ETHILON 7756GRecalls & reported problems
Ethicon, LLC · Black Monofilament Nylon, Nonabsorbable Surgical Suture · Class 2 · Implant
All models of ETHILON (133)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705031010840), received between Jul 16, 2025 and May 28, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Break” (12 reports).
8 other recalls cover devices cleared under the same FDA submission (K946173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break12
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT # 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319193 ↗MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT # 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319180 ↗MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT # 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319167 ↗MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT#: 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319165 ↗MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT # 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319163 ↗MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT # 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319160 ↗MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT#: 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319143 ↗MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT # 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319142 ↗MalfunctionMay 20, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT # 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25229199 ↗MalfunctionMay 20, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT, THE 10-0 ETHILON 7756G SUTURE BROKE MULTIPLE TIMES WHILE SURGEON WAS SUTURING. MULTIPLE SUTURE PACKS WERE PASSED TO STERILE FIELD AND MOST BROKE. ALL FROM LOT # 10BH3E. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25229197 ↗MalfunctionJul 16, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE TIP OF THE NEEDLE BROKE OFF OR WAS BENDING. THERE WERE NO ADVERSE PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22520691 ↗MalfunctionJul 16, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE TIP OF THE NEEDLE BROKE OFF OR WAS BENDING. THERE WERE NO ADVERSE PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22520690 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K946173. One submission often covers many models, so check the product and lot details in each record.
- Z-1516-2025 · Mar 7, 2025 · Open, ClassifiedETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
- Z-1097-2025 · Dec 20, 2024 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
- Z-2486-2020 · Jun 10, 2020 · Open, ClassifiedPERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
- Z-1841-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
- Z-1842-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
- Z-0747-2019 · Dec 18, 2018 · TerminatedPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
- Z-1839-2015 · May 8, 2015 · TerminatedEthicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
- Z-0492-2015 · Oct 17, 2014 · TerminatedEthicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE devices
- ETHILON 697GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE223 reports
- ETHILON 698GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE119 reports
- ETHILON 1696GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE118 reports
- Monosof SN-3695Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls3 reports
- McKesson S931BXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls2 reports
- McKesson S661GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls2 reports
- Surgilon 88861919-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls2 reports
- McKesson S697GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page