Skip to content

Halyard 45758Recalls & reported problems

Avanos Medical, Inc. · HALYARD* CORFLO* Nasogastric/Nasointestinal Feeding Tube with ANTI-IV* Connector and Stylet · Class 2

All models of Halyard (232)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10680651457582), received between Jul 23, 2019 and Oct 21, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Difficult to Remove2
  • Break1
  • Flushing Problem1

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2019Break; Difficult to Remove

    AN ELDERLY FEMALE PATIENT WITH A HISTORY OF OBSTRUCTIVE LUNG DISEASE REQUIRED DOBHOFF FEEDING TUBE INSERTION PER PROTOCOL. UNABLE TO REMOVE GUIDEWIRE. THE END PLASTIC PIECE ON THE GUIDEWIRE BROKE OFF, SO THE ENTIRE TUBE HAD TO BE REMOVED AND REPLACED. TUBE AND PACKAGING WERE KEPT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9213486 ↗
  • MalfunctionJul 23, 2019Flushing Problem; Difficult to Remove

    SINCE A FEW MONTHS AGO, STAFF REPORTING ISSUES THAT AFTER FEEDING TUBE INSERTION AND X-RAY CONFIRMATION OF PLACEMENT, THE GUIDE WIRE IS UNABLE TO BE REMOVED. STAFF ALSO REPORTING THAT THEY ARE ABLE TO FLUSH TUBE PRIOR TO INSERTION BUT NOT AFTER INSERTION. NO CHANGE IN PRACTICE OR PROCESS. THIS HAS OCCURRED AT LEAST 10 TIMES, 7 THAT WE AREA ABLE TO CONFIRM WITH THE SAME LOT NUMBER. UPON INVESTIGATION, STAFF REPORTING 6-8 TIMES LAST MONTH THAT THEY EXPERIENCED DIFFICULTY, BUT THOUGHT IT WAS THEIR INSERTION TECHNIQUES. THIS HAS OCCURRED IN OUR HOSPITAL AND IN OUR LONG-TERM ACUTE CARE HOSPITAL (LTACH) UNIT. WE HAVE ABOUT 4 TUBES WITH PACKAGING THAT WE WILL BE RETURNING TO THE COMPANY. NO PATIENTS HAVE BEEN DIRECTLY INJURED IN THESE INCIDENTS: HOWEVER, THEY HAVE HAD TO HAVE THE TUBES REMOVED, REPLACED, HAVE ANOTHER X-RAY TO VERIFY PLACEMENT. THIS HAS CAUSED SOME DELAY TO INITIATION OF MEDICATIONS AND OR FEEDINGS, BUT AGAIN, NO HARM OR ADVERSE OUTCOME TO PATIENTS. WE REPORTED A SIMILAR EVENT ABOUT A YEAR AGO. THE DEVICE INVOLVED IN THIS REPORT WAS A HALYARD PRODUCT, NOT A MEDTRONIC PRODUCT AS PREVIOUSLY DESCRIBED. MANUFACTURER RESPONSE FOR TUBE, FEEDING, HALYARD (PER SITE REPORTER). THE

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8815329 ↗

FDA clearance and classification

The FDA regulates product code FPD (tube, feeding) under 21 CFR 876.5980 as a Class II device. 21 CFR 876.5980 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Halyard 45758

Has the Halyard 45758 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10680651457582), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Halyard 45758?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jul 23, 2019 and Oct 21, 2019. In those reports, the most common specific device problems are “Difficult to Remove” (2), “Break” (1) and “Flushing Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Halyard 45758?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Halyard 45758, and is it still sold?

The FDA's device database lists Avanos Medical, Inc. as the company responsible for it (brand Halyard, model 45758). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TUBE, FEEDING devices

All TUBE, FEEDING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page