Deltec 21-4471-24Recalls & reported problems
Icu Medical, Inc. · Class 2 · Implant
All models of Deltec (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10610586032745), received between Dec 12, 2017 and Jun 23, 2026 — classified as 7 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K072657); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak4
- Leak/Splash2
- Material Puncture/Hole2
- Break1
- Contamination1
- Defective Component1
- Fracture1
- Hole In Material1
Patient problems
- Pain5
- No Clinical Signs, Symptoms or Conditions3
- Burning Sensation1
- Foreign Body In Patient1
- Infiltration into Tissue1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 23, 2026Contamination
IT WAS REPORTED THAT DURING NEEDLE INSERTION, THE PORT-A-CATH IMPLANTABLE ACCESS PORT, A WHITE FLAKE WAS NOTICED DURING ASPIRATION ON THE VENOUS PORT AND NO LONGER USED. THE PATIENT REQUIRED REPLACEMENT OF THE VENOUS PORT AND UNDERWENT ADDITIONAL SURGERY IN WHICH THESE FLAKES WERE FOUND. THERE WAS PATIENT INVOLVEMENT; HOWEVER, UNKNOWN HARM/ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25598031 ↗MalfunctionApr 13, 2026Material Separation
IT WAS REPORTED THAT THE CATHETER RUPTURED, AND A FRAGMENT MIGRATED AND BECAME LODGED IN THE ATRIUM OF THE HEART. AS A RESULT, THE PATIENT REQUIRED TRANSFER TO A CITY HOSPITAL FOR REMOVAL OF THE RETAINED CATHETER FRAGMENT. THE TREATING FACILITY EMPLOYS EXPERIENCED SURGEONS AND HAD NOT PREVIOUSLY ENCOUNTERED THIS TYPE OF COMPLICATION. THE RUPTURED CATHETER WAS REMOVED ON MARCH 24. THE CATHETER HAD ORIGINALLY BEEN IMPLANTED ON NOVEMBER 15, 2024.THERE WAS PATIENT INVOLVEMENT AND PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24860653 ↗InjuryOct 28, 2025Defective Component; Defective Component
IT WAS STATED THAT THE PATIENT HAS A BROKEN FRAGMENT OF PORT-A-CATH ON THE LEFT SIDE OF THE CHEST. DURING AN IMAGING TEST, IT WAS FOUND THAT THE TIP OF THE PORT-A-CATH IS IN AN EXTRAVASCULAR POSITION AND IS SHORT (10 CM INSTEAD OF THE USUAL 20 CM). REMOVAL OF THE PORT-A-CATH WAS ATTEMPTED IN THE OPERATING ROOM AND THE REMOVAL WAS INCOMPLETE AS THE DISTAL END WAS NOT RETRIEVED. A SCAN CONFIRMED THAT THE CATHETER TIP IS POSITIONED IN THE PATIENT'S RIGHT VENTRICLE FOR WHICH A HEMODYNAMIC ASSESSMENT AND A POSSIBLE CARDIAC SURGERY WILL BE DONE TO REMOVE THE REMAINING PORTION OF THE CATHETER. THERE WAS PATIENT INVOLVEMENT AND PATIENT HARM.
Patient problems: Foreign Body In Patient
MAUDE report 23402821 ↗MalfunctionJul 7, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE CATHETER RUPTURED INSIDE THE PATIENT. THE DISTAL PART WAS REMOVED FROM THE PATIENT WITH ADDITIONAL INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22412414 ↗MalfunctionNov 27, 2019Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
INFORMATION WAS RECEIVED INDICATING THAT A 7 CM LEAK OCCURRED TO A SMITHS MEDICAL DELTEC® PORT-A-CATH®II IMPLANTABLE ACCESS SYSTEM. IT WAS REPORTED THAT THE PATIENT HAD BEEN EXPERIENCING SHOULD PAIN. AN X-RAY WAS ORDERED AND REVEALED A LEAK UNDER THE CORRESPONDING CLAVICLE. THE PORT WAS EXPLANTED AND REPLACED. THE PHYSICIAN REPORTED THAT A HOLE IS MADE NEXT TO THE PORT ENABLE TO GET THE GUIDEWIRE IN FOR INSERTION. THERE WERE NO FURTHER REPORTED ADVERSE PATIENT EFFECTS.
Patient problems: Pain
MAUDE report 9386749 ↗InjurySep 24, 2019Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE DEVICE WAS CRACKED IN TWO PLACES. THIS ISSUE WAS CONFIRMED VIA A LINOGRAM. THE PATIENT EXPERIENCED INFILTRATION ABOVE THE PORT AS RESULT OF THE PRODUCT PROBLEM. THE PATIENT REPORTED FEELING A BURNING SENSATION. THE PORT WAS SUBSEQUENTLY EXPLANTED ON (B)(6) 2019. IT WAS ALSO REPORTED THAT THIS INCIDENT REQUIRED THE INSERTION OF A PICC LINE IN ORDER TO CONTINUE TREATMENT. THE PATIENT'S TREATMENT WAS DELAYED BY ONE DAY AS A RESULT OF THIS. NO PATIENT INJURY OR FURTHER COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT.
Patient problems: Burning Sensation
MAUDE report 9112164 ↗InjurySep 13, 2019Leak/Splash; Leak/Splash
INFORMATION WAS RECEIVED THAT WHILE A SMITHS MEDICAL IMPLANTABLE PORT WAS IN USE, NO REFLUX OCCURRED. XRAY HAS SHOWN A LEFT-SIDED, DYSFUNCTIONAL PORT WITH A SMALL DEFECT AT THE PROXIMAL END OF THE CATHETER. LEAKAGE OCCURRED TO SURROUNDING TISSUE WHERE CATHETER RUNS UNDER THE CLAVICLE. AS A RESULT, THE DEVICE WAS THEN REMOVED.
Patient problems: Infiltration into Tissue
MAUDE report 9037967 ↗InjuryFeb 16, 2018Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE EXPLANTING DOCTOR NOTICED HOLES IN THE CATHETER OF A DELTEC® PORT-A-CATH®II POWER P.A.C. IMPLANTABLE ACCESS SYSTEM. ISSUES WITH THE CATHETER HAD TO BE EXPLANTED WHEN ONE PATIENT COMPLAINED OF PAIN. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS OCCURRENCE.
Patient problems: Pain
MAUDE report 7278792 ↗InjuryFeb 16, 2018Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE EXPLANTING DOCTOR NOTICED HOLES IN THE CATHETER OF A DELTEC® PORT-A-CATH®II POWER P.A.C. IMPLANTABLE ACCESS SYSTEM. ISSUES WITH THE CATHETER HAD TO BE EXPLANTED WHEN ONE PATIENT COMPLAINED OF PAIN. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS OCCURRENCE.
Patient problems: Pain
MAUDE report 7278791 ↗MalfunctionFeb 16, 2018Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT THE EXPLANTING DOCTOR NOTICED HOLES IN THE CATHETER OF A DELTEC® PORT-A-CATH®II POWER P.A.C. IMPLANTABLE ACCESS SYSTEM. ISSUES WITH THE CATHETER HAD TO BE EXPLANTED WHEN ONE PATIENT COMPLAINED OF PAIN. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS OCCURRENCE.
Patient problems: Pain
MAUDE report 7278787 ↗InjuryFeb 16, 2018Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE EXPLANTING DOCTOR NOTICED HOLES IN THE CATHETER OF A DELTEC® PORT-A-CATH®II POWER P.A.C. IMPLANTABLE ACCESS SYSTEM. ISSUES WITH THE CATHETER HAD TO BE EXPLANTED WHEN ONE PATIENT COMPLAINED OF PAIN. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS OCCURRENCE.
Patient problems: Pain
MAUDE report 7278786 ↗InjuryDec 12, 2017Hole In Material; Hole In Material
INFORMATION WAS RECEIVED INDICATING THAT AT THE END OF THERAPY THE EXPLANTING DOCTOR NOTICED HOLES IN THE CATHETER OF THIS PORT. THE PORT HAD TO BE EXPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7110175 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K072657. One submission often covers many models, so check the product and lot details in each record.
- Z-2499-2020 · Jun 11, 2020 · TerminatedPORT-A-CATH II POWER P.A.C. Low Profile" system.
- Z-0190-2020 · Jul 8, 2019 · TerminatedGRIPPER Needles sold as part of PORT-A-CATH II SINGLE LUMEN SYSTEM. Labeled as the following item descriptions: 1. TRAY, PAC, VEN, TI, PU, 1.9MM, UL, 6FR/INTRO 1/EA; 2.
Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices
- PowerPort M.R.I., Groshong 9808560Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR3 deaths834 reports
- PowerPort ClearVUE 1608062Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR18 deaths571 reports
- PowerPort ClearVUE 5608062Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR31 deaths542 reports
- PowerPort M.R.I., ChronoFlex 1808000Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR19 deaths513 reports
- PowerPort M.R.I., ChronoFlex 1808060Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall15 deaths383 reports
- PowerPort M.R.I., Groshong 1808560Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR10 deaths304 reports
- PowerPort M.R.I., ChronoFlex 1829500Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR6 deaths197 reports
- BardPort M.R.I. 0602680Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR2 deaths179 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page