Deltec 21-4453-24Recalls & reported problems

Icu Medical, Inc. · Class 2 · Implant

All models of Deltec (97)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10610586032684), received between Sep 3, 2020 and Jan 24, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K072657); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • Leak/Splash1
  • Material Fragmentation1
  • Material Puncture/Hole1
  • Material Split, Cut or Torn1
  • Mechanical Problem1

Patient problems

  • Extravasation2
  • No Clinical Signs, Symptoms or Conditions2
  • Necrosis1
  • Patient Problem/Medical Problem1
  • Tissue Breakdown1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 24, 2025Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    IT WAS REPORTED THAT DURING THE OPERATION IT WAS DISCOVERED THAT THE CATHETER HAD A SMALL HOLE AT THE TRANSITION TO THE VENTRICLE. THERE WAS PATIENT INVOLVEMENT. THE PATIENT DETAILS: 57 YEARS OLD, GENDER: MALE. IT WAS REPORTED THAT THE PATIENT EXPERIENCED EXTRAVASATION OF CHEMOTHERAPEUTIC AGENT (OXALIPLATIN) INTO THE RIGHT DELTOIDEOPECTORAL REGION. THERE WAS PORT EXPLANTATION WITH LAVAGE ON (B)(6) 2024 WITH INTENDED SECONDARY WOUND HEALING. THE PATIENT ALSO EXPERIENCED TISSUE NECROSIS CAUSED BY THE LONG EFFECT OF THE TOXICITY OF THE CHEMOTHERAPY DRUG.

    Patient problems: Extravasation; Necrosis; Tissue Breakdown

    MAUDE report 21229476 ↗
  • InjuryNov 4, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE TUBE FROM THE CATHETER TORE OFF RIGHT AT THE CONNECTION POINT. THE HOSE WAS INSERTED CORRECTLY UP TO THE STOP AND IS CLEARLY VISIBLE THROUGH THE SMALL HOLE. THE CATHETER HAD TO BE RETRIEVED VIA THE GROIN, WHICH REQUIRED A PARTIAL HOSPITALIZATION. THERE WAS PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20600131 ↗
  • InjuryMay 9, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING REMOVAL, THE CATHETER BROKE OFF FROM THE CHAMBER AND MOVED TOWARDS THE HEART. THAT SAME DAY, THE PATIENT WAS TRANSFERRED TO A CLINIC FOR A MINIMALLY INVASIVE CATHETER INTERVENTION (MIKI) TO REMOVE THE FOREIGN BODY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16898662 ↗
  • MalfunctionSep 3, 2020Leak/Splash; Mechanical Problem; Leak/Splash; Mechanical Problem

    IT WAS REPORTED THAT THE DELTEC POWER PORT-A-CATH II WAS SET ON THE 22-JULY. THERE WAS NO INDICATION OF A CONTRAST MEDIUM EXTRAVASATION IN THE C-ARM. THIS WAS NOT OBSERVED ON 22-JULY. ON 23-JULY CHEMOTHERAPY (EPIRUBICIN) WAS STARTED. THE PATIENT COMPLAINED OF A PRONOUNCED PAIN IN THE PORT AREA. A CLEAR EXTRAVASATION WAS FOUND DURING A FUNCTIONAL TEST WITH A MOBILE X-RAY. THE PORT WAS REMOVED ON THE SAME DAY AS A RESULT OF THIS. NO FURTHER PATIENT COMPLICATIONS WERE NOTED. THERE WAS NO DISCOLORATION OF THE CATHETER, HOWEVER A LEAK WAS VERIFIED ON THE CATHETER JUST TWO CM DISTAL TO THE AFOREMENTIONED SITE. THE OPERATOR NOTED A SLIT-SHAPED LESION IN THE CATHETER. THE DEFECTIVE PRODUCT WAS RETURNED TO THE MANUFACTURER FOR FURTHER DEVICE INVESTIGATION.

    Patient problems: Extravasation; Patient Problem/Medical Problem

    MAUDE report 10488246 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K072657. One submission often covers many models, so check the product and lot details in each record.

  • Z-2499-2020 · Jun 11, 2020 · Terminated
    PORT-A-CATH II POWER P.A.C. Low Profile" system.
  • Z-0190-2020 · Jul 8, 2019 · Terminated
    GRIPPER Needles sold as part of PORT-A-CATH II SINGLE LUMEN SYSTEM. Labeled as the following item descriptions: 1. TRAY, PAC, VEN, TI, PU, 1.9MM, UL, 6FR/INTRO 1/EA;

Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices

All PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices →