Skip to content

PROLENE 8522HRecalls & reported problems

Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant

All models of PROLENE (358)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Prolene suture is a blue synthetic thread used to close tissue during surgery. Of its 90 reports, 3 involve injuries and no deaths are recorded. Reports mostly describe the thread breaking or splitting during surgery. There were 26 reports in the last 12 months, up from 6 before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
99
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 99 adverse event reports that list this exact device (DI 10705031016569), received between Dec 26, 2018 and Sep 30, 2026 — classified as 3 injuries, 96 malfunctions.

The most frequently reported specific device problem is “Break” (74 reports).

21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 9 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 9 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 9 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break74
  • Material Separation17
  • Material Frayed6
  • Adverse Event Without Identified Device or Use Problem2
  • Component Missing1
  • Defective Component1
  • Delivered as Unsterile Product1
  • Device Markings/Labelling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions85
  • No Consequences Or Impact To Patient9
  • Failure to Anastomose2
  • Insufficient Information2
  • Not Applicable2
  • Unspecified Tissue Injury1

Recent adverse event reports

The 25 most recent of 99 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951521 ↗
  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951520 ↗
  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951518 ↗
  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951517 ↗
  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951516 ↗
  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951515 ↗
  • MalfunctionSep 24, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889971 ↗
  • MalfunctionSep 16, 2026Break

    It was reported that a patient underwent an unknown procedure on an unknown date and suture was used. The suture was broken when the surgeon attempted to sew the tissue. Changed another one to continue the surgery, the same problem happened. Changed another one to complete the surgery. No adverse patient consequences were reported. Additional information was requested

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787693 ↗
  • MalfunctionSep 16, 2026Break

    It was reported that a patient underwent an unknown procedure on an unknown date and suture was used. The suture was broken when the surgeon attempted to sew the tissue. Changed another one to continue the surgery, the same problem happened. Changed another one to complete the surgery. No adverse patient consequences were reported. Additional information was requested

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787549 ↗
  • MalfunctionMay 19, 2026Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE GOT SPLIT IN THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. ENCLOSED PLEASE FIND DEFECT PHOTO. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25216796 ↗
  • MalfunctionMay 19, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MYOMECTOMY PROCEDURE ON (B)(6)2026 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN THE SUTURE WAS OPENED, THERE WAS AN INDENTATION ON THE SUTURE ,AND THE SUTURE BROKE DURING USE . CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25216505 ↗
  • MalfunctionMay 18, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25202353 ↗
  • MalfunctionMay 5, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE EASILY DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25086627 ↗
  • MalfunctionApr 20, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE . CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24928093 ↗
  • MalfunctionApr 20, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24928091 ↗
  • MalfunctionApr 20, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE . CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24928089 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LARGE VESSEL SURGERY ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. SEVEN PACKAGES THAT WAS UNPACKED BY THE DOCTOR BUT NOT USED, THEY WILL BE RETURNED FOR ANALYSIS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789153 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LARGE VESSEL SURGERY ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. SEVEN PACKAGES THAT WAS UNPACKED BY THE DOCTOR BUT NOT USED, THEY WILL BE RETURNED FOR ANALYSIS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789151 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LARGE VESSEL SURGERY ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. SEVEN PACKAGES THAT WAS UNPACKED BY THE DOCTOR BUT NOT USED, THEY WILL BE RETURNED FOR ANALYSIS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789150 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VALVE REPLACEMENT SURGERY ON (B)(6) 2026 AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789023 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VALVE REPLACEMENT SURGERY ON (B)(6) 2026 AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789009 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VALVE REPLACEMENT SURGERY ON (B)(6) 2026 AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789006 ↗
  • MalfunctionFeb 20, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24400850 ↗
  • MalfunctionDec 19, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT HEART SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. WHEN USING THIS SUTURE MATERIAL, SEVERAL THREADS ARE BROKEN AT LOW TENSION. IT HAPPENED WHILE TRYING THE KNOT IN THE THREAD. TISSUE AURICLE. THE THREAD HAS BROKEN IN THE MIDDLE IN RELATION TO THE NEEDLE. THERE WAS NO DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23850451 ↗
  • MalfunctionDec 19, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT HEART SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. WHEN USING THIS SUTURE MATERIAL, SEVERAL THREADS ARE BROKEN AT LOW TENSION. IT HAPPENED WHILE TRYING THE KNOT IN THE THREAD. TISSUE AURICLE. THE THREAD HAS BROKEN IN THE MIDDLE IN RELATION TO THE NEEDLE. THERE WAS NO DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23850447 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.

  • Z-1515-2025 · Mar 7, 2025 · Open, Classified
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
  • Z-1096-2025 · Dec 20, 2024 · Open, Classified
    PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
  • Z-2464-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
  • Z-2455-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
  • Z-2457-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
  • Z-2453-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
  • Z-2449-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
  • Z-2454-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
  • Z-2463-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
  • Z-2466-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE 122CM M1.5, D10189
  • Z-2448-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
  • Z-2451-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
  • Z-2452-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
  • Z-2460-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
  • Z-2465-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027

…and 6 more.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page