PROLENE 8523HRecalls & reported problems

Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant

All models of PROLENE (356)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Prolene 8523H is a blue nonabsorbable polypropylene surgical suture. Most of the 64 reports describe the thread snapping, splitting or fraying during surgery, or coming apart, and a few mention packaging contaminated with fiber or a needle pulling loose from the thread. There were 13 injuries and no deaths. No recalls name this device. Reports rose from 5 in the previous 12 months to 20 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
63
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 63 adverse event reports that list this exact device (DI 10705031016590), received between Dec 27, 2017 and May 28, 2026 — classified as 12 injuries, 51 malfunctions.

The most frequently reported specific device problem is “Break” (47 reports).

21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 16 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break47
  • Material Separation13
  • Material Frayed12
  • Device Contaminated During Manufacture or Shipping2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions45
  • Not Applicable11
  • No Consequences Or Impact To Patient7
  • Insufficient Information2
  • Adhesion(s)1
  • Unspecified Tissue Injury1
  • Urolithiasis1

Recent adverse event reports

The 25 most recent of 63 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2026Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY SISTER THAT A SUTURES BIFURCATED DURING CABG PROCEDURE. THE ISSUE WAS IDENTIFIED DURING INTRAOPERATIVE USE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25322583 ↗
  • MalfunctionFeb 23, 2026Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS PRODUCT WAS REJECTED AT INCOMING INSPECTION DUE TO A FOREIGN FIBER MATTER THAT WAS IN THE PACKAGE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24418021 ↗
  • MalfunctionFeb 23, 2026Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS PRODUCT WAS REJECTED AT INCOMING INSPECTION DUE TO A FOREIGN FIBER MATTER THAT WAS IN THE PACKAGE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24418013 ↗
  • MalfunctionNov 26, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HERNIA REPAIR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE SURGICAL LIGATION OF THE FEMORAL GROOVE HERNIA, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED AND A NEW SUTURE WAS IMMEDIATELY REPLACED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23658178 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534110 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534104 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534094 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534090 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534089 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534085 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534067 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534063 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534058 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534057 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534038 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534034 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534021 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534005 ↗
  • MalfunctionNov 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, IT WAS FOUND THAT THE SUTURE HAD BEEN BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23458180 ↗
  • MalfunctionJun 13, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES BREAK DURING SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22204753 ↗
  • MalfunctionJun 13, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES BREAK DURING SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22204752 ↗
  • MalfunctionJun 13, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES BREAK DURING SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22204750 ↗
  • MalfunctionApr 4, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOT-ASSISTED TOTAL GASTRECTOMY AND UROLIFT ON AN UNKNOWN DATE AND SUTURE WAS USED. THE NEEDLE HAD COME OFF DURING TRACTION ON THE SUTURE. THE TRACTION ON THE SUTURE MAY HAVE BEEN TOO STRONG, BUT IT IS VERY DANGEROUS FOR THE SUTURE TO COME OFF FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21772738 ↗
  • MalfunctionDec 11, 2024Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20901471 ↗
  • InjurySep 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BY A LSC SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED TO FIX THE VAGINA AND MESH. POST-OP, THE PATIENT HAD A URINARY STONE AFTER THE SURGERY AND THE BLADDER WAS CHECKED TRANSURETHRAL SCOPICALLY WITH A CYSTOSCOPE. THE STUMP OF THE SUTURE THAT HAD BEEN USED TO SUTURE THE VAGINA TO THE MESH, WAS FOUND TO HAVE PENETRATED THE BLADDER. THE ADHESION OCCURRED ON THE STUMP AND DEVELOPED INTO STONES. THE PATIENT UNDER WENT RE-OPERATION. THE STONE WAS DRAINED TRANSURETHRAL SCOPICALLY AND CUT THE STUMP OF THE SUTURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Adhesion(s); Unspecified Tissue Injury; Urolithiasis

    MAUDE report 20311209 ↗

FDA clearance and classification

Cleared through Special 510(k) K133356 on Dec 6, 2013 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Prolene Polypropylene Nonabsorbable Suture”. The FDA review took 36 days. FDA record ↗

The FDA regulates product code GAW (suture, nonabsorbable, synthetic, polypropylene) under 21 CFR 878.5010 as a Class II device. 21 CFR 878.5010 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.

  • Z-1515-2025 · Mar 7, 2025 · Open, Classified
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
  • Z-1096-2025 · Dec 20, 2024 · Open, Classified
    PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
  • Z-2455-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
  • Z-2454-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
  • Z-2463-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
  • Z-2449-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
  • Z-2458-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
  • Z-2464-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
  • Z-2465-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
  • Z-2452-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
  • Z-2450-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
  • Z-2456-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
  • Z-2461-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
  • Z-2448-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
  • Z-2460-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662

…and 6 more.

Common questions about the PROLENE 8523H

Has the PROLENE 8523H been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705031016590), based on the FDA recall file dated Nov 27, 2026. However, 21 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the PROLENE 8523H?

The FDA's MAUDE database lists 63 adverse event reports for this device, received between Dec 27, 2017 and May 28, 2026. In those reports, the most common specific device problems are “Break” (47), “Material Separation” (13) and “Material Frayed” (12); the most common patient problems are “Not Applicable” (11), “Adhesion(s)” (1) and “Unspecified Tissue Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PROLENE 8523H?

None of the reports are classified as deaths, 12 as injuries and 51 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PROLENE 8523H FDA approved?

It was cleared through Special 510(k) K133356 on Dec 6, 2013. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the PROLENE 8523H MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the PROLENE 8523H, and is it still sold?

The FDA's device database lists Ethicon, LLC as the company responsible for it (brand PROLENE, model 8523H). Its record shows it as in commercial distribution.