Halyard 13471Recalls & reported problems
Avanos Medical, Inc. · HALYARD* ON-Q* Pump with ONDEMAND* · Class 2
All models of Halyard (232)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10680651134711), received between Jun 21, 2017 and Jun 21, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Mechanical Jam” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K063530); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Mechanical Jam1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 21, 2017Mechanical Jam; Mechanical Jam
FILL VOLUME: 400 ML. FLOW RATE: UNKNOWN. PROCEDURE: R ANKLE TENDON REPAIR. CATHPLACE: UNKNOWN. IT WAS REPORT THAT A PATIENT EXPERIENCED AN ON DEMAND BOLUS BUTTON THAT WAS ABLE TO BE PRESSED DOWN, BUT CAME RIGHT BACK UP. THE ORANGE INDICATOR ALSO REMAINED AT THE BOTTOM. THE PATIENT DID NOT EXPERIENCE ANY SIDE EFFECTS. HOWEVER, THE PUMP WAS CLAMPED IMMEDIATELY, AND THE PATIENT'S WIFE WAS INSTRUCTED TO CONTACT THEIR ANESTHESIOLOGIST FOR FURTHER INSTRUCTIONS. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6657827 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K063530. One submission often covers many models, so check the product and lot details in each record.
- Z-2860-2011 · Apr 15, 2011 · TerminatedI-Flow ON-Q with Select-A-Flow Pumps, Intended for continuous and/or intermittent delivery of medication to and around surgical wound sites and/or close proximity to nerves for p
- Z-2133-2008 · Feb 14, 2008 · TerminatedOn-Q C-bloc 400 ml, 1-7 ml/hr Select-A-Flow + 5 ml/60 min ONDEMAND, REF CB005 indicated for continuous and/or intermittent delivery of medication (such as local anesthetics or narc
- Z-1869-2008 · Feb 12, 2008 · TerminatedON-Q PainBuster Infusion Pump (270 ml, 5 ml/hr), REF PS6504, Part No. 5001260. The pump is indicated for continuous and/or intermittent delivery of medication (such as local anest
Other PUMP, INFUSION, ELASTOMERIC devices
- EASYPUMP 4540018-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC313 reports
- Halyard 13568Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC86 reports
- ON-Q* Pain Relief System with SELECT-A-FLOW 13472Avanos Medical, Inc. · Pump, infusion, elastomeric73 reports
- Halyard 13472Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC68 reports
- EASYPUMP 4540008-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC54 reports
- DOSI-FUSER L25915-600D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- DOSI-FUSER L25915-500D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- Halyard 13561Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC42 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page