Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon PROLENE suture is a blue monofilament polypropylene surgical suture. Reports mostly describe the needle separating from the suture or material breakage during surgery, often during CABG; many note no patient harm, while some mention bleeding. There are 100 adverse event reports, with no deaths and no injuries. Reports rose from 4 in the prior 12 months to 34 in the last 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
102
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 102 adverse event reports that list this exact device (DI 10705031019348), received between Dec 24, 2019 and Sep 28, 2026 — classified as 102 malfunctions.

The most frequently reported specific device problem is “Break” (66 reports).

21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 15 reportsApr 2026: 2 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 6 reportsAug 2026: 0 reportsSep 2026: 6 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break66
  • Material Separation36
  • Material Frayed4

Patient problems

  • No Clinical Signs, Symptoms or Conditions87
  • No Consequences Or Impact To Patient11
  • Hemorrhage/Blood Loss/Bleeding4

Recent adverse event reports

The 25 most recent of 102 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923804 ↗
  • MalfunctionSep 28, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916739 ↗
  • MalfunctionSep 28, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916702 ↗
  • MalfunctionSep 28, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916691 ↗
  • MalfunctionSep 28, 2026Material Separation

    It was reported that a patient underwent an unknown procedure on an unknown date and suture was used. Several defected 7-0 sutures had the Needles popping off the thread too easily. No adverse patient consequences were reported. Additional information was requested.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916671 ↗
  • MalfunctionSep 28, 2026Material Separation

    It was reported that a patient underwent an unknown procedure on an unknown date and suture was used. Several defected 7-0 sutures had the Needles popping off the thread too easily. No adverse patient consequences were reported. Additional information was requested.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916660 ↗
  • MalfunctionJul 28, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50056924 ↗
  • MalfunctionJul 15, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25825327 ↗
  • MalfunctionJul 15, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25825323 ↗
  • MalfunctionJul 15, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25825322 ↗
  • MalfunctionJul 15, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25825321 ↗
  • MalfunctionJul 15, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25825319 ↗
  • MalfunctionMay 1, 2026Break; Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CV CABG SURGERY ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN CLOSING THE DISTAL ANASTOMOSIS, THE SUTURE BROKE, PULLED APART, FELL APART AND DISINTEGRATED. IT HAPPENED IN DIFFERENT PLACES ON EACH SUTURE, NOT WHERE THE NEEDLE AND SUTURE MEET THAT STAYED INTACT AND IT WAS NOT EXTRA TENSION ON THE SUTURE. THE PATIENT SHOWED BLEEDING ON THOSE ANASTOMOSIS SPOTS AND HEMOSTASIS PRODUCTS WERE USED ON TOP OF EACH SUTURE. ANOTHER LIKE DEVICE WAS USED TO FINISH THE CASE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25055668 ↗
  • MalfunctionMay 1, 2026Break; Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CV CABG SURGERY ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN CLOSING THE DISTAL ANASTOMOSIS, THE SUTURE BROKE, PULLED APART, FELL APART AND DISINTEGRATED. IT HAPPENED IN DIFFERENT PLACES ON EACH SUTURE, NOT WHERE THE NEEDLE AND SUTURE MEET THAT STAYED INTACT AND IT WAS NOT EXTRA TENSION ON THE SUTURE. THE PATIENT SHOWED BLEEDING ON THOSE ANASTOMOSIS SPOTS AND HEMOSTASIS PRODUCTS WERE USED ON TOP OF EACH SUTURE. ANOTHER LIKE DEVICE WAS USED TO FINISH THE CASE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25055664 ↗
  • MalfunctionMay 1, 2026Break; Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CV CABG SURGERY ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN CLOSING THE DISTAL ANASTOMOSIS, THE SUTURE BROKE, PULLED APART, FELL APART AND DISINTEGRATED. IT HAPPENED IN DIFFERENT PLACES ON EACH SUTURE, NOT WHERE THE NEEDLE AND SUTURE MEET THAT STAYED INTACT AND IT WAS NOT EXTRA TENSION ON THE SUTURE. THE PATIENT SHOWED BLEEDING ON THOSE ANASTOMOSIS SPOTS AND HEMOSTASIS PRODUCTS WERE USED ON TOP OF EACH SUTURE. ANOTHER LIKE DEVICE WAS USED TO FINISH THE CASE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25055657 ↗
  • MalfunctionMay 1, 2026Break; Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CV CABG SURGERY ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN CLOSING THE DISTAL ANASTOMOSIS, THE SUTURE BROKE, PULLED APART, FELL APART AND DISINTEGRATED. IT HAPPENED IN DIFFERENT PLACES ON EACH SUTURE, NOT WHERE THE NEEDLE AND SUTURE MEET THAT STAYED INTACT AND IT WAS NOT EXTRA TENSION ON THE SUTURE. THE PATIENT SHOWED BLEEDING ON THOSE ANASTOMOSIS SPOTS AND HEMOSTASIS PRODUCTS WERE USED ON TOP OF EACH SUTURE. ANOTHER LIKE DEVICE WAS USED TO FINISH THE CASE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25055652 ↗
  • MalfunctionApr 9, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE IN (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SURGEON TO THE SALES REP THAT MULTIPLE SUTURES WERE BREAKING. EVERY SUTURE THEY TRIED HAS BROKEN WHICH MEANS THIS HAPPENED MULTIPLE TIMES. NO OTHER INFORMATION PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24833500 ↗
  • MalfunctionApr 8, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY THE SURGEON TO THE SALES REP THAT MULTIPLE SUTURES WERE BREAKING. EVERY SUTURE THEY TRIED HAS BROKEN WHICH MEANS THIS HAPPENED MULTIPLE TIMES. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24815967 ↗
  • MalfunctionMar 26, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT MULTIPLE SUTURES BROKE WHILE DOING ANASTOMOSIS ON THE HEART OR CAROTID TISSUE. THEY WOULD BREAK ANYWHERE FROM THREE THROWS IN TO HALFWAY AROUND THE ANASTOMOSIS. CUSTOMER REPORTS THAT THIS HAPPENED TO 3 OF THEIR 4 SURGEONS IN AT LEAST 5 CASES, BEFORE SUTURES WERE PULLED FROM THE SHELF. THEY BECAME AWARE OF THE ISSUE 2/5 & AGAIN ON 2/7 PULLING ALL THE BOXES ON 2/9. THERE IS NO PATIENT OR END-USER INJURY REPORTED. DEVICES ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24706409 ↗
  • MalfunctionMar 26, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT MULTIPLE SUTURES BROKE WHILE DOING ANASTOMOSIS ON THE HEART OR CAROTID TISSUE. THEY WOULD BREAK ANYWHERE FROM THREE THROWS IN TO HALFWAY AROUND THE ANASTOMOSIS. CUSTOMER REPORTS THAT THIS HAPPENED TO 3 OF THEIR 4 SURGEONS IN AT LEAST 5 CASES, BEFORE SUTURES WERE PULLED FROM THE SHELF. THEY BECAME AWARE OF THE ISSUE 2/5 & AGAIN ON 2/7 PULLING ALL THE BOXES ON 2/9. THERE IS NO PATIENT OR END-USER INJURY REPORTED. DEVICES ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698632 ↗
  • MalfunctionMar 26, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT MULTIPLE SUTURES BROKE WHILE DOING ANASTOMOSIS ON THE HEART OR CAROTID TISSUE. THEY WOULD BREAK ANYWHERE FROM THREE THROWS IN TO HALFWAY AROUND THE ANASTOMOSIS. CUSTOMER REPORTS THAT THIS HAPPENED TO 3 OF THEIR 4 SURGEONS IN AT LEAST 5 CASES, BEFORE SUTURES WERE PULLED FROM THE SHELF. THEY BECAME AWARE OF THE ISSUE 2/5 & AGAIN ON 2/7 PULLING ALL THE BOXES ON 2/9. THERE IS NO PATIENT OR END-USER INJURY REPORTED. DEVICES ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698627 ↗
  • MalfunctionMar 26, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT MULTIPLE SUTURES BROKE WHILE DOING ANASTOMOSIS ON THE HEART OR CAROTID TISSUE. THEY WOULD BREAK ANYWHERE FROM THREE THROWS IN TO HALFWAY AROUND THE ANASTOMOSIS. CUSTOMER REPORTS THAT THIS HAPPENED TO 3 OF THEIR 4 SURGEONS IN AT LEAST 5 CASES, BEFORE SUTURES WERE PULLED FROM THE SHELF. THEY BECAME AWARE OF THE ISSUE 2/5 & AGAIN ON 2/7 PULLING ALL THE BOXES ON 2/9. THERE IS NO PATIENT OR END-USER INJURY REPORTED. DEVICES ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698625 ↗
  • MalfunctionMar 26, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT MULTIPLE SUTURES BROKE WHILE DOING ANASTOMOSIS ON THE HEART OR CAROTID TISSUE. THEY WOULD BREAK ANYWHERE FROM THREE THROWS IN TO HALFWAY AROUND THE ANASTOMOSIS. CUSTOMER REPORTS THAT THIS HAPPENED TO 3 OF THEIR 4 SURGEONS IN AT LEAST 5 CASES, BEFORE SUTURES WERE PULLED FROM THE SHELF. THEY BECAME AWARE OF THE ISSUE (B)(6) & AGAIN ON (B)(6) PULLING ALL THE BOXES ON (B)(6). THERE IS NO PATIENT OR END-USER INJURY REPORTED. DEVICES ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698617 ↗
  • MalfunctionMar 26, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT MULTIPLE SUTURES BROKE WHILE DOING ANASTOMOSIS ON THE HEART OR CAROTID TISSUE. THEY WOULD BREAK ANYWHERE FROM THREE THROWS IN TO HALFWAY AROUND THE ANASTOMOSIS. CUSTOMER REPORTS THAT THIS HAPPENED TO 3 OF THEIR 4 SURGEONS IN AT LEAST 5 CASES, BEFORE SUTURES WERE PULLED FROM THE SHELF. THEY BECAME AWARE OF THE ISSUE 2/5 & AGAIN ON 2/7 PULLING ALL THE BOXES ON 2/9. THERE IS NO PATIENT OR END-USER INJURY REPORTED. DEVICES ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698583 ↗
  • MalfunctionMar 26, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT MULTIPLE SUTURES BROKE WHILE DOING ANASTOMOSIS ON THE HEART OR CAROTID TISSUE. THEY WOULD BREAK ANYWHERE FROM THREE THROWS IN TO HALFWAY AROUND THE ANASTOMOSIS. CUSTOMER REPORTS THAT THIS HAPPENED TO 3 OF THEIR 4 SURGEONS IN AT LEAST 5 CASES, BEFORE SUTURES WERE PULLED FROM THE SHELF. THEY BECAME AWARE OF THE ISSUE 2/5 & AGAIN ON 2/7 PULLING ALL THE BOXES ON 2/9. THERE IS NO PATIENT OR END-USER INJURY REPORTED. DEVICES ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698582 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.

  • Z-1515-2025 · Mar 7, 2025 · Open, Classified
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
  • Z-1096-2025 · Dec 20, 2024 · Open, Classified
    PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
  • Z-2459-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
  • Z-2466-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE 122CM M1.5, D10189
  • Z-2451-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
  • Z-2465-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
  • Z-2449-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
  • Z-2450-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
  • Z-2463-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
  • Z-2458-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
  • Z-2464-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
  • Z-2455-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
  • Z-2454-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
  • Z-2457-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
  • Z-2461-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746

…and 6 more.