Halyard 46094Recalls & reported problems
Avanos Medical, Inc. · HALYARD* CORTRAK* 2 Nasogastric/Nasointestinal Feeding Tube with ENFit® Connector and Electromagnetic Transmitting Stylet · Class 2
All models of Halyard (232)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10680651460940), received between Oct 28, 2020 and Feb 14, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Malposition of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Malposition of Device2
- Difficult to Insert1
- Material Puncture/Hole1
- Sharp Edges1
Patient problems
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
- Perforation1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 14, 2024Malposition of Device; Malposition of Device
IT WAS REPORTED, ¿INCORRECT INSERTION INTO LUNG OCCURRED. VERY SMALL AMOUNT OF NUTRITION WAS MIS-ADMINISTERED INTO LUNG. THE MONITOR SHOWED THE INCORRECT INSERTION, BUT IT WAS NOT NOTICED DURING THE INSERTION AND ALSO FROM X-RAY DATA AFTER INSERTION.¿ NO MEDICAL INTERVENTIONS REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 18705519 ↗InjuryFeb 4, 2021Malposition of Device; Sharp Edges; Malposition of Device; Sharp Edges
IT WAS REPORTED THAT TWO GASTRIC MUCOSA PERFORATIONS OCCURRED DURING INSERTION OF THE NG TUBE. THE PERFORATIONS WERE LOCATED "AROUND UPPER STOMACH POSTERIOR WALL" AND WERE REPAIRED BY GASTRIC CLIPS. THE NG TUBE WAS REMOVED BY ENDOSCOPE. IT WAS NOTED THAT "DURING THE INSERTION, PASSING CARDIAC ORIFICE WAS DIFFICULT BUT LATER ON IT WENT THROUGH AND ADVANCE TO JEJUNUM." THE DOCTOR POINTED OUT THE TUBE HAD A "SLIGHTLY SHARP END." ONE WEEK AFTER THE EVENT, AN ALTERNATIVE TUBE WAS INSERTED UNDER ENDOSCOPE AND THE TWO PERFORATIONS WERE NOTED TO HAVE "RECOVERED." ADDITIONAL INFORMATION RECEIVED VIA INTERVIEW WITH A PHYSICIAN ON 29-JAN-2021 INCLUDED PATIENT INFORMATION AND THAT THE EVENT WAS "MOST LIKELY DUE TO PATIENT ANATOMICAL FEATURES."
Patient problems: Perforation
MAUDE report 11278153 ↗MalfunctionOct 28, 2020Difficult to Insert; Material Puncture/Hole; Difficult to Insert; Material Puncture/Hole
IT WAS REPORTED THAT DURING INSERTION OF THE NASOGASTRIC (NG) TUBE, "RESISTANCE WAS FELT AT AROUND CARDIA AND IT WAS DIFFICULT TO INSERT THE TUBE. THE USER TRIED TO MOVE THE TUBE IN AND OUT A FEW TIMES BENDING THE TUBE. THE TUBE WAS TORN AND STYLET CAME OUT FROM THE TEAR ON THE TUBE WHILE THE TUBE WAS PLACED IN THE PATIENT. NO PATIENT INJURY REPORTED." PER ADDITIONAL INFORMATION RECEIVED ON 13 OCT 2020, THE TUBE WAS NOT FLUSHED WITH WATER WHEN REMOVING THE STYLET. NO FURTHER INFORMATION HAS BEEN PROVIDED AT THIS TIME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10751696 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page