PROLENE 8741HRecalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
Quick summary
This Ethicon Prolene 8741H is a blue nonabsorbable polypropylene surgical suture. The 47 reports mostly describe the thread breaking or coming apart mid-operation, or the needle detaching from the suture during stitching, sometimes tangling and delaying surgery. No deaths and no injuries were reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 49
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 10705031000049), received between May 15, 2019 and Sep 22, 2026 — classified as 49 malfunctions.
The most frequently reported specific device problem is “Material Separation” (27 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation27
- Break23
Patient problems
- No Clinical Signs, Symptoms or Conditions43
- No Consequences Or Impact To Patient6
Recent adverse event reports
The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859341 ↗MalfunctionSep 21, 2026Break
It was reported that a patient underwent an unknown procedure on (b)(6)2026 and suture was used. It was reported by the Sales Rep, that suture continues to break despite fine previous use. They pulled lot and provided product. The device was returning. There were no patient consequences reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50842673 ↗MalfunctionSep 8, 2026Break
It was reported that a patient underwent an unknown procedure on (b)(6) 2026 and suture was used. During the procedure, the suture broke. There were no adverse patient consequences. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50670210 ↗MalfunctionJun 17, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE SURGICAL PROCEDURE, THE NEEDLE DETACHED FROM THE SUTURE MATERIAL WHILE SUTURING. THE NEEDLE-SUTURE SEPARATION OCCURRED DURING NORMAL USE. THERE WAS A SURGICAL DELAY REPORTED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25536250 ↗MalfunctionMay 20, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, DOCTOR WAS DOING CABG PROCEDURE, WHEN HE WAS DOING DISTAL ANASTOMOSIS WITH THE HELP OF PROLENE 8732H & 8741H THEN SUTURE GOT BREAKDOWN FROM THE MID POINT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25229326 ↗MalfunctionFeb 18, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP THAT THERE WAS NO CHANGE IN TECHNIQUE OR PROCEDURE THE SUTURE BROKE EASILY. DEVICE IS AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24379069 ↗MalfunctionJul 1, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC VASCULAR PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING CARDIAC VASCULAR SURGERY, NEEDLE DETACHES EASILY FROM THE SUTURE AND THE SUTURES ARE TANGLED. NO ADVERSE IMPACT TO PATIENT/USER WAS REPORTED. DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22371037 ↗MalfunctionJan 9, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFTING PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED ON THE VASCULAR ANASTOMOSIS, DURING CONTINUOUS SUTURING, THE SUTURE WAS BROKEN DURING USE. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21111252 ↗MalfunctionAug 15, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, DURING A CORONARY ARTERY BYPASS GRAFTING: CABG, DURING CONTINUOUS SUTURING, AND DURING LIGATION OF THE PERIPHERAL ANASTOMOSIS, IT HAD OCCURRED CONTINUOUSLY THE SUTURE BREAKAGE AT A POINT THAT WAS NOT THE LIGATURE POINT. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19991266 ↗MalfunctionAug 15, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6)2024 AND SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, DURING A CORONARY ARTERY BYPASS GRAFTING: CABG, DURING CONTINUOUS SUTURING, AND DURING LIGATION OF THE PERIPHERAL ANASTOMOSIS, IT HAD OCCURRED CONTINUOUSLY THE SUTURE BREAKAGE AT A POINT THAT WAS NOT THE LIGATURE POINT. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19991265 ↗MalfunctionAug 15, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6)2024 AND SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, DURING A CORONARY ARTERY BYPASS GRAFTING: CABG, DURING CONTINUOUS SUTURING, AND DURING LIGATION OF THE PERIPHERAL ANASTOMOSIS, IT HAD OCCURRED CONTINUOUSLY THE SUTURE BREAKAGE AT A POINT THAT WAS NOT THE LIGATURE POINT. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19991241 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A CABG (CORONARY ARTERY BYPASS GRAFT) PROCEDURE ON AN UNKNOWN DATE IN 2024, AND SUTURE WAS USED. DURING THE SURGERY, THE SURGEON EXPERIENCED SUTURES THAT WERE TEARING OFF. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18970772 ↗MalfunctionDec 19, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE IS COMING UNATTACHED FROM THE SUTURE. THE EXACT PASS OF THE SUTURE IS UNKNOWN WHEN THE SUTURE IS COMING APART FROM THE NEEDLE. CASE COMPLETED WITH LIKE DEVICE. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18356108 ↗MalfunctionDec 19, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE IS COMING UNATTACHED FROM THE SUTURE. THE EXACT PASS OF THE SUTURE IS UNKNOWN WHEN THE SUTURE IS COMING APART FROM THE NEEDLE. CASE COMPLETED WITH LIKE DEVICE. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18356103 ↗MalfunctionDec 19, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE IS COMING UNATTACHED FROM THE SUTURE. THE EXACT PASS OF THE SUTURE IS UNKNOWN WHEN THE SUTURE IS COMING APART FROM THE NEEDLE. CASE COMPLETED WITH LIKE DEVICE. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18356099 ↗MalfunctionDec 19, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE IS COMING UNATTACHED FROM THE SUTURE. THE EXACT PASS OF THE SUTURE IS UNKNOWN WHEN THE SUTURE IS COMING APART FROM THE NEEDLE. CASE COMPLETED WITH LIKE DEVICE. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18356098 ↗MalfunctionDec 19, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE IS COMING UNATTACHED FROM THE SUTURE. THE EXACT PASS OF THE SUTURE IS UNKNOWN WHEN THE SUTURE IS COMING APART FROM THE NEEDLE. CASE COMPLETED WITH LIKE DEVICE. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18356097 ↗MalfunctionDec 19, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE IS COMING UNATTACHED FROM THE SUTURE. THE EXACT PASS OF THE SUTURE IS UNKNOWN WHEN THE SUTURE IS COMING APART FROM THE NEEDLE. CASE COMPLETED WITH LIKE DEVICE. NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18356088 ↗MalfunctionSep 14, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. THE SUTURE WAS BROKEN DURING USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17750107 ↗MalfunctionApr 18, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A ANASTOMOSIS PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE CUSTOMER REPORTS THAT WITH THE NEW PACKAGING, THE 8741H SUTURE HAS THIS PROBLEM: THE NEEDLE DETACHES VERY EASILY FROM THE SUTURE, CREATING CONSIDERABLE INCONVENIENCE AND LITTLE SAFETY, BEING SUTURES DELICATE AND OFTEN USED IN PEDIATRIC PATIENTS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16765879 ↗MalfunctionApr 18, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A ANASTOMOSIS PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE CUSTOMER REPORTS THAT WITH THE NEW PACKAGING, THE 8741H SUTURE HAS THIS PROBLEM: THE NEEDLE DETACHES VERY EASILY FROM THE SUTURE, CREATING CONSIDERABLE INCONVENIENCE AND LITTLE SAFETY, BEING SUTURES DELICATE AND OFTEN USED IN PEDIATRIC PATIENTS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16765868 ↗MalfunctionApr 18, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A ANASTOMOSIS PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE CUSTOMER REPORTS THAT WITH THE NEW PACKAGING, THE 8741H SUTURE HAS THIS PROBLEM: THE NEEDLE DETACHES VERY EASILY FROM THE SUTURE, CREATING CONSIDERABLE INCONVENIENCE AND LITTLE SAFETY, BEING SUTURES DELICATE AND OFTEN USED IN PEDIATRIC PATIENTS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16765860 ↗MalfunctionApr 18, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A ANASTOMOSIS PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE CUSTOMER REPORTS THAT WITH THE NEW PACKAGING, THE 8741H SUTURE HAS THIS PROBLEM: THE NEEDLE DETACHES VERY EASILY FROM THE SUTURE, CREATING CONSIDERABLE INCONVENIENCE AND LITTLE SAFETY, BEING SUTURES DELICATE AND OFTEN USED IN PEDIATRIC PATIENTS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16765832 ↗MalfunctionApr 18, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A ANASTOMOSIS PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE CUSTOMER REPORTS THAT WITH THE NEW PACKAGING, THE 8741H SUTURE HAS THIS PROBLEM: THE NEEDLE DETACHES VERY EASILY FROM THE SUTURE, CREATING CONSIDERABLE INCONVENIENCE AND LITTLE SAFETY, BEING SUTURES DELICATE AND OFTEN USED IN PEDIATRIC PATIENTS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16765827 ↗MalfunctionApr 18, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A ANASTOMOSIS PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE CUSTOMER REPORTS THAT WITH THE NEW PACKAGING, THE 8741H SUTURE HAS THIS PROBLEM: THE NEEDLE DETACHES VERY EASILY FROM THE SUTURE, CREATING CONSIDERABLE INCONVENIENCE AND LITTLE SAFETY, BEING SUTURES DELICATE AND OFTEN USED IN PEDIATRIC PATIENTS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16765814 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2451-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
- Z-2457-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2456-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.