Skip to content

Deltec 21-4052-24Recalls & reported problems

Icu Medical, Inc. · Class 2 · Implant

All models of Deltec (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10610586023521), received between Nov 5, 2024 and Dec 12, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K830730); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Air/Gas in Device2
  • Crack1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2024Crack; Air/Gas in Device; Crack; Air/Gas in Device

    IT WAS REPORTED THAT THE IRRIGATION CANNULA WAS CONNECTED TO THE CATHETER, AND WHEN ASPIRATION OCCURRED, AIR CAME BACK INSTEAD OF LIQUID. THEY NOTICED THAT THE END OF THE CATHETER WAS CRACKED. THE DEVICE WAS OPERATED BY THE PHYSICIAN. THE EVENT OCCURRED INTRAOPERATIVELY; THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20915341 ↗
  • MalfunctionNov 5, 2024Material Split, Cut or Torn; Air/Gas in Device; Material Split, Cut or Torn; Air/Gas in Device

    IT WAS REPORTED THAT WHEN THE CATHETER WAS INSERTED AND ASPIRATION WAS ATTEMPTED, AIR RESURFACED INSTEAD OF BLOOD. IT WAS NOTICED THAT THE COATING OF THE CATHETER WAS TORN. A NEW CATHETER WAS USED AND THERE WERE NO NEGATIVE CONSEQUENCES FOR THE PATIENT. THE EVENT OCCURRED IN THE HOSPITAL, WHILE IN USE WITH THE PATIENT. THE OPERATOR OF THE DEVICE WAS A HEALTHCARE PROFESSIONAL. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20613318 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K830730. One submission often covers many models, so check the product and lot details in each record.

  • Z-0152-2016 · Sep 14, 2015 · Terminated
    PORT-A-CATH Plastic Hub Needles, 0.9mm (20G) x 19mm (3/4 in.) 90-degree. REF 21-2287-24 The Plastic hub PORT-A-CATH (PAC) needles are intended to access the implantable portal sy

Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices

All PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page