ETHILON 1663HRecalls & reported problems
Ethicon, LLC · Black Monofilament Nylon, Nonabsorbable Surgical Suture · Class 2 · Implant
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 10705031001732), received between Feb 13, 2018 and May 14, 2026 — classified as 22 malfunctions.
The most frequently reported specific device problem is “Material Separation” (14 reports).
8 other recalls cover devices cleared under the same FDA submission (K946173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation14
- Break5
- Detachment Of Device Component2
- Device Contaminated During Manufacture or Shipping1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Consequences Or Impact To Patient11
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2026Device Contaminated During Manufacture or Shipping; Material Deformation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, UPON UNPACKING, THE TIP OF THE NEEDLE IS BLUNT AND SEEMED TO BE BLACK. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25176273 ↗MalfunctionApr 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY SALES REP THAT THE SUTURE WAS NOT SECURELY SWITCHED INTO THE NEEDLE, SO THE NEEDLE FELL OFF. NOBODY GOT HURT DURING THE PROCESS BUT THE FACILITY DISPOSED OF THE ITEM. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21873852 ↗MalfunctionMar 5, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE; AND A SUTURE WAS USED. THE SUTURE BROKE DURING PULLING TO SUTURE THE WOUND. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18839430 ↗MalfunctionDec 4, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE DISTRIBUTOR THAT CUSTOMER STATES THAT THEY HAVE MULTIPLE ISSUES OF THE SUTURE BREAKING WHERE IT CONNECTS AT THE NEEDLE. ISOLATED TO THIS SPECIFIC LOT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18264151 ↗MalfunctionNov 29, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE DISTRIBUTOR THAT THE CUSTOMER STATES THAT THEY HAVE MULTIPLE ISSUES OF THE SUTURE BREAKING WHERE IT CONNECTS AT THE NEEDLE. ISOLATED TO THIS SPECIFIC LOT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE WILL BE RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18234562 ↗MalfunctionNov 29, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE DISTRIBUTOR THAT THE CUSTOMER STATES THAT THEY HAVE MULTIPLE ISSUES OF THE SUTURE BREAKING WHERE IT CONNECTS AT THE NEEDLE. ISOLATED TO THIS SPECIFIC LOT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE WILL BE RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18234561 ↗MalfunctionNov 29, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE DISTRIBUTOR THAT THE CUSTOMER STATES THAT THEY HAVE MULTIPLE ISSUES OF THE SUTURE BREAKING WHERE IT CONNECTS AT THE NEEDLE. ISOLATED TO THIS SPECIFIC LOT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE WILL BE RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18234557 ↗MalfunctionSep 6, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT THE CUSTOMER HAD SUTURES FROM SAME LOT BREAK DURING A PROCEDURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17691391 ↗MalfunctionSep 6, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT THE CUSTOMER HAD SUTURES FROM SAME LOT BREAK DURING A PROCEDURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17691386 ↗MalfunctionDec 15, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE ETHILON SUTURE HAS BEEN BREAKING WHEN CLOSING INCISIONS. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE HAS BEEN DISCARDED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15992086 ↗MalfunctionDec 12, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON 11/16/2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE ETHILON SUTURE HAS BEEN BREAKING WHEN CLOSING INCISIONS. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE HAS BEEN DISCARDED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15959934 ↗MalfunctionAug 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE POPPED OFF THE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8931358 ↗MalfunctionAug 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A KNEE SURGERY ON (B)(6) 2019 AND SUTURE WAS USED TO CLOSE THE SKIN. DURING THE PROCEDURE, THE SUTURE POPPED OFF THE NEEDLE WHEN DRAWING THROUGH THE SKIN. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8931330 ↗MalfunctionAug 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A KNEE SURGERY ON (B)(6) 2019 AND SUTURE WAS USED TO CLOSE THE SKIN. DURING THE PROCEDURE, THE SUTURE POPPED OFF THE NEEDLE WHEN DRAWING THROUGH THE SKIN. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8931329 ↗MalfunctionJun 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE NEEDLE POPPED OFF WHEN IN USE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8737232 ↗MalfunctionJun 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE NEEDLE POPPED OFF WHEN IN USE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8737230 ↗MalfunctionJun 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE NEEDLE POPPED OFF WHEN IN USE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8737211 ↗MalfunctionJun 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE NEEDLE POPPED OFF WHEN IN USE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8737208 ↗MalfunctionJun 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE NEEDLE POPPED OFF WHEN IN USE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8737203 ↗MalfunctionJun 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE NEEDLE POPPED OFF WHEN IN USE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8737187 ↗MalfunctionFeb 13, 2018Detachment Of Device Component; Detachment Of Device Component
IT WAS REPORTED THAT THE PATIENT UNDERWENT A VASCULAR PROCEDURE IN (B)(6) 2017 AND SUTURE WAS USED FOR WOUND CLOSURE. DURING THE PROCEDURE, THE SUTURE FROM THE NEEDLE. SUTURE FROM A DIFFERENT LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7266997 ↗MalfunctionFeb 13, 2018Detachment Of Device Component; Detachment Of Device Component
IT WAS REPORTED THAT THE PATIENT UNDERWENT A VASCULAR PROCEDURE IN (B)(6) 2017 AND SUTURE WAS USED FOR WOUND CLOSURE. DURING THE PROCEDURE, THE SUTURE FROM THE NEEDLE. SUTURE FROM A DIFFERENT LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7266981 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K946173. One submission often covers many models, so check the product and lot details in each record.
- Z-1516-2025 · Mar 7, 2025 · Open, ClassifiedETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
- Z-1097-2025 · Dec 20, 2024 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
- Z-2486-2020 · Jun 10, 2020 · Open, ClassifiedPERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
- Z-1841-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
- Z-1842-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
- Z-0747-2019 · Dec 18, 2018 · TerminatedPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
- Z-1839-2015 · May 8, 2015 · TerminatedEthicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
- Z-0492-2015 · Oct 17, 2014 · TerminatedEthicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE devices
- ETHILON 697GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE223 reports
- ETHILON 698GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE119 reports
- ETHILON 1696GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE118 reports
- Monosof SN-3695Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls3 reports
- McKesson S931BXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls2 reports
- McKesson S661GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls2 reports
- Surgilon 88861919-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls2 reports
- McKesson S697GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.