Skip to content

Halyard 46017Recalls & reported problems

Avanos Medical, Inc. · HALYARD* CORFLO* Nasogastric/Nasointestinal Feeding Tube with ENFit® Connector and Stylet · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10680651460179), received between Dec 13, 2019 and Dec 13, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Physical Resistance/Sticking2

Patient problems

  • No Consequences Or Impact To Patient1
  • Sedation1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT EVENTS. THIS IS THE SECOND OF TWO REPORTS. REFER TO 3011270181-2019-00071 FOR THE SECOND EVENT. IT WAS REPORTED WHEN PHYSICIAN TRIED TO REMOVE THE DEVICE, THE TUBE SEEMS TO GET STUCK IN THE SEPTUM. THE PHYSICIAN STATED AS HE WAS REMOVING THE DEVICE, HE ENCOUNTERED RESISTANCE, STOPPED PULLING, AND THEN THE NURSE TRIED. IT WAS JUST STUCK. THE PHYSICIAN HAD TO USE A GLIDE SCOPE AND PULLED THE DEVICE OUT THROUGH THE PATIENT'S MOUTH. THERE WAS NO REPORTED INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9465813 ↗
  • MalfunctionDec 13, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT EVENTS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 3011270181-2019-00072 FOR THE SECOND EVENT. IT WAS REPORTED WHEN PHYSICIAN TRIED TO REMOVE THE DEVICE, THE TUBE SEEMS TO GET STUCK IN THE SEPTUM. THE PHYSICIAN HAD TO SEDATE THE PATIENT AND REMOVE THEM FROM THE MOUTH. THE PHYSICIAN STATED AS HE WAS REMOVING THE DEVICE, HE ENCOUNTERED RESISTANCE, STOPPED PULLING, AND THEN THE NURSE TRIED. IT WAS JUST STUCK. THE PHYSICIAN THEN HAD TO SEDATE THE PATIENT, USE A GLIDE SCOPE AND PULLED THE DEVICE OUT THROUGH THE PATIENT'S MOUTH. THERE WAS NO REPORTED INJURY.

    Patient problems: Sedation

    MAUDE report 9465772 ↗

Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices

All TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page