Halyard 44566Recalls & reported problems
Avanos Medical, Inc. · MIC* Jejunal Feeding Tube with ENFit® Connector, 12 Fr · Class 2 · Implant
All models of Halyard (232)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10680651445664), received between Jan 29, 2020 and Jan 29, 2020 — classified as 1 death.
The most frequently reported specific device problem is “Malposition of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Malposition of Device1
Patient problems
- Aspiration/Inhalation1
- Death1
- No Code Available1
- Obstruction/Occlusion1
- Pneumonia1
- Sepsis1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathJan 29, 2020Malposition of Device; Malposition of Device
ADDITIONAL INFORMATION RECEIVED 04-FEB-2020 VIA MEDWATCH/ FDA USER FACILITY REPORT # MW (B)(4) NOTED: ELDERLY FEMALE PATIENT WITH UNRESECTABLE ESOPHAGEAL CANCER AND DYSPHAGIA UNDERWENT ELECTIVE FEEDING JEJUNOSTOMY TUBE REPLACEMENT. AFTERWARDS, SHE DEVELOPED SBO INITIALLY THOUGHT TO BE RELATED TO A TWIST IN THE SMALL BOWEL-THIS WAS SURGICALLY CORRECTED. AFTER THAT, HER SBO FAILED TO RESOLVE, AND REPEAT IMAGING SHOWED THAT THE 3ML BALLOON OF THE J-TUBE TO BE THE POINT OF OBSTRUCTION. THE BALLOON WAS DEFLATED, BUT BY THAT TIME SHE HAD ALREADY ASPIRATED, DEVELOPED PNEUMONIA, THEN SEPSIS AND DEATH. DOCTOR'S RECOMMENDATION IS TO FIND A DIFFERENT J-TUBE THAT DOES NOT HAVE A BALLOON. CURRENTLY THE MANUFACTURER DOES NOT OFFER THE FEEDING J-TUBE WITHOUT THE BALLOON. IT WAS THOUGHT THAT THE BALLOON WAS NOT LARGE ENOUGH TO CAUSE A SMALL BOWL OBSTRUCTION, BUT THIS IS A SENTINEL CASE THAT PROVES THE BALLOON IS LARGE ENOUGH TO CAUSE OBSTRUCTION.
Patient problems: Aspiration/Inhalation; Death; No Code Available; Obstruction/Occlusion; Pneumonia; Sepsis
MAUDE report 9641390 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗