Skip to content

Sensation 7Fr 40cc IAB with Accessories (Turkey) 0684-00-0470-09Recalls & reported problems

Datascope Corp. · Sensation 7Fr 40cc IAB with Accessories (Turkey) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10607567113135), received between Jan 15, 2025 and Sep 10, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 23, 2023Class IIOpen, Classified

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

    Z-2217-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Advance1
  • Failure to Sense1
  • Material Deformation1
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2026Failure to Advance; Material Deformation

    It was reported that during use, while placing the Sensation 40cc balloon catheter, the user stated that the catheter failed to advance due to damage to the distal tip and that the procedure was continued with a new catheter. The IAB inserted prior to surgery could not be withdrawn when removal was attempted during the procedure. Examination after removal revealed damage to the distal tip. Surgical intervention was not required for the balloon catheter removal. No harm to the patient reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50710095 ↗
  • MalfunctionJan 15, 2025Overheating of Device; Failure to Sense; Overheating of Device; Failure to Sense

    IT WAS REPORTED THAT DURING INTRA AORTIC BALLOON THERAPY (IAB) "FIBER OPTIC SENSOR FAILURE" WARNING APPEARED ON THE SCREEN. THE USER'S REASON FOR CALLING WAS THE "NO PRESSURE SOURCE AVAILABLE" WARNING. THERE WAS NO PATIENT HARM/INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21156927 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page