no patient involvement and no harm was reported. A Getinge Field Service Engineer (FSE) was dispatched to evaluate the unit. During the evaluation, the FSE determined that no data was displayed on the screen and identified the power supply assembly as defective. The power supply assembly requires replacement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50518404 ↗
  • MalfunctionAug 24, 2026Failure to Power Up

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50403953 ↗
  • MalfunctionAug 23, 2026Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50393666 ↗
  • MalfunctionAug 19, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50323203 ↗
  • MalfunctionAug 4, 2026Failure to Deflate

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50125838 ↗
  • MalfunctionAug 3, 2026Battery Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50113666 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page