Level 1 FC400-16Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Level 1 (52)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10695085000501), received between Jul 26, 2017 and Jan 22, 2021 — classified as 6 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Obstruction of Flow” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Obstruction of Flow5
- Burst Container or Vessel1
- Complete Blockage1
- Invalid Sensing1
Patient problems
- Pain4
- Obstruction/Occlusion3
- Abdominal Pain1
- Device Embedded In Tissue or Plaque1
- No Consequences Or Impact To Patient1
- No Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJan 22, 2021Obstruction of Flow; Obstruction of Flow
INFORMATION RECEIVED A SMITHS MEDICAL TEMPERATURE PROBES/LEVEL 1 FOLEY CATHETERS WAS CLOGGED. THE AFFECTED LOT NUMBERS WERE PROVIDED. THE PATIENT ENDURED DISCOMFORT AS BACTERIAL CONTAMINATION OCCURRED REQUIRING TREATMENT AND LONGER STAY IN HOSPITAL. ULTRASOUND SCAN WAS PERFORMED FOR CATHETER REINSERTION. PATIENT INFORMATION ON COVER PAGE.
Patient problems: Pain
MAUDE report 11218912 ↗InjuryJan 22, 2021Obstruction of Flow; Obstruction of Flow
INFORMATION RECEIVED A SMITHS MEDICAL TEMPERATURE PROBES/LEVEL 1 FOLEY CATHETERS WAS CLOGGED. THE AFFECTED LOT NUMBERS WERE PROVIDED. THE PATIENT ENDURED DISCOMFORT AS BACTERIAL CONTAMINATION OCCURRED ,REQUIRING TREATMENT AND LONGER STAY IN HOSPITAL. ULTRASOUND SCAN WAS PERFORMED FOR CATHETER REINSERTION. PATIENT INFORMATION ON COVER PAGE.
Patient problems: Obstruction/Occlusion; Pain
MAUDE report 11218895 ↗InjuryJan 22, 2021Obstruction of Flow; Obstruction of Flow
INFORMATION RECEIVED A SMITHS MEDICAL TEMPERATURE PROBES/LEVEL 1 FOLEY CATHETERS WAS CLOGGED. THE AFFECTED LOT NUMBERS WERE PROVIDED. THE PATIENT ENDURED DISCOMFORT AS BACTERIAL CONTAMINATION OCCURRED REQUIRING TREATMENT AND LONGER STAY IN HOSPITAL. ULTRASOUND SCAN WAS PERFORMED FOR CATHETER REINSERTION. PATIENT INFORMATION ON COVER PAGE.
Patient problems: Obstruction/Occlusion; Pain
MAUDE report 11218472 ↗InjuryJan 22, 2021Obstruction of Flow; Obstruction of Flow
INFORMATION RECEIVED A SMITHS MEDICAL TEMPERATURE PROBES/LEVEL 1 FOLEY CATHETERS WAS CLOGGED. THE AFFECTED LOT NUMBERS WERE PROVIDED. THE PATIENT ENDURED DISCOMFORT AS BACTERIAL CONTAMINATION OCCURRED REQUIRING TREATMENT AND LONGER STAY IN HOSPITAL. ULTRASOUND SCAN WAS PERFORMED FOR CATHETER REINSERTION. PATIENT INFORMATION ON COVER PAGE.
Patient problems: Obstruction/Occlusion; Pain
MAUDE report 11217776 ↗MalfunctionJan 20, 2021Obstruction of Flow; Obstruction of Flow
INFORMATION RECEIVED A SMITHS MEDICAL TEMPERATURE PROBES/LEVEL 1 FOLEY CATHETERS WAS CLOGGED. THE AFFECTED LOT NUMBERS WERE PROVIDED. THE PATIENT ENDURED DISCOMFORT AS BACTERIAL CONTAMINATION OCCURRED REQUIRING TREATMENT AND LONGER STAY IN HOSPITAL. ULTRASOUND SCAN WAS PERFORMED FOR CATHETER REINSERTION. PATIENT INFORMATION ON COVER PAGE.
Patient problems: Abdominal Pain
MAUDE report 11202905 ↗MalfunctionJan 19, 2021Complete Blockage; Complete Blockage
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL CATHETER WAS CLOGGED. THIS ISSUE HAPPENED AT LEAST 5 TIMES DURING THE LAST THREE WEEKS. THERE WERE NO REPORTED ADVERSE EVENTS.
Patient problems: No Information
MAUDE report 11194522 ↗InjuryAug 18, 2017Invalid Sensing; Invalid Sensing
IT WAS REPORTED THAT DURING USE OF THIS LEVEL 1® TEMPERATURE-SENSING FOLEY CATHETER, THE TEMPERATURE DISPLAY WAS INACCURATE, CAUSING DELAY IN TREATMENT FOR INFECTION AND HYPOTHERMIA. THE CATHETER HAD BEEN IN USE APPROXIMATELY 15 DAYS PRIOR TO THIS ISSUE. THE READING MEASURED 36.1° WHEREAS THE CENTRAL CATHETER MEASURED 38.6°. MEDICAL TREATMENT WAS PRESCRIBED, AND THE INFECTION AND HYPOTHERMIA RESOLVED. NO PERMANENT INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6804980 ↗InjuryJul 26, 2017Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE BALLOON OF A LEVEL 1® FOLEY CATHETER TEMPERATURE SENSOR BURST UPON INSPECTION IN THE PATIENT'S BLADDER. THE CATHETER AND BALLOON PIECES WERE REMOVED. ALL PIECES OF THE BALLOON WERE ACCOUNTED FOR. NO PERMANENT INJURY WAS REPORTED, AND THE EVENT OUTCOME WAS REPORTED AS RESOLVED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6744939 ↗
Other CATHETER, UROLOGICAL devices
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50614C. R. Bard, Inc. · CATHETER, UROLOGICAL883 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53614GC. R. Bard, Inc. · CATHETER, UROLOGICAL692 reports
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50616C. R. Bard, Inc. · CATHETER, UROLOGICAL2 deaths443 reports
- GENTLECATH GLIDE 421567Convatec, Purchasing Department · Catheter, Urological342 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53616GC. R. Bard, Inc. · CATHETER, UROLOGICAL301 reports
- Bardex® Foley Catheter Silicone Coated 0167V22SC. R. Bard, Inc. · CATHETER, UROLOGICAL187 reports
- Bardex® Foley Catheter Silicone Coated 0167V20SC. R. Bard, Inc. · CATHETER, UROLOGICAL175 reports
- Bardia® Coude Olive Tip Urethral Foley Catheter 802516C. R. Bard, Inc. · CATHETER, UROLOGICAL170 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page