Deltec 21-4453-24Recalls & reported problems
Icu Medical, Inc. · Class 2 · Implant
All models of Deltec (97)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10610586032684), received between Sep 3, 2020 and Jan 24, 2025 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K072657); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak1
- Leak/Splash1
- Material Fragmentation1
- Material Puncture/Hole1
- Material Split, Cut or Torn1
- Mechanical Problem1
Patient problems
- Extravasation2
- No Clinical Signs, Symptoms or Conditions2
- Necrosis1
- Patient Problem/Medical Problem1
- Tissue Breakdown1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJan 24, 2025Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT DURING THE OPERATION IT WAS DISCOVERED THAT THE CATHETER HAD A SMALL HOLE AT THE TRANSITION TO THE VENTRICLE. THERE WAS PATIENT INVOLVEMENT. THE PATIENT DETAILS: 57 YEARS OLD, GENDER: MALE. IT WAS REPORTED THAT THE PATIENT EXPERIENCED EXTRAVASATION OF CHEMOTHERAPEUTIC AGENT (OXALIPLATIN) INTO THE RIGHT DELTOIDEOPECTORAL REGION. THERE WAS PORT EXPLANTATION WITH LAVAGE ON (B)(6) 2024 WITH INTENDED SECONDARY WOUND HEALING. THE PATIENT ALSO EXPERIENCED TISSUE NECROSIS CAUSED BY THE LONG EFFECT OF THE TOXICITY OF THE CHEMOTHERAPY DRUG.
Patient problems: Extravasation; Necrosis; Tissue Breakdown
MAUDE report 21229476 ↗InjuryNov 4, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE TUBE FROM THE CATHETER TORE OFF RIGHT AT THE CONNECTION POINT. THE HOSE WAS INSERTED CORRECTLY UP TO THE STOP AND IS CLEARLY VISIBLE THROUGH THE SMALL HOLE. THE CATHETER HAD TO BE RETRIEVED VIA THE GROIN, WHICH REQUIRED A PARTIAL HOSPITALIZATION. THERE WAS PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20600131 ↗InjuryMay 9, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING REMOVAL, THE CATHETER BROKE OFF FROM THE CHAMBER AND MOVED TOWARDS THE HEART. THAT SAME DAY, THE PATIENT WAS TRANSFERRED TO A CLINIC FOR A MINIMALLY INVASIVE CATHETER INTERVENTION (MIKI) TO REMOVE THE FOREIGN BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16898662 ↗MalfunctionSep 3, 2020Leak/Splash; Mechanical Problem; Leak/Splash; Mechanical Problem
IT WAS REPORTED THAT THE DELTEC POWER PORT-A-CATH II WAS SET ON THE 22-JULY. THERE WAS NO INDICATION OF A CONTRAST MEDIUM EXTRAVASATION IN THE C-ARM. THIS WAS NOT OBSERVED ON 22-JULY. ON 23-JULY CHEMOTHERAPY (EPIRUBICIN) WAS STARTED. THE PATIENT COMPLAINED OF A PRONOUNCED PAIN IN THE PORT AREA. A CLEAR EXTRAVASATION WAS FOUND DURING A FUNCTIONAL TEST WITH A MOBILE X-RAY. THE PORT WAS REMOVED ON THE SAME DAY AS A RESULT OF THIS. NO FURTHER PATIENT COMPLICATIONS WERE NOTED. THERE WAS NO DISCOLORATION OF THE CATHETER, HOWEVER A LEAK WAS VERIFIED ON THE CATHETER JUST TWO CM DISTAL TO THE AFOREMENTIONED SITE. THE OPERATOR NOTED A SLIT-SHAPED LESION IN THE CATHETER. THE DEFECTIVE PRODUCT WAS RETURNED TO THE MANUFACTURER FOR FURTHER DEVICE INVESTIGATION.
Patient problems: Extravasation; Patient Problem/Medical Problem
MAUDE report 10488246 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K072657. One submission often covers many models, so check the product and lot details in each record.
- Z-2499-2020 · Jun 11, 2020 · TerminatedPORT-A-CATH II POWER P.A.C. Low Profile" system.
- Z-0190-2020 · Jul 8, 2019 · TerminatedGRIPPER Needles sold as part of PORT-A-CATH II SINGLE LUMEN SYSTEM. Labeled as the following item descriptions: 1. TRAY, PAC, VEN, TI, PU, 1.9MM, UL, 6FR/INTRO 1/EA;
Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices
- PowerPort M.R.I., Groshong 9808560Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR3 deaths834 reports
- PowerPort ClearVUE 1608062Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR18 deaths571 reports
- PowerPort ClearVUE 5608062Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR31 deaths542 reports
- PowerPort M.R.I., ChronoFlex 1808000Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR19 deaths513 reports
- PowerPort M.R.I., ChronoFlex 1808060Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall15 deaths383 reports
- PowerPort M.R.I., Groshong 1808560Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR10 deaths304 reports
- PowerPort M.R.I., ChronoFlex 1829500Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR6 deaths197 reports
- BardPort M.R.I. 0602680Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR2 deaths179 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026