Halyard 95649Recalls & reported problems
Avanos Medical, Inc. · HALYARD* MIC-KEY* GJ Feeding Tube-16-3.5-45.0-7-10 · Class 2 · Implant
All models of Halyard (232)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10680651956498), received between Jan 19, 2018 and Jan 19, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Fluid/Blood Leak1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 19, 2018Fluid/Blood Leak; Appropriate Device Problem Term/Code Not Available; Fluid/Blood Leak; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE FEEDING TUBE WAS EXCHANGED DUE TO LEAKAGE AND THE TUBE COMING OUT. UPON EXAMINATION, THE DOCTOR NOTICED THAT THE G PORT WAS ON THE J PORTION OF THE TUBE, AND THE J PORT WAS ON THE G PORTION OF THE TUBE. THE PARENTS DENIED SWITCHING OR TAMPERING WITH THE PORTS. PER ADDITIONAL INFORMATION RECEIVED ON 15 JAN 2018, THE FEEDING TUBE WAS IN PLACE FOR 86 DAYS AND THERE WAS NO KNOWN ADVERSE AFFECTS ON THE PATIENT. PER ADDITIONAL INFORMATION RECEIVED ON 18 JAN 2018, THE MOTHER OF THE PATIENT STATED THAT THE TUBE BEGAN LEAKING FROM THE GASTRIC PORT SEVERAL WEEKS AGO, AND EVENTUALLY THE TUBE CAME OUT PART WAY. THE MOTHER PUSHED IT BACK IN PLACE AND PUT MORE WATER IN THE BALLOON. ON INITIAL INSPECTION, THE CLOSURE DEVICE THAT USUALLY CLOSED THE GASTROSTOMY AND JEJUNOSTOMY PORT WAS NO LONGER PRESENT ON THE TUBE. IN ADDITION, THE ANTERIORLY FACING PORT WAS LABELED GASTROSTOMY AND THE LATERAL FACING PORT WAS LABELED JEJUNOSTOMY, THE OPPOSITE OF WHAT IS NORMALLY SEEN WITH THESE TUBES. THE MOTHER DID NOT SPECIFICALLY REPORT THAT THE PORTS CAME LOOSE AND WERE REPLACED. THE FEEDING TUBE WAS EXCHANGED UNDER FLUOROSCOPIC GUIDANCE. PRIOR TO THE EXCHANGE, THE PHYSICIAN INJECTED A
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7207071 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page