PROLENE 8523HRecalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
Quick summary
This Ethicon Prolene 8523H is a blue nonabsorbable polypropylene surgical suture. Most of the 64 reports describe the thread snapping, splitting or fraying during surgery, or coming apart, and a few mention packaging contaminated with fiber or a needle pulling loose from the thread. There were 13 injuries and no deaths. No recalls name this device. Reports rose from 5 in the previous 12 months to 20 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 63
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 63 adverse event reports that list this exact device (DI 10705031016590), received between Dec 27, 2017 and May 28, 2026 — classified as 12 injuries, 51 malfunctions.
The most frequently reported specific device problem is “Break” (47 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break47
- Material Separation13
- Material Frayed12
- Device Contaminated During Manufacture or Shipping2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions45
- Not Applicable11
- No Consequences Or Impact To Patient7
- Insufficient Information2
- Adhesion(s)1
- Unspecified Tissue Injury1
- Urolithiasis1
Recent adverse event reports
The 25 most recent of 63 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 28, 2026Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY SISTER THAT A SUTURES BIFURCATED DURING CABG PROCEDURE. THE ISSUE WAS IDENTIFIED DURING INTRAOPERATIVE USE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25322583 ↗MalfunctionFeb 23, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS PRODUCT WAS REJECTED AT INCOMING INSPECTION DUE TO A FOREIGN FIBER MATTER THAT WAS IN THE PACKAGE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24418021 ↗MalfunctionFeb 23, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS PRODUCT WAS REJECTED AT INCOMING INSPECTION DUE TO A FOREIGN FIBER MATTER THAT WAS IN THE PACKAGE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24418013 ↗MalfunctionNov 26, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A HERNIA REPAIR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE SURGICAL LIGATION OF THE FEMORAL GROOVE HERNIA, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED AND A NEW SUTURE WAS IMMEDIATELY REPLACED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23658178 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534110 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534104 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534094 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534090 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534089 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534085 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534067 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534063 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534058 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534057 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534038 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534034 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534021 ↗MalfunctionNov 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC GASTROSTOMY FOR DYSPHAGIA, THE PRODUCT WAS USED ON THE GASTRIC MUCOSA FOR SUTURING. DURING THE SUTURING PROCEDURE, WHEN THE NEEDLE WAS PASSED THROUGH THE TISSUE AND PULLED, THE SUTURE WAS BROKEN A FEW TIMES AT A POINT A LITTLE AWAY FROM THE SWAGE PART. THEREFORE, THE SUTURE WAS REPLACED WITH A NEW ONE AND THE SUTURING WAS PERFORMED AGAIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534005 ↗MalfunctionNov 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, IT WAS FOUND THAT THE SUTURE HAD BEEN BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23458180 ↗MalfunctionJun 13, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES BREAK DURING SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22204753 ↗MalfunctionJun 13, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES BREAK DURING SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22204752 ↗MalfunctionJun 13, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES BREAK DURING SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22204750 ↗MalfunctionApr 4, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOT-ASSISTED TOTAL GASTRECTOMY AND UROLIFT ON AN UNKNOWN DATE AND SUTURE WAS USED. THE NEEDLE HAD COME OFF DURING TRACTION ON THE SUTURE. THE TRACTION ON THE SUTURE MAY HAVE BEEN TOO STRONG, BUT IT IS VERY DANGEROUS FOR THE SUTURE TO COME OFF FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21772738 ↗MalfunctionDec 11, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20901471 ↗InjurySep 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BY A LSC SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED TO FIX THE VAGINA AND MESH. POST-OP, THE PATIENT HAD A URINARY STONE AFTER THE SURGERY AND THE BLADDER WAS CHECKED TRANSURETHRAL SCOPICALLY WITH A CYSTOSCOPE. THE STUMP OF THE SUTURE THAT HAD BEEN USED TO SUTURE THE VAGINA TO THE MESH, WAS FOUND TO HAVE PENETRATED THE BLADDER. THE ADHESION OCCURRED ON THE STUMP AND DEVELOPED INTO STONES. THE PATIENT UNDER WENT RE-OPERATION. THE STONE WAS DRAINED TRANSURETHRAL SCOPICALLY AND CUT THE STUMP OF THE SUTURE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Adhesion(s); Unspecified Tissue Injury; Urolithiasis
MAUDE report 20311209 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2451-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2457-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2450-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
- Z-2456-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
- Z-2454-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
- Z-2449-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.