Skip to content

Norm-O-Temp® 111ZRecalls & reported problems

Gentherm Medical, LLC · Total Body Hyperthermia System · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10613031861549), received between Apr 19, 2022 and Mar 14, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 20, 2019Terminated

    The manual does not provide instruction stating that exceeding 40 Celsius for extended periods can cause tissue damage.

    Z-2355-2019 · Root cause: Labeling False and Misleading · FDA record ↗

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1
  • Leak/Splash1
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 14, 2025Leak/Splash; Leak/Splash

    CUSTOMER REPORTS THAT THIS DEVICE HAS A LEAK. ADDITIONAL INFORMATION: DURING THE SERVICE EVALUATION, IT WAS OBSERVED THAT THERE WAS A HOLE IN THE RESERVOIR DUE TO THERMAL DECOMPOSITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21605093 ↗
  • MalfunctionApr 19, 2022Overheating of Device; Improper or Incorrect Procedure or Method

    THE CUSTOMER STATES THAT THE UNIT OVERHEATED. THE UNIT WAS IN THE OPERATING ROOM, PREPPING FOR A PAITENT BUT WAS NEVER USED FOR THE PAITENT. THE UNIT WAS LEFT ON AND RUNNING OVERNIGHT WITHOUT OVERSIGHT. THE CUSTOMER NOTICED THAT THE UNIT HAD CAUSED DAMAGE TO ITSELF AND TO OBJECTS AROUND THE DEVICE. THE CUSTOMER CLAIMS THAT THEY DID NOT DISCOVER THE UNIT MALFUNCTION UNTIL THE NEXT DAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14141356 ↗

Other System, Thermal Regulating devices

All System, Thermal Regulating devices →